    HeNe Laser Testing, Adjustment, Repair

    Sub-Table of Contents



  • Back to Sam's Laser FAQ Table of Contents.
  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    HeNe Problems and Testing

    How Can I Tell if My Tube is Good?

    A variety of faults can result in a HeNe tube not working properly. These can be the result of a large number of problems including a bad tube, a bad power supply, incorrect input voltage (or configuration), incompatibility between the tube and power supply, bad or incorrect ballast resistor, or defective wiring.

    If all you have is a HeNe tube but no power supply, see the section: Testing a HeNe Laser Tube Without a Compatible Power Supply for ways to determine if the tube is good. The following applies to both bare HeNe tubes and laser heads though some of the inspection and/or tests will require removing the tube from any enclosure.

    Several types of problems can prevent a HeNe tube from lasing properly or make it hard to start:

    It probably doesn't make sense to spend a lot of effort, time, or money to revive a 1 mW HeNe tube that can be replaced for $15. However, if you are ambitious or a new tube cannot be substituted easily (e.g., due to mounting restrictions), see the sections starting with: Repairing Leaky or Broken HeNe Tubes.

    Testing a HeNe Laser Tube Without a Compatible Power Supply

    First time laser enthusiasts are often confronted with this problem: Becoming the proud owner of a HeNe laser tube but no HeNe laser power supply and possibly no specs for the tube or assurance that it is even good. What to do? How can a quick inexpensive test be done to determine if the tube works?

    First, you need to determine the tube's power connections. See the section: Identifying Connections to Unmarked HeNe Tube or Laser Head if you aren't sure.

    There are many ways to power a HeNe tube for the purposes of seeing if it produces a beam. Almost anything that can provide enough voltage to get a few mA through the tube will result in at least a momentary flash of laser light out the end if the tube is good. There won't be any way of determining output power or whether the tube meets specs, but the knowledge that it lases at all may be enough to take the next step - the purchase or construction of a proper power supply.

    It is easy to use the family microwave to see if the tube is gas-intact if the tube will fit inside. See the section: Using a Microwave Oven to Evaluate and Revive HeNe Laser Tubes. While this won't tell you if the tube lases, if it fails this test, there is no need to go further.

    To test for lasing, current must be passed through the bore of the tube. A couple of options for a quick test power supply are:

    Even a high voltage AC supply with appropriate current limiting can be used safely for a few seconds only. (I've been sent HeNe laser tubes which have been operated on AC because the owner copied some power supply design off the Web and didn't know any better. The output power (what of it there was) gradually declined over a few minutes and then there was none.) And even with the rectifier voltage, the tube will be restarting once per cycle which is hard on it so don't run that for too long either. None of these are suitable to operate a HeNe tube continuously unless proper filtering and starting circuitry is added to turn it into a proper HeNe laser power supply.

    Don't go overboard though: Too high a voltage applied in the wrong place can arc straight though the glass at which point you have a rather boring high-tech sculpture. :( A very high current can also damage the tube very quickly, thus the need for the current limiting ballast resistance.

    With these power supplies driving the tube, if there is any output beam, even if it is weak or in the form of short flashes, the tube is probably good. However, there is no way to tell if it meets specs since HeNe laser output power is only maximum over a narrow range of tube current and these quick test power supplies are at most controlling only average current, not instantaneous current as would be the case with a real HeNe power supply. But, at least you know the tube isn't dead.

    Quick and Dirty Universal HeNe Laser Test Power Supply

    The following simple device can be used to confirm that a HeNe laser tube or head is not totally dead. It won't be able to help with output power or be safe to run for more than a few seconds but should produce some coherent output from any HeNe laser from 0.5 to 40 mW or more.

    It consists of the following components in series:

    Wire the output of the transformer in series with the rectifier(s) and ballast resistors. The positive output goes to the anode of the HeNe head or tube; the negative to the cathode. It doesn't matter whether the laser has an internal ballast resistor. Insulate everything VERY well. :)

    Powering the laser should result in flashes of coherent light, probably at the power line frequency (60 or 50 Hz). The amount of light will not be that impressive even with a perfectly good high power laser since the current is nowhere near optimal for any length of time, if ever. However, the presence of laser output would confirm that there is life.

    WARNING: Since centertap of transformer secondary should be grounded, both outputs of the power supply will be floating with respect to ground. Take care.

    HeNe Tube Lases but Color of Discharge Changes Along Length of Bore

    I've only seen this on a couple of HeNe tubes but the appearance is quite strange. At the anode-end of one, which I call the "Northern Lights" tube, the discharge color is perfectly normal - white-ish red-orange (salmon color), just as would be expected in any self respecting HeNe tube. However moving toward the cathode-end of the tube, the color changes to what appears to be a much more blue-ish white color (relative to the normal white-ish red-orange, though in absolute terms it may be more of a pink), typical of some sort of contamination. However, having a tube with the behavior provided a way (as you shall see below) of relating power output to the color of its discharge.

    In this condition, the tube still lased at a power level which relative to its rated output, is approximately proportional to how much of the bore has the correct color. In this sample, about 2 mW for a tube specified at 4 to 5 mW. I don't believe the starting or operating voltage has been affected very much.

    The explanation that makes the most sense is that due to the discharge current in the bore, the few N2 and O2 atoms (and any other party poopers that may have entered without an invitation) are being ionized and pushed toward the cathode of the tube leaving the desired helium and neon atoms to play at the anode-end. The contamination, whether due to a manufacturing problem or an air leak, is so marginal that nearly all of the unwanted atoms are swept from about half the length of the bore. However, the other HeNe tube I have like this had the color change in the exact opposite direct - correct at the cathode but blue-ish-pink at the anode, also reduced power. I now suspect that it may have been internal contamination. More research is needed. :)

    Another unusual characteristic of the Northern Lights tube was that the output power (what of it there is) peaked at a current somewhat higher than expected (8 mA as opposed to the 6 or 6.5 mA typical of this size tube). I don't know whether this is simply due to the overall contamination or that more of the nasty unwanted ions being swept from the bore when running at a higher current.

    This tube had an unfired getter which provided a means of cleaning up the contamination without a refill. A few weeks later, I got around to making the attempt. And the results are.... See the section: Repairing the Northern Lights Tube.

    Starting Problems and Hard-to-Start Tubes

    Some tubes seem to practically start on their own. Other won't perform even when you stand on your head, hold your breath, and provide the proper chants and sacrifices. :-) Note that this may not be a problem under your control. Apparently, even the Ph.D. physics types at major laser companies may not understand why apparently identical HeNe tubes from the same production run may vary in their ability to start by a large amount. See the section: About Hard Start HeNe Tubes.

    However, it could also be that your power supply operating voltage, ballast resistor, and other factors may need modification. Of course, if the system used to work reliably and suddenly died, an actual power supply or wiring problem is most likely though a dead HeNe tube is also possible especially if the system has been unused for several years. But don't overlook the unlikely, but not impossible situation where your line voltage is low for some reason! Check it first. The discussion below is somewhat oriented to the situation where a HeNe tube or laser head is being assembled with a power supply (or parts have been replaced) and the combination just doesn't want to work properly. However, some of it also applies to actual failures as well. Where the power supply itself is suspect, see the section: Power Supply Measurements, Testing, Repair.

    There are several types of possible behavior depending on how well the power supply, ballast resistance, and HeNe tube are matched up, and if any of these as well as the wiring, are faulty. You first need to determine if the discharge is being initiated at all. If the starting voltage is adequate, there will be momentary flashes that may be extremely short and weak and only visible in a darkened room but operating current may not actually follow. Under marginal conditions, operating current will flow in response to the starting voltage but won't be maintained. These flashes will be brighter and longer in duration. The result may be a nice flashing laser. In fact, this progression is exactly what will be seen when operating a HeNe laser tube from a power supply on a Variac as the voltage is increased: Short flashes followed by longer flashes and at some point, a steady beam.

    WARNING: If your HeNe tube doesn't start after a reasonable length of time (like a minute), don't leave the power supply on overnight in a futile attempt to get it going. Starting is a stressful time for power supply components, especially some wide compliance designs, and an extended period with the very high starting voltage on parts of the circuitry may result in total failure. It could also result in electrical breakdown (arcing) inside the laser head or cable. If the laser is flashing, this may be ultimately bad for the tube as well. Turn it off, step back, and try to determine what is wrong.

    Where the power supply components and/or wiring is exposed and subject to dirt and grime, first, carefully clean everything to eliminate possible sources of electrical leakage, which can affect operation, particularly the very low current starting circuit. As an experiment, try warming up the unit (which drives off conductive moisture) with a hair dryer or heat gun on the 'low heat' setting. This may enable it to start more easily confirming the need for some housekeeping. :)

    First, vacuum and/or dust it off with a soft brush, then use mild detergent and water followed by isopropyl alcohol (rubbing or medicinal is fine as long as there are no additives). Give it ample time to dry completely. The hair dryer or heat gun can be used to help it along. You may now find that your starting problems have disappeared!

    If your tube or head has an external starting loop or tape (see the section: Power Requirements for HeNe Lasers), it must be cleaned thoroughly as well (or maybe it has become disconnected, is broken, or has shorted to the case!).

    There is also a possibility that something else is shorting out the power supply, possibly only when enough voltage is applied so it won't show up with an ohmmeter test. Sometimes, the ballast resistor inside cylindrical laser heads will arc to the case. This can be checked with an HV insulation tester or more easily for most people, by removing the end-cap(s) and visually inspecting (as well as smelling!) for evidence of arcing, or by disconnecting the anode wire and driving the tube directly from the power supply with an external ballast resistor.

    Assuming none of this helps, there are three types of behavior: (1) No action of any kind, (2) an occasional flash possibly at random intervals, and (3) a periodic flashing laser which never settles down to normal steady operation. However, the behaviors and their causes are not really always independent so read through all of the possibilities before replacing components or ripping your system apart!

    1. Tube does not fire at all. This means that the tube itself appears totally dead with no flashes inside and no evidence of a beam, even for an instant.

      This generally means that the starting voltage is inadequate for the tube or isn't reaching it, there are other circuit problems, or the tube is bad. In rare cases, shining a light into the tube will allow it to start. Tubes with longer and narrower bores (capillaries) will generally require greater starting voltage and your power supply may just not be up to the task. While tube manufacturers generally specify a starting voltage of 7 to 10 kV (or higher), typical tubes will fire with 3 to 5 times their operating voltage. Thus, a tube that runs on 1,700 VDC will probably start on 5,400 to 8,500 VDC.

      • There may be too much leakage in the anode circuit preventing the buildup of adequate starting voltage. The problem may be in the power supply itself or in the wiring to the HeNe tube. Corona discharge or arcing can result from inadequate insulation or component spacing as well as sharp points in the wiring or connections. Dirt and grime may also reduce the insulation resistance. A sizzling and/or ticking sound along with the aroma of ozone are indications of this sort of HV breakdown. Highly humid conditions may make the situation worse. For pulse (trigger) type starters, there may be too much capacitance as well.

        In the case of an enclosed laser head with a HV (e.g, Alden) connector, HV cable, and internal (potted) ballast resistor, there may be a breakdown in one of these components and it may only show up when starting voltage is applied (not with an ohmmeter).

        Allow the laser to attempt to start for 15 or 20 seconds and turn off the power supply. Immediately pull the Alden connector out of the power supply and discharge it on a metal surface. A nice long (e.g., 1/4") spark indicates that the starting voltage is probably adequate from the power supply and breakdown in the wiring is not likely. If there is little or no spark, either the starting voltage is low or zero, or there is a broken connection between the connector and the tube resulting in not much capacitance to store a charge.

        Here are two more tests for this situation:

        • Remove the negative connection from the ballast resistor assembly entirely - float it so the starting voltage cannot arc to anything. Connect the negative directly to the cathode of the HeNe tube or laser head case as appropriate.

        • Disconnect the anode of the HeNe tube and substitute your own ballast resistor and wiring.

        If the tube now starts, one of the original components was faulty (most likely the potted ballast resistor assembly if the negative connection runs through it) and this will need to be replaced.

      • There may be an open ballast resistor or break in the wiring inside an enclosed laser head. Check with an ohmmeter (with power off and the tube capacitance discharged!) all the way from the power supply or Alden connector to the tube terminals or mirror mounts (as appropriate). Don't forget to check the negative (return) connection and cathode ballast resistor (if present). Some laser heads have the wires simply clipped on or use little push-on connectors to attach to the tube mirror mounts, exhaust tip-off, or through-glass pins, and these have been known to pop off in shipping.

      • If you need to increase the input to start or obtain any sort of response but then must back it off substantially to reduce the tube current to the proper value, low starting voltage or one of the other related problems is indicated.

        Assuming the power supply and wiring check out and the tube is good, the only solution is to boost the starting voltage or use a different type of starting circuit (inverter instead of voltage multiplier, for example).

      • You may have an extremely hard-to-start tube. Whether this is just normal for your particular model tube or this particular sample, is due to it being old or unused for a long time, it is just tired with many hours under its belt, or some other problem, the result is that the specified starting voltage does not have any effect.

        Note that newly manufactured tubes requiring more than a second or so to start using a compatible power supply are usually rejected as defective and may end up in the hands of surplus dealers who may sell them as 'new' even though they don't meet specs. Thus, you may be more likely to end up with one of these hard starting tubes!

        • If a particular tube doesn't start right away because the tube itself is hard starting, cycle the power supply on and off a few times at about 1 second intervals. If it's going to start, it usually will by the third or forth try. I suppose the rapid rate of rise of the starting voltage does something useful. :) (How well this works will depend on the actual tube and power supply but most power supplies should survive the abuse.) While probably not a long term solution, if the tube starts, at least you know it isn't dead.

        • If the voltage doesn't discharge in a reasonable amount of time after powering down due to a lack of bleeder resistor, it may help to discharge the capacitance of the tube and power supply manually with a well insulated high value resistor (a few hundred k ohms to a few M ohms) before attempting to restart. Again, the rapid rate of rise of the starting voltage helps to ionize the gas. Where the tube starts reliably when this is done, it may be possible to add a high value (e.g., several hundred M ohm) resistor permanently across the power supply output.

        • For HeNe laser tubes with exposed bores like Spectra-Physics side-arm tubes, touching various glass parts of the tube (NOT the high voltage connections!) might also provide a capacitive path for the starting pulse. This applies more to pulse starters or those with s substantial component of high voltage AC, not the typical voltage multiplier with a DC output. Older Spectra-Physics exciters tend to have up to 50 percent of the starting voltage in the form of an AC waveform. Double check that the laser head case is grounded - a missing ground may result in starting problems. (More on this below.)

        • Shine a light on the bore in various places to see if that has any effect. On high mileage (probably) tubes, as with high mileage neon lamps or fluorescent lamp starters, a few photons can help ionize the gas. Even shining a flashlight into the front of the laser may do the trick. This isn't always effective but appears to be more common than I had though and is easy to check.

          And, it may not only be high mileage tubes. I recently discombobulated (translation: disassembled to harvest its organs) a vintage HeNe laser-based LaserDisc player and found a little incandescent lamp buried near the bore of the laser tube. It is even documented in the service manual (which includes the assembly procedure for the optical pickup or "slider" as Pioneer calls it), but there is no discussion as to its purpose. However, the way it's wired in suggests that the lamp is not in the original design. The only possible explanation is that it was there to help a possibly hard starting HeNe laser to start. It must have been included to be able to use tubes that otherwise would fail to start quickly enough. Nonetheless, for a manufacturer include a light bulb to aid starting in production units is so strange. I did have a professor whose motto was "If it works, use it". :) That's also my motto when dealing with slow starting HP/Agilent metrology lasers like the 5501B, 5517B, and others, which have a similar quirk: With the cover off, they would start normally, or at least fairly quickly. But with the cover on, they could take a minute or more to start. I now routinely add a high brightness LED shining on the back-end of the laser tube if the starting time is so excessive that it might be flagged as an error.

        • Another test is to try the tube with reverse polarity on its input. Connect the positive output of the power supply to the ballast resistor (don't omit this!) and then to the cathode (can electrode) end of the HeNe tube. Connect the negative of the power supply to the anode of the tube. You are only doing this for testing! Do not be tempted to leave the tube wired this way permanently should it actually start. And definitely no more than an instant for yellow or green HeNe laser tube.

          Based on tubes I have tested, the starting voltage is much lower with the anode and cathode connections interchanged. However, the voltage drop across the tube when running with reverse polarity is much higher than with correct polarity. Thus, the tube may not run within the normal operating voltage range of your power supply even if the discharge is initiated - more likely it will just pulse.

          Nonetheless, even if it just pulses, at least you know the tube is not totally dead. If the tube is otherwise undamaged, there should also be an indication of (at least weak) laser output from the business end of the tube. Perhaps, all you need is a power supply with higher starting and/or operating voltage. An inverter type starter using a flyback transformer appears to be particularly good for hard-to-start tubes. Unfortunately, I do not know of any reliable way of determining the likelihood of success without actually trying it.

          I have one 5 mW HeNe tube that requires (depending on its mood) as much as 15 to 20 kV to start (it should be less than about 10 kV). However, once started, it runs with a normal operating voltage of about 1,800 VDC.

          WARNING: Do not let the HeNe tube run for any length of time with reverse polarity as damage may occur due to heating and sputtering at the anode end of the tube.

        • Carefully heating the tube (with a hair dryer, NOT a propane torch!) may help to get it started in some cases. I've only heard of one instance of this and have not tried it myself but it should be easy to experiment. A tube that responds to this treatment is probably one what just has trouble starting the first time after being off for awhile but restarts easily when warm. However, what may really be going on is that you are heating the power supply circuit board nearby reducing leakage in the anode circuit (see above) or bringing a ground in close proximity to the tube (see below).

    2. Tube flashes occasionally for the very briefest of instant, possibly at random intervals. There may or may not be a laser beam accompanying the flashes. Depending on the situation, it may just appear to give up and revert to (1), above. This may be a protective feature of certain power supplies or an indication that the starting pulse is being diverted by an electrical leakage path, corona, or arcing somewhere.

      This sort of behavior is probably more likely with a pulse type starter but can occur with other types as well. What is likely happening is that the energy is insufficient to fully ionize the gas inside the bore of the HeNe tube so the discharge doesn't 'catch'.

      In addition to the other possibilities listed above and below:

      • Check for missing grounds. Where a pulse starter is used or a marginally compatible power supply where the high rate of rise of the starting voltage is needed to start it is possible that like many fluorescent lamp fixtures, a grounded metal plate is needed near or in contact with the tube to provide a capacitive path to aid initial ionization of the gas. Operating the laser (or fixture) on an ungrounded outlet, not installing a foil shield or not properly grounding it, or replacing the HeNe tube with one that doesn't have such a 'feature', might result in erratic starting - perhaps no action at all or maybe an occasional flash but not actually remaining on or taking a long and random time to 'catch'. If putting your hand on the tube (carefully avoiding the high voltage terminals!) results in consistent starting, lack of proper grounding would be confirmed. (This applies to uncooperative fluorescent fixtures as well.) Add a wrap of aluminum or copper foil and attach it to earth ground. I have only heard of one case that might have been due to a missing ground so I suspect this is quite an unlikely scenario!

      • Make sure the anode leads are as short as possible to minimize stray capacitance.

    3. Tube flashes momentarily - longer than an instant possibly up to a second or two - but does not 'catch'. During the time the discharge is on, there is a laser beam.

      What happens is that the discharge is initiated but the voltage drops too much at the tube anode and the discharge goes out. This cycle repeats resulting in a flashing HeNe laser.

      To produce a stable discharge, the following must be satisfied:

      • The sum of the effective resistance of the power supply and the ballast resistor and the (incremental) negative resistance of the tube (dV/dI at the operating point) must be greater than 0.

      • The voltage across the tube must be above the minimum for the tube at the operating current.

      • The current must be above the minimum for the tube/power supply/ballast resistor combination.

      These factors are not independent. Since the negative resistance and sustaining voltage of the tube are not normally specified and depend on current, some amount of trial and error may be required to achieve consistent stable operation but in most cases it really is very easy.

      Cycling behavior can be due to several factors:

      • Poor power supply voltage regulation or excessive ripple. Until the tube fires, there is essentially no load on the supply resulting in much greater voltage than under load. Except for a high compliance type of design where this is needed to produce the starting voltage, minimizing this difference will improve stability and reduce the voltage needed for stable operation.

        If the transformer or inverter drops too much under load, the tube voltage may fall below the minimum for the tube/ballast combination as soon as it starts. This cycle will repeat continuously or it occasionally may catch.

        Use a higher voltage and larger ballast resistor, and/or increase the uF value of the main filter capacitor (and/or the one in the DC supply to an inverter type supply as well if it isn't regulated).

        Minimum capacitor values for less than 5 percent voltage ripple (typical voltage and current requirements):

        • Line operated supplies: .5 to 1 uF (2000 V, 5 mA).
        • Inverter output: .005 to .01 uF (10 kHz, 1,800 V, 4 mA).
        • Unregulated inverter input: 15,000 to 20,000 uF (12 V, 1 A).

        Actual ripple in the current to the tube may be several times greater than this since it depends on the change in voltage with respect to the total effective resistance of the PS+tube+ballast resistor combination). However, the resulting ripple in the optical output power will be 2 to 10 times lower than the ripple in the current depending on operating point. The lowest will occur around the tube's optimal current specification.

      • Ballast resistor too large for the operating voltage. The operating current falls too low resulting in increased (magnitude of) negative resistance. Once the total system resistance goes negative, the discharge becomes unstable and goes out. The result is a flashing laser like a neon bulb relaxation oscillator.

        For an unregulated power supply, increase the operating voltage and/or decrease the ballast resistance.

        For a regulated power supply, decrease the ballast resistance so that the voltage for the desired operating current falls within its compliance range.

      • Too much stray capacitance and/or inductance in anode circuit. The system is behaving like a relaxation oscillator as the capacitance charges and then discharges through the tube. The wiring inductance causes the current from the main supply to lag too far behind the starting current and the discharge goes out.

        Shorten the wiring - minimize the distance between the power supply and ballast resistor, the ballast resistor, and tube anode, and don't use long runs of high voltage coax (which may have higher capacitance). Increasing the energy of the starting circuit slightly may help as well.

      • For laser heads in particular, the additional capacitance resulting from the metal case may increase the minimum stable tube current by up to 1 mA or more - and thus require changes in the power supply and/or ballast resistor. So, if you tested the HeNe tube and power supply on your workbench but the enclosed system is unstable, this may be the reason.

      • Power supply polarity is reversed. The voltage drop across a HeNe tube operated with the cathode and anode interchanged is higher than under normal conditions. However, required starting voltage is much lower. The result is likely to be a pulsing laser. Double check your wiring and terminal connections. I have also seen commercial power supplies mislabeled! See the section Making Measurements on HeNe Laser Power supplies if you need to actually test for reverse polarity.

    Also see the sections: How Can I Tell if My Tube is Good?, About Hard Start HeNe Tubes, Testing a HeNe Laser Power Supply, Power Supply Construction Considerations, and Adding a Start Wire.

    About Hard Start HeNe Tubes

    As noted elsewhere, apparently identical HeNe tubes made on the same production line may differ widely in their starting performance. While one sample may start absolutely instantly, another one that is indistinguishable may take several minutes to wink on. A newly manufactured HeNe tube that requires more than a second or so to start using a compatible power supply is generally rejected as defective. Guess who may end up with these - surplus dealers, and ultimately, you! The condition may get worse with use so some high mileage tubes could indeed be a lot worse in this department - possibly to the point of being virtually impossible to start even if they were within spec when new.

    As far as I can determine, the fundamental physics behind this phenomenon may not even be well understood by the major laser companies. The only meaningful data is statistical, because even a give tube with a given power supply will have dramatically different start times from attempt to attempt, as will tubes built side-by-side through the entire production process.

    Tubes that are kept in dark cold environments for long periods of time don't tend to start well. But, once one of these tubes is started successfully, restarts will likely be instantaneous, or at least reasonably quick. However, left overnight, they will revert to being uncooperative.

    Also lower fill pressures and cleaner tubes make for hard starting - not to mention power supply variables.

    Some manufacturers (e.g., Melles Griot) use a conductive 'start-tape' running the length of the tube attached to the anode electrode to aid in starting. It's not even really proven that this improves performance (and I've found that it can be a source of electrical breakdown problems. I've never noticed any difference in the speed of starting after removing the start-tape). Uniphase had a pointed electrode inside the anode mirror sleeve to aid in starting but it isn't obvious that it made any statistical difference either. There has even been talk of using a trace of radioactive gas (as used to be common in neon indicator lamps and glow tube fluorescent starters), but this of course would probably not be a popular idea today!

    A given production line may still have hard-start related yield problems from time to time (which kind of suggests the Ph.D physicists don't understand it). Funny thing is, no one can tell anything that's different on a hard-starter versus a regular one.

    Adding a Start Wire

    As noted elsewhere, many HeNe laser heads have some type of starting wire or start tape running from the anode along the tube to the vicinity of the cathode. In most cases, this actually has little or no effect on starting, and may end up becoming a problem where electrical leakage can cause starting problems. The reason it doesn't usually help is that with the common coaxial tube construction with a full length gas reservoir, the electric field produced by the high dV/dt of the initial starting voltage rise can't be anywhere near the bore, where it's needed. This is particularly bad with Melles Griot and Aerotech tube construction since the discharge path is not near the glass surface anywhere. On Hughes and NEC tubes, there is a location near the cathode where the center of the tube is exposed.

    For some hard start tubes that otherwise run well without current drop out problems, adding a start wire directly from the anode with a few turns wrapped around the tube near the cathode may help starting dramatically. For example, on Spectra-Physics side-arm laser tubes, there is a section next to where the cathode attaches to the main bore where adding a start wire may work quite well. This allowed an SP-155 which would almost never start at normal line voltage to start instantly first time every time.

    Use a well insulated wire connected before the ballast resistors. Wrap the end a few times around the narrowest section of the tube near the cathode or use a metal clip. Some experimentation may be required. Just try not to zap yourself excessively during testing. :)

    However, if used with a some linear power supplies having a voltage multiplier starter (probably without a final diode/filter stage), the tube must run stably substantially above its lower current dropout point or else the start wire will tend to turn the tube off as well as turn it on and the result will be a flashing laser, which is usually not a good thing. Adding a HV diode in series with the start wire might help, but I haven't tried it.

    As added protection for you and the power supply, instead of connecting the start wire directly to the raw high voltage output, install a series capacitor with bleeder. I would suggest 1 nF with a 100M ohm resistor across it. Both should be rated for the peak starting voltage available from the power supply, typically 8 to 15 kV.

    Possible Causes of No Output with a Normal Discharge Glow

    Where the HeNe tube starts - there is a stable glow discharge and it is the correct color (bright white-ish red-orange, see the section: How Can I Tell if My Tube is Good?), there are a few possibilities that are not due to a bad HeNe tube or laser head:

    And, for other-color HeNe tubes which have much lower gain for a given length than red HeNes, all of the above may apply. The following comments were prompted by questions about a non-lasing short green HeNe tube (similar to a Melles Griot 05-LGR-024, 215 mm in length:

    (From: Lynn Strickland (stricks760@earthlink.net).)

    Those things are touchy, touchy, little SOBs. They usually have an almost flat HR and OC combination. If it does lase, it will probably be a few tenths of a mW at best. Probably have to walk the beam AND tweak both ends for any hope. Try some magnets too, for 3.39 micron suppression. In general, low power greens are a bitch to tweak.

    Note that the green discharge is more 'pink' (red tubes more 'orange'). Fill mixture is a little different, but the different color mostly due to lower fill pressure - which is why greens have shorter lifetimes than red.

    Possible Causes of Low Output with a Normal Discharge Glow

    Where the discharge color is normal and the tube current is known to be reasonably correct but the output is weak, there can be a variety of causes. (Of course, most of this also applies to no output but this is covered in the section: HeNe Tube Problems and Testing.) The following applies when the output power just isn't as great as expected:

    Doughnut Mode Beam from TEM00 HeNe Laser

    Occasionally, a HeNe (or other) laser designated as TEM00 will actually output a beam which looks like a doughnut or torus - with a hole in the middle. It's running with a not quite TEM00 mode structure - a doughnut mode. Since this should be a TEM00 laser with a single transverse mode Gaussian beam profile, that's normally considered a bug. If it's outputting decent power (compared to its specifications), the cause isn't likely to be dirt inside the tube or imperfect mirror alignment.

    More likely, the manufacturer accidentally used too large a bore for the length of the resonator and the mirror curvature. For example, if this is a green (543.5 nm) HeNe laser, they may have used a bore sized for a red (632.8 nm) HeNe laser by mistake resulting in a mode diameter that is too large. Or, it might have been designed on the hairy edge, size-wise, in an attempt to get as much power as possible out of the tube and the engineers weren't lucky that day. With a correctly sized bore, slightly incorrect mirror alignment would result in lower power but maintain a TEM00 beam profile.

    If you had been the original owner, the laser might have been replaced under warranty. As it is, you now have what I generally call an "interesting" laser. :) Or, since the specifications are often only with respect to "95% mode purity", if the hole represents less than 5 percent of the power, maybe it's considered acceptable, though I can't imagine anyone being entirely happy with a laser that's supposed to be TEM00 having a hole in the middle of the beam unless all they care about is the number of photons per second!

    However, even with too wide a bore, it may be possible to adjust the mirror alignment to obtain a TEM00 beam. Gain access to the cathode-end mirror mount (to avoid a shocking experience) and allow the laser to warm up completely in the orientation that will be used. Aim the laser at a screen of some sort like a white business card, and gently press sideways on the mirror mount from orientations every 45 degrees or so (top, bottom, left, right, and the diagonals). Closely examine the spot shape to see if it pops into TEM00 at any time while still maintaining near-maximum power. If it does, it should be possible to adjust the alignment as described in the sections starting with: Problems with Mirror Alignment. A power meter will be useful to assure that the output power is still near maximum. It may in fact end up being maximum with the TEM00 beam, though some other beam profile when cold. Very likely, the alignment will be very critical. Several attempts with multiple warmup cycles (to allow the metal mirror mount stems to settle in) may be needed to achieve consistent behavior. I adjusted a green HeNe laser head that formerly had a doughnut beam. It now starts out TEM01 but becomes a beautiful TEM00 once it warms up, and significantly exceeds spec'd power. But I haven't decided whether a consistently interesting doughnut beam or a boring high quality TEM00 beam only when warmed up is preferable. :) CAUTION: You could make the laser worse or dead by attempting mirror alignment. So unless you've done this successfully before, it may be best to leave well enough alone and enjoy the unusual behavior. There's nothing else you can do about it!

    Unstable or Flickering HeNe Tube

    Where your HeNe tube starts and lases normally but is unstable, flickering or going out and then restarting whenever the power line voltage dips slightly or for no apparent reason, the problem may be power supply or HeNe tube related. It is also not unusual for this to start happening after the system has been on for awhile due to characteristics of one or more of the components changing slightly. The following are not mutually exclusive:

    A different power supply or slight adjustments or modifications may make your HeNe tube happy, at least temporarily. However, where the HeNe tube is an inexpensive vanilla flavored variety, replacement may be the easiest solution if it turns out to be marginal. :-)

    Sputtering or Erratic HeNe Tube

    This behavior is somewhat similar to the some of the compatibility problems described above. However, it could be much more random and may only occur during warmup.

    The symptoms are that the tube may start normally but then go off and restart, possibly quickly and unpredictably. One possible cause is a bad internal connection between the cathode can and its attachment to the mirror mount where the negative lead of the power supply is hooked up. The type of construction susceptible to this malady is where a 'nipple' on the end of the aluminum cathode can is swaged (pressed/squished) into the mirror mount rather than actually being attached by spot welding or via a spring contact. After many thermal cycles, the swage can loosen resulting in intermittent contact especially as the tube heats and parts expand. (This is sort of the same problem that aluminum house wiring can have if improper termination techniques or devices are used, but in that case, the consequences can be much more disastrous!) Any sort of high resistance increases the required tube voltage since the mirror mount has a higher 'cathode fall' voltage drop. The discharge will likely go out and the power supply will then attempt to restart. In some cases, the discharge may strike to the mirror mount itself (look for a glow near the mirror) and if this persists, will eventually destroy the mirror. (See the section: Damage to Mirror Coatings of Internal Mirror Laser Tubes) After the tube warms up sufficiently, since aluminum expands faster than steel or Kovar, the problem may disappear once the connection tightens. However, until then, the intermittent contact and many restarts is hard on the power supply and nearby mirror.

    Assuming the power supply and tube are properly matched and the power supply isn't defective, this is a defective HeNe tube. No cure is possible. This is a relatively unusual problem (I've only seen it in two (2) HeNe laser tubes so far) so first check external connections and make sure your HeNe tube and power supply are properly matched. If its maximum voltage is marginal, as the tube heats up, the voltage drop may increase just enough to result in erratic behavior. However, one possible difference between this and a bad cathode connection is that with the latter, the condition may clear up once the tube heats up since the expansion of the aluminum cathode will improve contact. The marginal voltage situation will just get worse. The power supply itself could also be defective. The easiest way to determine which is at fault is to swap the PSU and/or tube with known good units.

    I've seen this malady on a few older Melles Griot HeNe lasers. On newer ones, a thin metal conductor has been added - presumably in response to this type of failure - connecting the spider/bore support (which is in good contact with the cathode can) directly to the end-cap. It is spot welded at both ends.

    Also see the section: Unstable or Flickering HeNe Tube.

    Alden Connector Problems and Repair

    While Alden connectors rarely give problems, a loose or corroded contact inside the plastic shell can result in all sorts of intermittent behavior including hard starting and repeated restarts or flickering, as well as possible power supply failure. Though rare, arcing is also possible from a damaged cable, generally where a HV coax enters the Alden shell. Any heating of the Alden connector - even if NOT accompanied by 6 foot flames - is an indication of trouble with the contacts as there should be absolutely no detectable power dissipation in the connector! :) So, before you blame the laser head or power supply, make sure the Alden connections are secure.

    The male pins can be cleaned with a file or fine sandpaper. The female contacts can be tightened by wedging a small flat-blade screwdriver inside the connector between the side and the plastic contact support. Entire replacement Alden connectors are readily available or can be salvaged from dead laser heads or power supplies. To avoid the possibility of arcing or a shocking experience, use at least 2 layers of heat-shrink tubing for insulation with a minimum of 3/4" beyond the bare wire sections, and stagger the splices. Add another layer of heat-shrink over the completed splices.

    And replacements are available (other than from salvaged lasers!). See Alden Connectors Catalog.

    Cyclical Variations in Output Beam Power

    A HeNe laser that is in good condition will produce an output beam that is quite stable and will have no visible (at least by eye) variations in output power though a laser power meter will show fluctuations over various time scales of a few percent even after warmup unless it is high performance amplitude stabilized laser. A typical laser tube will have a "mode sweep" specification of between 2 and 20 percent depending on type and size (smaller tubes typically have poorer performance). As the tube warms up, this cycling will start out slowly, speed up and peak, and then taper off as the system stabilizes thermally.

    Note that if the discharge is actually going on and off, the cause is entirely different - an incompatibility with the power supply, incorrect ballast resistor, low line voltage, etc. See the section: Unstable or Flickering HeNe Tube.

    However, sometimes you will find a laser that exhibits significant periodic variations in output intensity even where the discharge is perfectly stable. There are two types of phenomena depending on the period of the power cycles:

    Reasons for Short HeNe Tube Lifetime

    As noted, sealed HeNe should last many thousands of hours with no noticeable degradation in performance.

    If you are experiencing excessively short life (e.g., a month instead of years), the first things to check are operating current and polarity. See the section: Making Measurements on HeNe Laser Power Supplies. Of course, if you omitted the ballast resistor, life will likely be very short. :-(

    If the HeNe tube and power supply are mismatched, one can damage the other. For example, running a 1 mW HeNe tube on a power supply designed for a 35 mW HeNe tube may not only result in too high a current by design (e.g., 8 mA instead of 3 mA) but may also result in much higher current if the compliance range of the power supply is exceeded (i.e., the voltage across the HeNe tube is much lower than the power supply can handle). Conversely, attempting to power a 5 mW HeNe tube using the power supply from a barcode scanner (designed for a .5 to 1 mW HeNe tube) will likely result in a blown power supply. Just because the high voltage connectors mate and/or the tube lights up doesn't imply anything about compatibility! Also note that maximum optical output occurs at the optimum operating current - too high or too low and it goes down. (Operating current for yellow, orange, and green HeNe tubes is even more critical than for the common red variety so setting these up with an adjustable power supply or adjusting the ballast resistance for maximum output is recommended.)

    New and even used HeNe tubes and power supplies from reputable surplus dealers will generally last a long time if not abused. But, much of what you get at swap meets and hamfests has been pulled from equipment for one reason or another. So, the problems you are experiencing may have nothing to do with your setup!

    (From: Lynn Strickland (stricks760@earthlink.net).)

    Speaking as a non-physicist....

    There are so many variables in a gas discharge, it's a game of averages. That's why the power supply business can be so tricky - and why, for the power supplies you can look inside of, you see so many modifications. That, and the rate at which electronic components go obsolete keeps it in a continuous state of flux (no pun intended).

    Reasons for the variability in lifetime and failure mechanisms from design to design revolve around design fill pressure and gas mix, operating current, distance from capillary bore-end to cathode, optical design (some designs are more sensitive to misalignment than others). Also power supply variability, ballast resistor value differences, operating current tolerances (often set at, say, +/-0.2 mA).

    Gas lasers can be a pain, but for a lot of applications, they're still the most cost effective solution -- in some cases the only solution.

    Power Variations Due to Lack of Mirror Substrate Wedge

    The only surfaces of a HeNe laser cavity that should be aligned are those of the two mirrors. Any others that are perpendicular to the optic axis will cause reflections that can affect the power and stability of the laser by forming a Fabry-Perot etalon - a secondary resonant cavity. (Note that the following discussion does *not* deal with the stability aspects of reflections - which is a very complex subject unto itself - but only with how they affect the waste beam and main beam output power.) Perhaps surprisingly, the outer surfaces of the mirror glass or substrate are a possible source of reflections that can cause peculiarities, if not actual problems, in the laser's power stability. Both the HR and OC mirrors can be at fault and affect the power from the main beam out the front of the laser as well as the waste beam out the back. But it is power variations of the waste beam that are usually most dramatic, except for some cases of goofups in manufacture, mostly of "other color" lasers..

    So, here are some tests to determine the source of the ripples:

    Of course, looking for ghost beams or the "dab of Epoxy" (or index matching fluid) test will be conclusive. And, it's always possible that there are problems at both ends of the laser!

    The only conclusion here can be that while it's easy to reduce the power variation due to a non-wedged HR significantly (even a smudge or fingerprint on the glass will do a fair job!), built-in wedge with absolutely no parallel surfaces to reflect directly back into the tube is really essential for both the HR and OC mirrors to achieve perfection. Index matching cement will still leave some reflection at the boundary and even 0.01 percent - which would be very good - will still result in a 4 percent variation in output power! That such a low reflectivity can produce this much of an effect is quite counter-intuitive. :) There is no easy solution for the OC mirror at all.

    One possibility that will work for both mirrors is regulated temperature control of the mirror itself. This is used in some Laboratory for Science stabilized HeNe lasers which include active circuitry in a little widget clamped to the OC mirror to maintain its temperature such that the etalon transmission is maximum. But simply temperature controlling the tube isn't adequate since even if locked to a particular mode, there is no guarantee that the mirrors themselves will be maintained at a constant temperature.

    The variation in output power for either HR or OC etalon effects is small and might go unnoticed for most applications. The variation in waste beam power, though typically a much greater percentage of the average waste beam power, is even more likely to go unnoticed since it's, well, usually wasted. Whatever waste beam peculiarities may be present with one sample of any model tube doesn't necessarily mean they all will behave similarly since what happens at the back of the tube or even the slight output beam ripples would not impact the laser's important specifications.

    But where the waste beam is used for something like implementing a stabilized HeNe laser, a laser tube without this malady should be selected if possible. Adding an external wedged optic to the HR is also a possibility but unless the index matching of the glue is essentially perfect, there will always be some residual reflection and even 0.001 percent will still result in more than a 1 percent in waste beam power variation.

    For a particularly interesting case study where this did matter, see the section Melles Griot Yellow Laser Head With Variable Output.

    HeNe Tube Use and Life Expectancy

    You often hear that lasers like to be run to keep them healthy and maximize life.

    For both types of HeNe tubes (as well as other lasers), power and beam quality will peak only after some warmup period. So it makes sense to keep the laser energized continuously over the course of an application where these are critical but this has no bearing on any need to turn the laser on just to keep it healthy.

    Here is a chart of very rough guidelines for evaluating HeNe lasers. This is based solely on my observations with only minimal input from those who should know about this sort of stuff like major laser companies:

     Characteristic            New      Middle Age    High Mileage    End-of-Life
    --------------------------------------------------------------------=----------
     Starting (1)              Easy        Easy       Easy to hard     Very hard
     Operating voltage (2)     Spec        +5%            +10%        +25% or More
     Dropout current (3)       Low         Low           Medium           High
     Output power (4)         1.5-3X     1.25-2X         1-1.5X          0-0.5X
     Discharge color (5)      Normal      Normal        More Pink      Pink-White
     Brown crud in bore (6)    None        Some           Much            Much
    

    1. Most tubes will start easily, at well below the spec'd maximum starting voltage. However, there are exceptions, which hopefully end up in the dumspter, or more likely, the surplus market. :) Starting may become more difficult as the tube is used, though not always. But, at end-of-life, it could be almost impossible.
    2. When new, the tube operating voltage will be close to the spec'd value. As the tube is run, it will increase gradually until end-of-life where it may become dramatically higher.

    3. Dropout current on a new tube should be well below the operating current but will gradually increase with use until it may reach the operating current. Then, the tube will start but not remain on, flashing or flickering. Increasing the ballast resistance with a good portion of it very close to the tube anode may reduce the dropout current substantially on a healthy tube, but will have little effect on a high mileage one. For example, the following data were measured for a healthy Zygo 2 mW tube:

           Ballast     Dropout
          Resistance   Current
        ------------------------
             68K       4.75 mA
             80K       4.5  mA
             90K       4.25 mA
            107K       3.75 mA
            134K       3.25 mA
            161K       3.0  mA
            188K       2.75 mA
      

      For this tube, adding about 36K ohms very close to the anode enabled the laser to be stable at the fixed power supply current of 4.5 mA allowing it to be retrofitted into a Spectra-Physics laser head. With the original 75K ohm ballast, this was marginal and the discharge became unstable after warmup. Of course, the operating voltage increased by about 160 V, so the power supply needed to be capable of dealing with that.

    4. Output power on average may actually increase during the early stages of the tube's life due to initial overfill, but will then decrease gradually until end-of-life, where it will nose dive. Output power may also change as a result of slight mirror misalignment caused by multiple thermal cycles but this is fully correctable.

    5. With use, the gas pressure inside the tube decreases resulting in a trend toward becoming more pink. But at end-of-life, sputtering results in a tendency to become more white or purple. By then, output power may have droped by 95 percent, if there is any output at all.

    6. Brown crud develops with use, though not all tubes will have any. Where present, in many cases there will be very distinct striations in the brown crud, indicating plasma oscillations, but this appearance alone may not imply much of anything about the tube's health.

    See the section: Case Study of 145 Melles Griot 05-LHR-640 HeNe Laser Tubes for an example of the behavior of one specific model. I do not know to what extent these data apply to other HeNe lasers.

    HeNe Laser Tube Gas Pressure and End-Of-Life

    The most significant mechanism is causing aging in a HeNe laser tube run under normal operating conditions (rated current, reasonable number of starting cycles), is cathode sputtering which traps gas molecules under/in the metal film and gradually reduces the gas pressure inside the tube. Since sputtering is more severe at low pressure, beyond a certain point, this process accelerates leading to a rapid deterioiration in performance near end-of-life. The "pickling" of the aluminum cathode surface to reduce its work function can only do so much.

    In addition to the discharge color becoming less intense and pinker, the increase in operating voltage, and eventually, the decline in output power, at least three other measurable values can be indicators of low pressure:

    1. Output power versus tube current: At the normal pressure at which most HeNe laser tubes are filled, the output power peaks at the nominal operating current, and then declines with increasing current. However, as gas pressure decreases, the current at which the peak occurs will increase, and eventually output power will simply increase with increasing tube current.

      However, there are some strange situations where mirror alignment can appear to affect the nominal operating current. And, wide bore multimode HeNe lasers may have a recommended operating current which is below the current for maximum power to extend tube life.

    2. Dropout current: This is the lowest current where the discharge will be stable, regardless of the ballast resistance and wiring/tube capacitance. For new red (632.8 nm) HeNe tubes, it's usually lower than 75 percent of the nominal current, with some down below 50 percent. However, as the gas pressure decreases, the dropout current goes up. When it reaches the nominal operating current, the tube may be unusable even if the output power is still decent. Increasing the operating current may extend the life somewhat but it's probably time to look for a replacement tube. (Of course, a high dropout current could also be due to power supply, ballast, or wiring problems, so check for those first!) On a healthy tube, adding ballast resistance (25K or 50K ohms) very close to the tube anode may reduce the dropout current enough to make the laser usable, but this will also increase the operating current required of the power supply.

    3. Spectral line ratio: The ratio of the intensities of two emission lines in the discharge spectra can be used as an indicator of the tube gas pressure, more or less independent of the He:Ne fill ratio. When the ratio of 501.6 nm to 587.6 nm (both He lines) exceeds about 0.5, the pressure is declining below the range where most new to mid-life laser tubes are operating. By also measuring the Ne line at 585.3 nm, it should be possible to tell even more about the laser tube's health. Of course, even if you can find these lines with a diffraction grating, a monochronometer and sensitive amplifier will be needed to make the measurements. (See the section: Monochromators)

    For more on evaluation of HeNe lasers using these techniques, see Nondestructive Analysis for HeNe Lasers (IBM Research Report). This 1979 paper includes several graphs showing the relationship of these quantities to gas pressure.

    Note that since green, yellow, and some "other color" HeNe laser tubes are filled at reduced pressure to achieve maximum gain, their life expectancy is generally lower than for red ones, and the measurable quantities will be correspondingly modified. It will be interesting to look at their spectral lines as well.

    Examples of HeNe Laser Tube Bore Spectral Measurements

    I measured the three critical He/Ne spectral lines discussed above for several HeNe lasers:

                                     He (1)      Ne      He (2)               H2
     ID#  Manufacturer     Model    501.6 nm  585.3 nm  587.6 nm  He(1)/Ne  656 nm
    -------------------------------------------------------------------------------
      1   Melles Griot  05-LHR-006    0.33      0.55      1.00      0.60      --
      2   PMS           LSTP-1010     0.31      0.53      1.00      0.58      --
      3   American Opt. 3100          1.64      0.73      1.00      2.25      --
      4   Aerotech      LS4P          0.69      1.12      1.00      0.62      --
      5   Aerotech      LFT250        0.42      0.67      1.00      0.53     3.08
      6   PMS           Ohmeda 507    0.35      0.65      1.00      0.54      --
      7   PMS           LSTP-1010     0.40      0.67      1.00      0.60     2.75
      8   SP            130           1.19      2.63      1.00      0.45      --
    

    Brown Deposits Inside HeNe Tube Bore

    Many surplus HeNe laser tubes - both hard-seal and soft-seal - will have various amounts of a brown material coating portions of the inside of the bore, usually toward the anode-end. The presence of these unsightly deposits has no significant impact on operation or power output directly, but is an indication that the tube has seen a lot of use. And, on average, tubes with a lot of brown crud may be harder to start and require higher voltage to run - but that's more related to the high mileage than the preseance of brown crud in the bore. On high power lasers with IR suppression magnets, the brown stuff will generally collect near the magnets with obvious effects of N and S polarity. I've yet to see an explanation for this phenomenon in any laser reference. One suggestion from someone from a major HeNe laser manufacturer was that it was material sputtered off the anode but the one below makes more sense.

    (From: Chris Leubner (cdleubner@ameritech.net).)

    The usual cause is silicon being freed from the oxygen in the glass due to the intensely hot plasma on it. The ionized oxygen ends up reacting with the getter or cathode leaving elemental silicon film behind causing that brown look. In some tubes it will make a zebra or tiger stripe pattern on the bore that is a dead giveaway of both long use and plasma oscillation. On larger tubes that use magnets for IR suppression (Zeeman splitting), the magnetic fields smash the plasma into the tube wall and increases the rate of dissociation of the glass. The oxygen, which is a gas, will disperse throughout the tube and combine with the more reactive materials in it, namely the getter or cathode. The silicon will remain behind wherever it was separated because it is not volatile and relatively difficult to ionize. I do not know why it appears first on the anode end. My guess is probably due to the larger number of negative ions there reacting with the silica in the glass via this reaction: SiO2+2Ne-1=SiO+O-2+2Ne. Then SiO+2Ne-1=Si+O-2+2Ne.

    Care of HeNe Laser Tubes

    The often quoted party line is that soft-seal HeNe lasers should be run periodically to maintain their health and that this is not necessary for hard-seal tubes. The true story is a bit more complex.

    All modern internal mirror HeNe laser tubes use hard-seal construction where everything but the mirrors (where the required high temperatures would destroy the coatings) use glass-to-metal seals. Mirrors are either sealed with frit (low temperature glass powder which acts as a sort of solder for glass), optical contacting, or are fully enclosed inside the glass envelope. None of these seals leak on any time scale that matters unless the processing was defective. Melles Griot quotes a 12 year shelf life but in reality, it's virtually unlimited.

    Note that frit is quite soft compared to even optical glass so don't unnecessarily abuse the mirror seals. Those with large amounts of frit like Melles Griot and Siemens are fairly robust. But the mirrors on those with only a thin frit line like Aerotech and Uniphase may pop off if whacked the wrong way. Unless your intent was to salvage the mirrors, this would be bad news. However, even those robust Melles Griot globs of frit can be cracked (which is just as bad) by a metal tool like a pipe used in an attempt to adjust mirror alignment.

    However, there are still many external mirror HeNe lasers that use soft-seals for the Brewster window(s) and these show up surplus with varying degrees of leakage. Tubes of the same age may differ greatly in their condition, apparently due to large variations in the rate of leakage. Where the discharge color is still a pastel but quite bright - somewhat more pink than normal, even with a bluish tinge - just running the tube for a few hours or days may clean it up irrespective of the condition of the getter because the cathode itself acts as a getter - a very slow one but good enough to scavenge a small amount of contamination. The typical discharge color that is still salvageable would be the "Minor" examples in Color of HeNe Laser Tube Discharge and Gas Fill, perhaps slightly worse. Even a HeNe tube that doesn't lase at all may benefit from this simple treatment. Periodically running soft-sealed HeNe laser tubes without getters or with exhausted getters is recommended. A few hours every month is probably adequate and this will extend their life considerably, possibly indefinitely. This is much preferred compared to restoring power once it's gone. Note that any detectable (by eye) change in discharge color will be accompanied by a significant drop in output power. As the tube is operated, the discharge color will gradually approach the correct one. The last place where a normal color appears will be the expanded regions of tubing (e.g., in the glass tube that joins the side-mounted cathode to the bore in a Spectra-Physics laser). Here, the normal color is a nice orange but will tend toward pink or pinkish-blue with contamination.

    Remarkably, for a soft-seal tube, the bottom of the "Minor" samples may actually be easier to salvage by running for a few hours. I've revived both a very old SP-130B as well as a not quite so old SP-120 using this simple treatment. Both these lasers were discarded because based on the color of the discharge, the original owners thought they were too far gone for there to be any hope. The SP-130B only recovered to about one third its rated power (but it is over 30 years old!). Running it every few days for a couple minutes appears as though it will maintain that power indefinitely. (I actually run it for less than a minute daily.) The SP-120 was restored to essentially new specifications, as was a PMS tube from a Raman gas analyzer. (See the various case studies in the section: Reports from Sam's HeNe Laser Hospital.)

    However, if the discharge color is highly saturated red or blue (the bottom two examples in the above diagram) and/or there are visible striations of the discharge in the expanded regions of tubing, all hope is probably lost as no amount of operation or getter reactivation will make enough difference to matter. But there is nothing to lose by running the tube for awhile to see if a miracle occurs. :)

    When powering a HeNe tube with an off-color discharge, keep in mind that the operating voltage may be quite different than normal especially initially and may overstress the power supply if it doesn't have enough compliance. A brute force unregulated power supply on a Variac can also be used, adjusting the Variac to maintain a more or less constant current at the rated value for the tube. It's also nice to monitor the laser's output (assuming there is any eventually!) with a laser power meter to keep track of how the patient is responding to treatment. What may happen is that the power will initially increase, then decrease as the tube heats up and internal parts outgas, then gradually decrease again as the cathode acting as a getter scavenges the contaminants, and then level off. This process may take several hours or days. Powering the laser on successive occasions may result in increasing power levels if the process wasn't complete and this seems to work better in general than simply running the laser continuously. One theory is that the power declines because parts of the tube become hot enough for previously trapped gases to go back into circulation. In some cases, this results in a permanent *decrease* is power which is not recoverable. In other words, your mileage may vary.

    Hard-seal red (632.8 nm) HeNe tubes generally will not respond to these sorts treatments since there should be essentially no leakage over any time scale that matters. The gain - as modest as it may be - is suffient that any improvement may be detectable only by careful power measurements before and after. But there can be exceptions. I did have a modern Melles Griot internal mirror HeNe tube that had an off-color discharge and low power. Running it for several hours didn't help at all but activating the getter with my Solar furnace rig completely cured it permanently (it's been over two years now with no degradation in discharge color or output power so this tube isn't a "leaker" but must have not have been properly processed at the factory). See the section: Repairing the Northern Lights Tube.

    However, for "other color" HeNe lasers, particularly yellow and green ones which have very low gain (about 1/20th of red), running even a hard-seal tube for a few hours *before* thinking about touching mirror alignment can make the difference between nothing and something, even if that something is small.

    I've found some hard-seal HeNe laser tubes where the gas fill was obviously contaminated on the shelf. One example was the HeNe laser tube from a Hewlett Packard 5501A two-frequency (Zeeman split) laser head that hadn't been used in about 15 years. It wouldn't lase at all when first powered up. After running for a total of about 12 hours, it has recovered probably to essentially normal output. This type tube is of very high quality construction and no doubt was very expensive with glass-to-metal seals for electrical connections and mirrors fully enclosed inside the glass envelope. Leakage is unlikely so it must have been internal outgassing over time. Thus, even hard-seal tubes can suffer from soft-seal maladies! :) After being idle for about 2 years, the power had again declined, but only to about 25 percent of the recovered level. Running for awhile again restored it, with a rapid recovery to about the 50 percent level in a minute or so and back to 100 percent in a few hours.

    Note that end-of-life tubes will often show an off-color discharge which may be mistaken for leakage. Output power will be low or zero and there will often be evidence of shiny metallic sputtering deposits on the glass near the cathode can - a dead giveaway that the tube is end-of-life. They will also likely be hard to start with a very high operating voltage. On Melles Griot tubes, there will be shiny metallic deposits on the glass opposite the three or four holes at the rear end of the cathode can. On Hughes-style tubes, it will be on the glass at the cathode end of the tube. These tubes will not respond to any known treament.

    Tip-Off Comments

    The following ONLY applies to metal exhaust tubes; those made of glass have obvious issues of their own but if there's any problem with them, it's generally obvious!

    The metal exhaust pipe that was used to evacuate and back-fill the tube on most HeNe and small ion lasers, colloquially called the "tip-off" or "pinch-off", is sealed by a special very expensive tool that squeezes the pipe shut with incredible force and then severs it entirely. The objective is to cause the metal to cold-weld and thus be vacuum-tight. This is not like a metal compression fitting in plumbing where the connection doesn't leak simply due to cold flow of the metal and a tight fit; the two sides of the exhaust pipe metal actually become one. Or at least that's the story.

    Usually, it takes a deliberate effort to actually get the tip-off to leak, requiring filing or cutting with a hacksaw or bending over with a BIG wrench. Doing the latter is more likely to crack the glass-to-metal seal rather than affecting the integrity of the tip-off. Squeezing the exhaust tube with a pair of pliers to open it up probably won't work either. In rare cases, where some speck of something got in between the two halves of the tube while being sealed, there might be a slow leak or a weak spot subject to failure with minimal abuse, but this is somewhat unusual.

    On most Melles Griot HeNe laser tubes, there is a glob of Epoxy or something similar covering the end of the tip-off. It's there to prevent the sharp edge from attacking unsuspecting humans, not so much to protect the tip-off from damage. Other manufacturers who care about your flesh might put a piece of heat shrink tubing or a rubber protector over it. Still others haven't had enough law suits and don't bother covering it at all. :)

    However, don't be tempted to file the edges of the tip-off to smooth them. On some tubes, even a small amount of filing will result in a leak. Put your own glob of Epoxy over the end if desired.

    Also, although bending (a longer) tip-off shouldn't affect its integrity, if this is required (for example, to provide clearance to fit the mirror mount stem into a tight space) hold both sides so that no stress is applied to the end-cap since this may crack the glass-to-metal seal.

    Troubleshooting an External Mirror HeNe Laser

    In addition to all of the problems of internal mirror HeNe tubes, external mirror lasers are subject to dirty optics and much more prone to have misaligned mirrors.

    Like their internal mirror counterparts, the general appearance of the output when non-lasing will be a diffuse blue, blue-green, or purple spot but no red light. If there is any evidence of a red beam, something may be marginal but it is lasing.

    If it won't start, then the tube could be up to air or there could be a power supply problem. Try another power supply if available. Or, see the section: How Can I Tell if My Tube is Good? for info on using a low level RF or microwave source to check for ionization.

    Assuming the tube lights up, follow the steps below to narrow down the cause:

    1. If possible, check the tube current. Though unlikely, a discharge current much much greater than optimum will result in low or no output beam (as well as overheating of the tube, ballast resistor(s), and power supply components.

    2. Compare the color of the discharge in the narrow bore/capillary with Color of HeNe Laser Tube Discharge and Gas Fill. The comments about output apply to red and maybe orange HeNe tubes; yellow and green HeNe tubes will likely produce no output at all unless the gas fill is nearly perfect. The normal appearance is a white-ish red-orange generally described as "salmon color" though there can be a fair range from more orange to more pink for a good tube depending on the exact gas fill He:Ne ratio and pressure. Lower pressure tends toward pink and may be normal for low gain non-red HeNe tubes to boost gain at the expense of tube life. Some of these may also be even more of a white-ish shade. To confirm, check the discharge spectrum, preferably with a spectroscope but a diffraction grating or prism may be adequate. It should be similar to the combination of the helium and neon spectra in Bright Line Spectra of Helium and Neon. If the color inside the bore appears normal, check the color of the discharge where it isn't as constricted - the color should be quite orange. (This can be seen in the funnel area near the anode on most internal mirror tubes or the expanded tubing sections on those like Spectra-Physics side-arm cathode tubes with exposed capillaries.)

      • Where both these colors are correct (salmon and orange), gas fill is probably not the immediate problem.

      • If the discharge color in the expanded areas is the same as the bore or more towards white or blue, the gas fill is somewhat contaminated and marginal - lasing may not occur (definitely not for yellow and green HeNe tubes).

      • A slightly more pinkish discharge in the bore but with the normal orange color elsewhere may indicate low gas pressure and near/at end of life. There is no cure (but as noted above, this may be normal for some tubes).

      Firing the getter (if any) or just running the tube for an extended period of time may clean up any slight contamination (but won't help low gas pressure). However, if it is very pink, blue, purple, or white, a significant amount of air has leaked in over time, probably via the soft-sealed Brewster windows, and the only cure is likely to be a tube transplant. This is probably the most common problem with older external mirror HeNe lasers. Unfortunately, it isn't cost effective to refill them and replacement tubes are likely to be very expensive - if they are available at all.

    3. Check the mirrors, Brewster windows, and any other intra-cavity optics for damage and clean them if necessary using the proper optics cleaning technique.

      See the section: Cleaning of Laser Optics for the recommended procedure.

    4. The only remaining cause of a non-lasing laser with a proper discharge color and clean optics is mirror alignment. If the laser was dropped (Ack!) or someone decided the alignment screws were loose and tightened them, see the section: >Sam's Approach for Aligning an External Mirror Laser with the Mirrors in Place or the more general procedures starting in the section: External Mirror Laser Cleaning and Alignment Techniques.

      Of course, this assumes that the optics are correct for the laser or that someone didn't remove a mirror for use in their science fair project! Note that alignment is super critical, especially for a long HeNe laser. Thus, if misalignment is found to be the problem, it may require a lot of patience, determination, and the proper jigs, to remedy it. You won't succeed by luck alone (though luck may play a part)!

    External Mirror HeNe Laser Health Checklist

    When considering the acquisition of a large-frame HeNe laser, here are some specific things to look at or questions that can be asked of the owner which will help to determine if the laser is likely to be functional near original specifications. The following applies directly to external mirror Spectra-Physics HeNe laser tubes with minor modifications for other manufacturers/models:

    Once the laser can be powered up, check the discharge color in the bore. It should be similar to the bright white-ish red-orange or 'salmon' color at the top of Color of HeNe Laser Tube Discharge and Gas Fill, or of any other fully functional HeNe laser tube. If it does not, either the tube is soft-seal and has leaked, or it has been very totally abused. See the sections starting with: HeNe Tube Use and Life Expectancy. If the discharge color looks good, then very likely mirror alignment is all that is needed to achieve at least a substantial fraction of full power.

    Can I Increase Output Power Using the Waste Beam from the HR?

    When operating a bare HeNe laser tube, you've no doubt noticed the weak beam that exits from the supposedly totally reflecting mirror (the HR or High Reflector). (This assumes it isn't covered with tape or paint.) So, could the power from the output end of the laser be increased by putting another mirror behind the HR?

    The quick answer is that this might be possible in theory.

    The practical answer is: forget it.

    The long answer is too involved to go into here but if the extra mirror were properly aligned AND an exact multiple of 1/2 wavelength of 632.8 nm from the other mirror AND if there were no losses from the non-AR coated HR surfaces, part of the wasted power might appear at the output.

    But, in the end, all you would gain at most would be the couple microwatts that escapes out the HR. :) The lost power isn't much on most tubes. For those occasional tubes where the output is significant from the HR (either because of a mistake in manufacture or by design), there might be more benefit but as a practical matter, there is no way to satisfy all the conditions in a stable manner without a fancy feedback loop, if at all.

    (From: Steve Roberts (osteven@akrobiz.com).)

    Assuming it's a standard TEM00 mode HeNe and not a multimode laser, you'd see little tiny increases and decreases in the power on a very sensitive power meter as the mirror was translated toward and away from the existing rear mirror. But you would not really recover any of the rear beam, in fact you'd confuse the lasing going on inside the main cavity somewhat, and at certain possible "magic" combinations of external reflector and distance, cause lasing to actually cease. In practice, HeNe lasers tend to run by default at their maximum possible gain for a given combination of tube optics.

    If you want to see one wink out or flicker, precisely anchor it to a stable bench and then use a third flat mirror some distance away on a precision mount to reflect the output back down the bore. When the reflected beam is 180 degrees or so out of phase with the wave in the cavity, it will wink and flicker.

    (From: Sam.)

    I wonder about this...

    To actually interfere with lasing in a typical HeNe laser may be more difficult than Steve claims. While flickering and apparent instability will be seen if this experiment is done with a common HeNe tube, it may only be a result of the output beam interfering with itself outside the cavity when reflected back to the OC. This could appear to be confusing lasing but may actually not cause any substantial effect inside the cavity. Monitoring the waste beam (as noted below) can be used to determine whether the behavior is due to external or internal interference. If it's only external, the waste power will be almost unaffected (just the portion of the reflected output beam that gets back through both the OC and HR). This is likely to be less than 0.1 percent of the output power or a couple percent of the waste beam power at most. However, if actual lasing is being affected, the waste beam power will fluctuate significantly - up to (as Steve suggests), total wink-outs. :)

    (From: Bob.)

    On a somewhat related side note, there is at least one commercial instrument I know of that focuses the output of a HeNe laser onto a surface, and has a highly sensitive photodetector behind the HR of the laser (the arrangement Steve mentioned, but in reverse). As the surface the light is focused on moves back and forth in relation to the laser, the photodiode detects changes in output power out the back end. Basically, this is a form of a Fabry-Perot interferometer which can be used to very precisely measure small distances.

    Reducing the Output Power of a HeNe Laser

    Anyone who asks to do this is generally referred to a good therapist :) but there can be valid reasons to limit the output power from a laser to a fixed maximum. One such reason would be to fall within a particular laser safety classification like Class IIIA (5 mW). There are a variety of techniques to reduce the output power of a HeNe laser but only some are reversible:



  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    Connections, Electrical Characteristics, Output Power

    Identifying Connections to Unmarked HeNe Tube or Laser Head

    Where you can visually inspect the wiring, this is trivial as we all know that the large aluminum 'can' electrode is the cathode (negative) terminal.

    CAUTION: While most modern HeNe tubes use the mirror mounts for the high voltage connections, there are exceptions and older tubes may have unusual arrangements where the anode is just a wire fused into the glass and/or the cathode has a terminal separate from the mirror mount at that end of the tube. Take note of the cathode arrangement in particular because the tube will still lase perfectly if you attach to the mirror mount but instead of the actual cathode but that will result in sputtering near the mirror which is about the worst place for this - similar to running the tube on reverse polarity. (Miswiring the anode might result in no or weak lasing but probably no permanent damage.)

    Alden high voltage connector

    The two pin Alden is by far the most common connector used for attaching HeNe laser heads to HeNe laser power supplies. They are used by almost all manufacturers and for lasers almost all sizes. The shorter (narrower) side goes to the anode (positive) and the longer (fatter) side goes to the cathode (negative). When such a connector is present, there will also be a ballast resistor (typically about 75K ohms) built into the HeNe tube assembly or laser head between the Alden's positive terminal and the anode.

                                  ________-__
                     Anode (+) ==|________|  |---_______
                             _____________|  |  |_______ HV Cable   
              Cathode (-) ==|_____________|__|---
                                          -
    

    Or see High Voltage Cable with Male Alden Connector. This one is built with separate wires and appears to have a ballast resistor built into the anode (red) lead (or maybe it's just a wart!). Many use coax similar in thickness to RG58U for the HV cable instead.

    Note: Genuine Alden brand connectors will have the name stamped on the plastic. Some power supplies may come with Alden compatibles without identification. This probably doesn't matter in any way, shape, or form, except as an indication that the power supply manufacturer installed the connector onto existing wiring or saved a few cents. :) For complete info, go to Amphenol Alden Products Company. Go to "Products", "High Voltage Connectors and Cable Assemblies".

    Three pin in-line high voltage connector

    Some larger HeNe lasers (mostly from Siemens and Spectra-Physics) use a connector somewhat similar to the standard Alden but it is more rectangular with 3 pins instead of 2. And, the pins on both of the connectors (male and female) are recessed to avoid the shocking experience of touching the pins on a recently detached laser head and getting zapped!

    CAUTION: The pinouts for Siemens and Spectra-Physics lasers have the HV pins swapped! Using the wrong one may result in very rapid destruction of laser head and/or power supply, not to mention a possible shocking experience!

      Pin Location     Siemens     Spectra-Physics
     ----------------------------------------------
      Square end     Earth Ground   Earth-Ground
      Middle         Cathode (-)      Anode (+)
      Round end       Anode (+)      Cathode (-)
    

    For these large laser heads, there may also be a small ballast resistor in series with the cathode lead. Bypassing it will reduce operating voltage requirements and the laser will probably still work fine though the claim is that stability will be better with it when used with the recommended power supply.

    Two pin 20 kV (possibly Alden) connector:

    This one is larger than the standard Alden and has a square pin and round pin. The square pin is the cathode and the round pin is the anode, at least for Melles Griot lasers.

    Three pin round High voltage connector

    Some Spectra-Physics lasers use a special 3 pin round connector (view is looking toward power supply):

                   O  Positive (Anode)
                   1
          GND  O 3   2 O  Negative (Cathode)
               
                   o  Interlock Prong
    

    The GND may not actually be present on some power supplies. In most cases, it is already connected to the negative elsewhere. The interlock prong activates a microswitch in the power supply to complete the primary-side circuit only if the power supply and laser head are securely attached. This provides protection for the power supply but isn't present on all models. (If your laser refuses to lase and there is no interlock prong, it's possible that the power supply requires it. It's either fallen or broken off, or the power supply isn't the one intended for your laser head.)

    No standard high voltage connector

    However, suppose the whole thing is sealed and all we have are some dangling wires or an unusual unmarked connector? Here are some guidelines. Try to obtain agreement on several of the following tests as no single one is necessarily a guarantee of correct identification:

    Getting the HeNe Tube Out of a Laser Head Intact

    If the tube is bad and won't be salvaged, more drastic means can be used. However, assuming you want to extract it intact and with a chance of being reused, the choices narrow dramatically.

    With rectangular laser heads, the actual HeNe tube will probably be mounted in a sane fashion - with screws and clamps for example. So, no problem if you have the correct screwdrivers. In rare cases (particularly for modern large ones, you may find that there is a cylindrical laser head inside! And some (like those from Spectral) seem to be glued together.

    For cylindrical laser heads, the tube may be mounted in a variety of ways. Just getting the end-caps off can be a fun experience as well. They may be mounted with screws or set screws (for which Murphy's law states you won't have the correct hex wrench), rivets (some drilling required), just glue (which will likely be hard and brittle by the time you need to do this - probably an advantage). As for the tube, there may be (plastic) set screws at 3 or 4 points around the outside in two locations - front and back. In this case, loosening the set screws should allow the tube to be slid out of the housing. If it still doesn't move, check for additional anchors or wiring connections at either end. If it still doesn't move, there may be some RTV, hot melt glue, or other adhesive in a hidden locations still securing it.

    Often, you will find that the tube itself has been set in place with Silicone RTV forced through holes on the side to keep it there. Unfortunately, removing these tubes intact appears to be right up there with dropping bare eggs from 10th story windows and having them survive unbroken in the level of difficulty department. :) However, it can be done without dynamite. (But, before going through any of the following RTV removal gymnastics, determine if the adhesive is actually something less stable than RTV, see below.)

    As we all know, Silicone RTV, a.k.a. GE Bathtub Caulk, be it white, black, or clear, is impervious to virtually everything but a good sharp blade. If there is enough clearance around the tube, it may be possible to slip a thin strip of metal in there and carefully slice the RTV from each end. I've done this to extract a couple of HeNe tubes intact. The first was dead (up to air) so I wasn't too worried about breaking it. I used thin aluminum strips (e.g., roof flashing) from either end and through the fill holes to grind away at the RTV until the tube could be removed - surviving with just a few scratches as aluminum is softer than glass! This literally took HOURS! However, there is often not even enough clearance for this to be possible. For my laser head, this was the case on *opposite* sides at each end even for the .015" aluminum. Only when enough RTV had been removed on the side with more clearance could it be worked loose. (In addition to the tube being dead, it had been mounted skewed in its cylindrical prison - someone must have had a really bad day when this thing was put together!) The second tube was weak (putting out only about 1/3 mW when it should have been 2 mW). It came out quite easily (still putting out only 1/3 mW) as the adhesive was localized and could be sliced with a single pass of my 'tool' for each small glob of RTV.

    Sometimes a hard non-RTV type adhesive is used in a similar manner to the RTV. For this, a narrow coping saw or model maker's saw blade between the tube and housing should work quite well.

    If you don't care about saving the housing, very carefully use a hacksaw to remove it as close as possible to the adhesive clumps (near the ends of the HeNe tube). This will make it easier to get at the glue with a thin knife, saw, razor blade, or that roof flashing. A copper tubing cutter may even work for this but go real slow or the distortion of the housing may crunch the tube. :(

    One might think a chemical exists capable of dissolving RTV that isn't totally toxic and disgusting. Such a substance would make this task a whole lot easier. Is there?

    (From: Mark Schweter (schweter@mail.bright.net)).

    Short of ashing the assembly (which will strip your wires for you too!), not really. (Considering the NON_toxic, NON_disgusting requirements - assuming you mean Silicone RTV, fuming HF or HNO3 comes to mind!)

    Fully cured RTV is fairly stable, unfortunately.

    You might try a NaOH solution to digest the RTV, if nothing else, it'll take the aluminum 'can' off! (NO smoking in the area PLEASE - HYDROGEN is released!)

    A thought occurs to me.... Get a 'slitting saw' or 'burr' and slice the aluminum can lengthwise, several times. Use a hot-knife to peel away the RTVed sections. Then use the hot-knife to pare RTV off glassware. My Weller soldering gun used to have one.

    (From: Mark Shipley (mark@startrek.com).)

    I have successfully removed an old Hughes HeNe tube from such a head by using an old piano wire (violin, cello, etc., as long as the wire was wound, it would work). (You hated the practicing, anyhow! :) --- Sam.)

    Pass the wire down the side of tube, anchor the end, say in a vice and slowly work the tube back and forth pressing the caulking against the wire. The wound wire cuts away at the caulking and after not too much time you should free the tube.

    (From: Dave (ws407c@aol.com).)

    I have yet to have a problem removing end-caps from the Melles Griot HeNe laser heads I have had after my tried and true tested method. :-)

    Fill a coffee cup about 3" high with BOILING hot water and let the head sit in it for about 10 min's. Repeat 3 times and the cap pops off by hand no problem. After it is removed, run your thumb around the inside to remove the remaining glue. Use a hair dryer to clear up the condensation inside the head from the process.

    Repeat for the other end. This has worked for most of the PMS and Uniphase heads as well.

    Removing a tube from ANY head is a cinch (if you're willing to sacrifice the aluminum cylinder) by using a hacksaw. There is no need to remove the end-caps in this case. First remove any set-screws. In Melles Griot heads there are usually two sets of 3 (alternating with glue-only holes). Use a sharp blade or Dremel(tm) tool to cut a slot in the plastic and then just unscrew them (COUNTER-CLOCKWISE!). Next roll the head across a table while making a mark around the middle of the head to follow with the hacksaw. Saw slow and carefully as not to nick the tube. The metal is soft and wont take too long to cut. When the cut is finished, squeeze some liquid dish-washing detergent (Ivory, etc...) into the head followed by some water. Give it a shake and then twist one way with the left hand and twist the other way with the right and the glue will give way most easily. :-) Make sure there are no set-screws hidden in the RTV or whatever it is. Once one end of the case frees up, cut the wires and pull it off. From here do the same for tube in one hand and half of head in the other. Once the tube is free and still soapy, pick off the rest of the glue and "starter tape". Then wash off the tube with fresh water and use a hair dryer to dry it off to prevent any trace of rust.

    I have done this over and over again without any problems or stress to the laser tube.

    (From: Sam.)

    CAUTION: If all you have removed using the hot water trick is one or both end-caps, DON'T attempt to run the tube until you are sure all moisture is gone from inside the head. Otherwise, there may be corona/arcing at various places which at the very least, will make it hard to start and may cause damage to the head and/or power supply.

    I have simplified this tube removal technique a bit if the end-caps have been taken off and don't need to cut the cylinder at all. Remove the six (6) nylon set-screws by first scribing with a sharp knife or Dremel cutoff wheel and turn COUNTERCLOCKWISE. Then carefully use a paper clip or knife blade to dig out the hot-melt glue in the other six (6) holes. (This helps to free up the attachment to the inner wall.) Put the head into hot water for a couple of minutes. Hot water from the tap is probably adequate and a bit of dish washing liquid won't hurt to make it slide easier. The heat also expands the aluminum faster than the glass. Then, finger pressure alone on the metal cathode end-bell should be sufficient to break any remaining attachment of the hot-melt glue and slide the tube a fraction of an inch inside the cylinder. Then, just push it back out from the other end. It may take several applications of hot sudsy water to loosen the tube but if the set-screws have been removed and the hot-melt glue holes cleared, it should work eventually. I've done this with several heads without damage to the tube inside.

    CAUTION: Since this is basically a fragile glass bottle you're trying to get out with some force (though hopefully not much), accidents can happen. Therefore, provide some protection between the tube and your fingers when pushing.

    I know this works with hot-melt glue-mounted tubes. This includes most or all newer Melles Griot tubes but some older Aerotech tubes use generous amounts of very resilient RTV which may not loosen up at all. If you're real lucky, your tube is just held in place with set-screws, like NEC. And some Aerotech heads have minimal blobs of RTV with set-screws doing most of the work.

    Long Uniphase HeNe laser heads (e.g., 1145/P) have 4 sets of RTV holes. But even after gouging out all the RTV, the tube may still be held securely with some sort of rubber padding leaving the tube still locked in place. Trying to push it out is made more difficult by the rock-hard poured-in-place ballast resistor assembly at the anode-end which only leaves the mirror mount exposed. Trying to push on the mirror mount at that end is asking for trouble because that's about the weakest part of the tube where the glass structure is only about 1 inch in diameter and relatively thin. The hacksaw approah may be the only option if you want the tube intact.

    But, the following would appear to be the definitive word on dealing with Melles Griot and other HeNe laser heads that use a non-RTV type rubber for mounting the tubes. Test a bit of the adhesive to determine if some heat will soften it - if so, your task is much easier.

    Replacing the HeNe Tube in a Melles Griot Laser Head

    So you found a new tube with 10 times the output power of that tired, bedraggled, worn old one taking up space inside a cylindrical laser head and it would be nice to replace it. This is possible as noted above but the following is perhaps the semi-official way of doing it and applies directly to most Melles Griot cylindrical laser heads but others may be similar.

    Note that some of the recommeded procedures will stink up the house so you may want to do this somewhere else like someone else's house. :)

    (From: Lynn Strickland (stricks760@earthlink.net).)

    The HeNe tube is usually mounted and aligned using nylon screws, then potted with RTV Silicone or hot-melt glue, and then the screws are cut off.

    1. Take off the end-caps (if it has any) by first clamping the head lightly in a vice. Heat the ends of the housing with a heat-gun, one at a time, and when the epoxy starts to smell, pull the end-caps off with a pair of locking pliers. Piece of cake, but don't burn yourself. Oh, some heads have pressed-in metal end-caps. If it does, you're on your own. (I have been known to use a pipe cutter though).

    2. Look at the head and find the potting holes. Half will probably be Nylon screws. Heat up an exacto blade and melt a small slot into the screws. Then unscrew them, and save them for later.

    3. If the alternate holes are hot-melt (and they probably are), put the entire head in the oven for 10 to 15 minutes to soften the hot-melt, then push the tube out of the housing. It kind of smells up the kitchen if you're not careful, and I take no responsibility if you burn down your house.

    4. Suspend the new tube inside the housing, using Nylon screws. (You saved, them, correct?)

    5. Ideally, you would then re-tweak the mirror alignment once it's mounted inside the housing. A filed down hex wrench can be used on the locking collar at the cathode/output-end but it isn't easy and you'll probably be okay to skip it. A better method is to carefully drill access holes for a hex wrench opposite each of the three adjustment screws in the side of the cylinder. Tweaking the anode-end mirror without getting shocked is the real challenge though and I would definitely recommend skipping it! Putting the end-caps back might help stabilize it a bit more, but you've won most of the battle already.

    (From: Daniel Matthews (daniel@wpmedia.com).)

    To disassemble, I first remove the screw in plugs by slicing into them with a hobby knife and then unscrew them. After that, I put on a pair of thick gloves and heat them in front of a ready heater until they're hot enough to push the tube out of the aluminum housing. Then, I clean the melted rubber off of the glass.

    I also have heads here that I reassembled. I put the centering plugs back in, screwed them all down flush leaving the tube snug and centered. Then, I inject black RTV Silicone into the other holes. After the RTV it cures then I trim the plug with a razor blade to leave a smooth fill level with the aluminum. Just looking at it, you can't tell they were ever disassembled.

    (From: Sam.)

    So with RTV, the next guy to attempt to disassemble the head will be using all the 4 letter words. :)

    Determining Electrical Characteristics of Unmarked HeNe Tube or Laser Head

    So, you found this fabulous HeNe laser in the dumpster and would like to power it. There are no markings of any kind - not even the manufacturer is known. If it is a laser head, the tube itself may be labeled - if you can get to it nondestructively. Where even the tube is unmarked, start by narrowing down the range of expected electrical characteristics:

    The rest is experimentation. You will need an HeNe laser power supply capable of handling a tube with the worst case voltage and current based on its size. Make sure you include a 75 K ohm ballast resistor of adequate wattage (10 W will be sufficient for anything up to 10 mA). A laser head will usually have an internal ballast resistor. Make sure the polarity is correct - see the section: Identifying Connections to Unmarked HeNe Tube or Laser Head.

    Once you get the tube to light, adjust the current for maximum beam intensity. Running at slightly higher than optimal current won't do any immediate damage but shouldn't be allowed to continue for too long. It's best to do this with a laser power meter but your standard complement of eyeballs will be close enough for most purposes. If using a meter (you probably won't notice the following effects visually), give the tube a few seconds to stabilize after a change in current - sometimes the power output may initially increase but then settle back to a lower level and you might as well operate the tube at the lowest current that results in maximum output. Then, label the HeNe tube or laser head with your findings so you will know how to deal with it the next time you pull it out of the cabinet. :-)

    Measuring the Negative Resistance of a HeNe Tube

    This is probably more of an academic exercise than anything else. However, a change in the tube's voltage versus current characteristics could indicate a problem with gas fill composition or pressure, or damaged or deteriorated electrode(s). A comparison between the suspect tube and an identical model that works could be revealing.

    For currents within and well beyond the normal operating range, a HeNe tube acts as a negative resistance - reducing the current results in an increase of tube voltage and vice-versa. Reducing current also results in an increase in the magnitude of the incremental negative resistance. Below 2 mA or so for a typical small HeNe tube, this magnitude rises so quickly that it is impossible to maintain a discharge even with very large values of ballast resistance. Going the other way, at some very large current (probably measured in amps), the incremental resistance turns positive (just before the tube melts or explodes!). For any given HeNe tube, power supply, and ballast resistor combination, there will be a range of current over which the discharge will remain stable. This is roughly the range over which the negative resistance of the tube plus the effective resistance of the ballast resistor, power supply, and regulator (if used) remains positive.

    Measuring resistance, negative or otherwise, is just a matter of determining the relationship of voltage to current for the device. It is trivial for common electronic components but more complicated for HeNe tubes due to the high voltage (particularly the starting voltage) produced by the power supply. (See the section: Making Measurements on HeNe Laser Power Supplies.) However, if you have a high impedance high voltage probe for you DMM or VOM, or a high voltage meter, it can be left attached even during starting without fear of a melt-down (though even its high resistance and small capacitance may alter tube behavior and/or prevent starting).

    One straightforward approach will require the following:

    A typical circuit is shown below:

    
                       Rb                     Rm
         HV+ o--------/\/\------+-------+----/\/\----+
                      75K       |Tube+  |    20M     |
                              .-|-.     |             / Close ONLY after
                              |   |     o         S1 |   tube has started!
                              |   |     +            o 
                          LT1 |   |     V            +
                              |   |     -           VOM (20M input, reads V/2)
                              ||_||     o            -
                              '-|-'     |            o
                      Rs        |Tube-  |            |
         HV- o---+---/\/\---+---+-------+------------+
                 |    1K    |
                 o -      + o
                 Current (I)
               1V/mA or direct
    
    

    Note: Where the VOM or DMM is connected after the starter (to the tube or head), a power supply with a high impedance parasitic voltage multiplier starting circuit is recommended to minimize the risk of damage to your meter should the tube drop out during the tests. The load of the meter will prevent such a circuit from developing significant damaging voltage. See the chapter: Complete HeNe Laser Power Supply Schematics for some suitable designs.

    To provide additional protection for your meter, consider putting a series stack of neon bulbs (NE2s, about 90 V each) across its input to bypass any voltage greater than the expected value while the tube is lit. For example, if the maximum range of your meter is 1 kV, use 11 or 12 NE2s.

    For the following, I assume the circuit above.

    I ran some tests on several small HeNe tubes using the following slightly modified circuit:

    
                   Rb                    Rm
     HV+ o--------/\/\------+-------+---/\/\---+
                  75K       |Tube+  |   15M    |
                          .-|-.     |           / Close ONLY after
                          |   |     |       S1 |   tube has started!
    (From AT-PS1,         |   |     o          +----+-----+
     AT-PS2B, or          |   |     +          |    |     |
     05-LPM-379       LT1 |   |     V       Rc | Cb |     o 
     depending            |   |     -       2M /   _|_    +
     on the tube.)        |   |     o       +->\   ---   DMM (10M input - Adjust Rc
                          ||_||     |       |  /    |     -    so that DMM reads
                          '-|-'     |       |  \    |     o    exactly V/10.)
                  Rs        |Tube-  |       |  |    |     |
     HV- o---+---/\/\----+--+-------+-------+--+----+-----+
             |    1K     |                     .01uF,1000V
             | +-------+ |
             +-| 10 mA |-+ M1 (Panel meter plugged into current sense test
             - +-------+ +     points on AT-PS1 or AT-PS2B front panel,
                               or in-line meter adapter for 05-LPM-379.)
    
    

    Depending on the voltage requirements of the tube, I used either Aerotech Model PS1 HeNe Laser Power Supply (AT-PS1) (most tubes up to 1 mW), Aerotech Model PS2B HeNe Laser Power Supply (AT-PS2B) (tubes above 1 mW), or a Melles Griot 05-LPM-379 (for the 05-LHR-640 tube). Current control was via the adjustable internal regulator when using AT-PS1 or 05-LPM-379 but with a Variac for AT-PS2B (its regulator is currently disabled). Both of the Aerotech units have parasitic voltage multiplier starters and with the circuit wired as shown above, even if the tube cuts out, the maximum voltage doesn't go above about 2.5 or 4 kV for the AT-PS1 and AT-PS2B, respectively (maximum of 400 V at the DMM itself). The output of the 05-LPM-379 may go somewhat above 400 V under these conditions but the Radio Shack DMM I'm using doesn't seem to mind.

    And, yes, S1 is just a clip lead. :)

    The following charts summarizes the results (I was too lazy to graph these data or take measurements every .1 mA!):

              | Melles G.  Metrologic Spectra-P. Uniphase   Aerotech   Melles G.
              |  LHR-002     ?????       88         098       LT2R      LHR-080
      Current |  .5-1 mW     .8 mW     1.25 mW     1 mW       2 mW       2 mW
       I(n)   | V(n) R(n)  V(n) R(n)  V(n) R(n)  V(n) R(n)  V(n) R(n)  V(n) R(n)
     ---------+-------------------------------------------------------------------
      2.5 mA                          1135       1103
      3.0 mA    1141                  1095 -73K  1064 -70K
      3.5 mA    1110 -61K*  923       1062 -59K  1033 -57K  1667
      4.0 mA    1080 -58K   896 -46K* 1036 -46K  1007 -42K* 1631 -64K  1519
      4.5 mA    1052 -49K   877 -31K  1016 -31K*  991 -30K  1603 -51K  1480 -69K
      5.0 mA    1031 -37K   865 -23K  1005 -24K   977 -26K  1580 -47K  1450 -58K
      5.5 mA    1015        854 -21K   992 -23K   965 -22K  1556 -44K* 1422 -50K
      6.0 mA                844        982        955       1536 -40K  1400 -38K
      6.5 mA                                                1516 -36K  1383 -29K*
      7.0 mA                                                1500       1371
      8.0 mA
      8.5 mA
    
              | Melles G.   Melles G.   Melles G.
              |  LHB-570     LHR-050     LHR-640
      Current |   4 mW        5 mW      0.5-1 mW
       I(n)   | V(n) R(n)   V(n) R(n)   V(n) R(n)
     ---------+-------------------------------------------------------------------
      2.5 mA
      3.0 mA    1130                    955
      3.5 mA    1100 -50K               922  -55K
      4.0 mA    1080 -40K               900  -47K
      4.5 mA    1060 -35K   2013        875  -39K*
      5.0 mA    1045 -28K   1970 -73K   861  -29K
      5.5 mA    1032 -27K   1940 -50K   846  -29K
      6.0 mA    1018 -24K   1920 -35K   832  -24K
      6.5 mA    1008 -23K*  1905 -29K*  822
      7.0 mA     995 -22K   1891 -24K
      8.0 mA     986        1881
      8.5 mA     980
    

    The '*' denotes the approximate recommended operating current for the tube (more or less guessed if the data wasn't available!). Below the lowest current listed for each tube, the magnitude of the (negative) resistance increased beyond the point where stability could be maintained with the 75K ballast resistor and the tube would not remain lit. It is interesting that the two lowest power tubes (both 12.5 cm long, bore approximately 0.5 mm) have their operating points close to the dropout current. Rb for these tubes is typically increased to 100K or more to assure stability.

    The LHB-570 is a wide bore multimode one-Brewster HeNe tube so the 4 mW is actually only valid for a particular OC mirror. Note the low operating voltage and magnitude of of the negative resistance for this tube.

    I'll add other tubes as the opportunity presents itself.

    Due to the effects on the V-I characteristics with temperature, there was some drift in the readings. For example, going to the highest current listed above for a particular tube and then back to the lowest current resulted in perhaps a 1 to 2 percent change in voltage until the tube cooled down.

    More sophisticated analysis is left as an exercise for the student. :)

    Determining Output Power

    This would be easy with a laser power meter. However, most of us are not so fortunate as to own such an instrument. See the section: What Makes a Laser Power Meter So Expensive?. There are two aspects of this same problem:

    1. What is the specified power output of a given HeNe tube?
    2. Is your tube really producing its specified output power?

    Note: There will often be a CDRH safety sticker (usually yellow or white) on the HeNe tube or laser head. The wattage listed on this sticker is NOT a reliable indication of output power. It is an upper bound and may be much higher than either the rated or actual output power. For example, a .5 mW laser will likely have a safety sticker value of 1 mW; a 1 or 2 mW laser will show 5 mW; and a 12 mW laser may show 15 or 25 mW. Some unscrupulous or careless HeNe laser or tube resellers will list this as the power output of the device - buyer beware! Few people can or will check this. If it sounds to good to be true, it probably is. :-(

    1. There are a few ways of determining the tube's specified output power:

      • Some manufacturers code the (usually minimum) output power into the model number. For example, Aerotech tubes and laser heads have a model number that is of the form: XYZ where X is the model designation, Y is the output power in mW (e.g., 2 = 2 mW, 05 = .5 mW, etc.) and Z is either R or P denoting a random or linearly polarized beam respectively.

      • Match the model number of the tube or laser head to the manufacturer's catalog listing. This may be easier said than done since many surplus tubes either don't have a model number printed on them or are old enough (but still perfectly good) so that the model is no longer listed in a current catalog. The manufacturer (if they still exist) will know and contacting them may be worth the effort. However, don't expect an overly enthusiastic response if you are asking about a 10 year old $20 HeNe tube! Any information so obtained may not be accurate either.

      • Attempt to compare the physical dimensions with those of tubes with known output power. This is not very reliable as the output power of a tube of identical diameter and length can easily vary by a factor of two or more by design or just due to sample-to-sample variations (at the time of manufacture, tubes are selected and sold based on their actual output power but they may appear to be physically identical).

      • Try to locate an indication on the tube itself of *measured* power output. Very often, the actual power output determined at the time of manufacture will be hand written or printed somewhere on the tube. This may be on the glass or metal shield (if used) or one of the ends. It may be on the outside of a laser head on the manufacturer's specification sticker (not the safety sticker, see above) or concealed inside. Examine every nook and cranny and the tube's secrets may very well be revealed!

      • Attempt to use the following equation to calculate expected output power:
                                              q * L
                           Po = T * A * I * (------- - 1)
                                              T + B
        
        Where:

        • T is the output coupler (OC) transmission in percent.
        • A is the cross sectional area of the beam.
        • I is a saturation parameter.
        • q is the small signal gain.
        • L is the gain length.
        • B is the sum of all cavity losses.

        For the typical internal mirror HeNe laser tube, q =.15/m and B will be close to 0 assuming there is no internal Brewster plate or etalon. A and L can be measured for your HeNe tube. Unfortunately, T and I are likely to be unknown but they can perhaps be estimated by comparison with another HeNe tube having a known power output. This would make an excellent exercise for the student! :-)

        However, what this equation does show is that all other factors being equal, when comfortably above the lasing threshold of (q * L)/(T + B) > 1, output power is proportional to bore length times its cross sectional area. But we already knew that!

        Of course, as noted above, the actual output power for any given sample tube of identical construction and dimensions can easily vary by a factor of two. The calculated value is at best the theoretical maximum - when the tube is new (or at its peak if initially overfilled with helium to compensate for loss over time), under ideal conditions, and possibly only on alternate Thursdays! :)

    2. It is much tougher to determine if the output of your HeNe tube is actually correct without a calibrated laser power meter. However, comparisons can be made.

      Maximum output power isn't achieved instantly for an HeNe laser when power is applied. Typically, it starts at 75 to 85 percent of its final value and reaches that only after a 10 to 20 minute warmup period. For long tubes or large frame lasers, an hour may be needed for the output power to stabilize. I've also noticed that power seems to peak and then decline slightly for many tubes during this warmup period. I don't know if this is an inherent properly due to the increasing temperature of the bore or just a matter of mirror adjustments not being optimal. Power also may take a few seconds or longer to stabilize after even a small change in operating current. Depending on where you are on the current versus output curve, it may go up and stay up, go down and stay down, or do one of these and then return to nearly its former value.

      In addition, for high power really long HeNe tubes (e.g., 15 mW or more) and/or unconventional HeNe tubes used in high quality lasers, there may be other physical factors affecting power output including mirror micro-adjustments, need for IR line suppressing or discharge stabilization magnets, rigid temperature and external force stabilized mounting, and even tube orientation (like: This Side Up!). In fact, where you have a weak beam or even no beam at all, gently pressing in the center of these long tubes (which bends them ever so slightly) can be a useful technique to determine which way the mirror alignment is off without actually touching the mirror mounts (though you will have to do this eventually to make the adjustments). In fact, just touching one side of the tube with your hand will cool it slightly and may result in a significant change in output power due to the change in mirror alignment due to thermal contraction and bending of the tube!

      For lasers with very long bores that are exposed (e.g., the SP-127), there may be one or more adjustments along the length of the bore to fine adjust its straightness. While slight misadjustment of these won't result in no beam, it could certainly greatly reduce power output.

      See the section: How Can I Tell if My Tube is Good?. However, none of these should be a major factor for small common inexpensive HeNe tubes (though there still may be some effects).

      • First, confirm that you are supplying the proper operating current. The output beam power will be maximum at the proper current - lower on either side. A power supply with a broken regulator could be producing greatly increased current which will result in much reduced output (and excessive heating, sputtering, and shortened tube life - and probably won't do the power supply much good either). At 2 to 3 times the rated current, there may be no beam at all!

      • If you have access to a working HeNe laser or a new HeNe tube with known output, this is best as the wavelength will be the same. But, keep in mind that a 4:1 ratio of beam intensity represents a perceived brightness ratio that is closer to 2:1.

      • If you have a (diode) laser pointer of known power and wavelength, it can be used. However, this gets to be complicated if the wavelength AND power differ (as is likely with many laser pointers (670 nm) at this time. In addition, power levels for laser pointers are maximums and the actual power is not generally known.

      Estimating relative power works better on your finger or palm (don't worry, you won't even be able to detect a 5 mW HeNe beam on your flesh from the any heating effect but don't do this with a 20 W argon laser!) in the raw beam than on a white card unless the beam is first spread out using a lens or equivalently and more easily accomplished, you view the spots through a lens to make them appear fuzzy. In either case, the amount of perceived beam spread depends on output power and the difference is much more apparent than just looking at a tiny bright dot.

      Both the perceived brightness AND the size of the spot will vary with HeNe beam power. After a little practice, estimating the output power will become second nature - sort of like recipe measurements: "just use a pinch of salt in the stew!". However, if you have a collection of neutral density filters, you can use these to match brightnesses which may be just a bit more precise! The laser power meter would be even better. :-)

      For relative power measurements, either of the simple laser diode based laser power meters described starting in the sections: Sam's Super Cheap and Dirty Laser Power Meter will actually work quite well. If you can calibrate one of these with a HeNe laser of known power output, better than 5 percent accuracy is easily achieved.

      Just give the laser enough warmup time to stabilize (10 minutes for a small HeNe tube, up to an hour for an 8 foot long SP-125!). See the section: Measuring HeNe Laser Output Power for additional tips.

    Measuring HeNe Laser Output Power

    Follow these steps to reliably measure the output power of your HeNe laser using a laser power meter (either homemade or a real one!):
    1. Set up the laser and power meter so that the entire beam falls on the sensor but covers as much of it as possible (or convenient). For a very small silicon photodiode, this may mean a combination of putting the sensor close to the laser (careful of the high voltage if a bare HeNe tube!) and using a focusing lens to reduce its size (but not so small as to concentrate it in a single dot!).

    2. Allow the laser to warm up. Initial power is generally much lower than after the tube has been on for 10 to 20 minutes (even more time may be needed for a long HeNe tube or large frame HeNe laser to stabilize). The average power may start out reduced by 15 to 25 percent or more and gradually creep up to its final value, often overshooting a bit before settling down. (I don't know whether the overshoot behavior is a fundamental characteristic of a HeNe laser or a symptom of less than perfect mirror alignment. However, I've seen it on too many tubes to be a coincidence.) In addition, for a normal (non-stabilized and non-precision) HeNe laser, expect a short term oscillation in power typically over a 1 to 5 percent range with a period of 10 to 30 seconds due to longitudinal mode cycling - just get used to it! (The amplitude for longer tubes may be lower since a larger number of longitudinal modes can be active simultaneously and their effects average out.) This may continue (though possibly at a reduced amplitude) even after infinite warmup time.

    3. Set the HeNe tube current to its recommended value or adjust it to maximize power output. I have found that the latter current is often lower than the manufacturer's current specifications. With the latter technique, allow time for the power to stabilize after a small change. For example, reducing the current by .5 mA may result in an immediate increase in output power of 5 percent, but this may settle back down and depending on where on the curve you are, may end up lower or higher than the original setting.
    Once the laser has warmed up, you are ready to take a reading. Where the power is varying due to mode cycling, unless you have a data acquisition system and data processing software, the best I can suggest is to eyeball the max, min, or average value of the readings as desired. This really isn't too difficult. :)

    A silicon photodiode or solar cell based power meter is quite linear with respect to laser beam power. For maximum accuracy, subtract or zero out the dark current (with the sensor covered) and locate the sensor far enough from the laser output aperture to minimize pickup of the glow of the discharge (though neither of these is a serious source of error unless you are measuring in the microwatt range).

    For lasers of other wavelengths, sensitivity may be quite different. For example, based on the Typical Silicon Photodiode Spectral Response, it may be 50 percent or more lower at 488 nm from an argon ion laser. For near-IR, silicon photodiodes will probably have a somewhat higher sensitivity which then drops off but is still usable beyond 1,064 nm. However, the exact response curve is dependent on many factors so the power meter really needs to be calibrated for each wavelength.

    Testing of Non-Red HeNe Lasers

    HeNe lasers producing yellow (594.1 nm), orange (611.9 nm), or green (543.5 nm) beams are much more finicky with respect to everything! This is due to the fact that the gain for these lines is much lower than that at the common red (632.8 nm) wavelength. Power supply current, mirror alignment, and even the mounting of longer tubes, must be exactly right or performance may be affected dramatically.

    For visible non-red HeNe lasers:

    HeNe lasers producing IR (1,152.3 nm, 1,523.1 nm, or 3,391.3 nm) shouldn't be nearly as critical, at least with respect to losing the beam entirely, as these have much higher gain than red tubes. However, power output and beam quality could still suffer where the conditions are not optimal.

    For more information, see the sections starting with: Problems with Mirror Alignment and the chapter: HeNe Laser Power Supplies.

    Identifying the Manufacturer of a HeNe Laser Tube

    For the most part, this is a curiosity thing. The future of the Universe won't be affected if you don't know who made your HeNe tube. However, it is sometimes useful to be able to identify the specific model and knowing the maker can help so the dimensions and style can be matched to a catalog or database entry.

    Some physical characteristics of HeNe tubes from various manufacturers are summarized below. Except for Melles Griot, this is from a rather limited sample so just use it as starting point. Unless otherwise noted, mirror mounts are the common 'you bend it' type. Some specific models - usually old, long, or other (than red) color tubes may have actual three-screw adjusters (not locking collars but permanently attached versions of the type shown in Typical HeNe Tube with Three-Screw Adjusters Added). Really old tubes will have mirrors Epoxied to fixed glass or metal mounts with no possibility of adjustment (though for those with exposed bores such as many Spectra-Physics models, very slight distortion of the glass will affect alignment though it's hard to devise a way of stabilizing any improvement.

    In the summaries below, where the HeNe product line of a company has been acquired by some other manufacturer, the original company's name is listed first since that's the one you're likely to see with respect to surplus lasers.

    The tubes from the following manufacturers are really very similar in terms of overall design (though I would assume that company proprietary details vary significantly). Most of the descriptive details are simply here for curiosity purposes and to help identify a totally unmarked laser. See Typical HeNe Laser Head for an example of this tube construction and its mounting in a cylindrical laser head. Tubes using the mirror mounts for both power supply connections are by far the most common for modern internal mirror HeNe lasers:

    HeNe laser tubes of conventional design

    HeNe laser tubes of more interestng but oddball design

    The following are include some tubes based on slightly different architectures:

    Really old HeNe soft-seal tubes may be of all-glass constructions as the bendable mirror mount techique hadn't been invented yet:

    Older large Aerotech tubes had a side-arm for the cathode only, with the remainder of the tube being of conventional design with a large gas reservoir. Really old tubes might have side-arms for both cathode and anode. Older tubes of from various manufacturers were also more likely to have an adjustable mirror mount at one end at least.

    An Assortment of Old HeNe Laser Tubes 1 and and An Assortment of Old HeNe Laser Tubes 2 show a few other styles where the manufacturers have not been positively identified. (They have been arranged with the output facing to the right.) Some additional information, photos, and diagrams may be found in Vintage Lasers and Accessories Brochures with actual examples of many of these in the Laser Equipment Gallery under "Photos of Assorted Helium-Neon Lasers".

    And one truly ancient design some people call the "J-Tube" due to its shape is shown in Vintage J-Tube-Style HeNe Laser Tubes. The neon sign electrode cathode is a tip-off that these are really old tubes, probably from the late 1960s or early 1970s. I doubt any still work today but they do make decent conversation pieces. ;-)

    I'm sure there are all sorts of exceptions and a HeNe tube may appear in style to be a particular manufacturer but could have some other origin (like a foreign clone). In other words, your mileage may vary. :)

    Also see the section: A Far East HeNe Laser Tube.



  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    Internal Mirror HeNe Tube Optics

    Cleaning HeNe Laser Optics

    Fortunately, this is almost a non-issue for internal mirror HeNe tubes as there is only one accessible surface that matters and it is only in the output beam - not part of the laser resonator. Thus, no amount of gunk or dirk on its surface can affect laser action in any detectable way. However, if it is not clean, the output beam may become diffused or distorted. And, eventually the Anti-Reflection (AR) coating and the surface of the glass itself may be etched permanently by finger oils which turn acidic and/or damaged by gritty dirt. Of course, there may be external optical components like lenses, mirrors, and prisms that need to be cleaned and could also be damaged from abuse or neglect.

    Those who maintain lasers professionally will insist on the use of laboratory (gas chromatograph or spectroscopic) grade methanol and acetone. For small internal mirror HeNe laser tubes and their optics, this really isn't necessary. The type of isopropyl alcohol sold in drug stores designated medicinal (91%) is quite acceptable but you will have to gently dry off the cleaned surface - the impurities will result in a cloudy film if just allowed to dry. Even rubbing alcohol (70 percent) will work in a pinch. However, if you are cleaning the mirrors of an external mirror laser, see the section: Cleaning of Laser Optics.

    The surfaces of Brewster windows are somewhat sturdier than mirror coatings but without knowing the precise material, assume they are still relatively soft. When cleaning a Brewster window with the tube powered and aligned (e.g., there is an intra-cavity beam), my criteria for 'clean' is when the scatter off the outside surface is less than or equal to the scatter off the inside (inaccessible) surface. (Scatter here means the fuzzy spot of light appearing on the surface, not the actual reflection.) Unless the tube is damaged or defective, the inside surface should be about as clean as possible!

    Lens tissue is best, Q-tips (cotton swabs) will work. They should be wet but not dripping. Be gentle - the glass and particularly the AR coating on the output mirror surface (and other optics) is soft. Wipe (don't press!) in one direction only - don't rub. Also, do not dip the tissue or swab back into the bottle of alcohol after cleaning the optics as this may contaminate it. The alcohol should be all you need in most cases but some materials will respond better to acetone or just plain water. Just blowing on the surface so it fogs and wiping very gently may help to rid it of the last traces of residue from the alcohol. (Unless you have spectroscopic grade solvents, this latter method is probably best for clearing the dust that invariably settles on the surfaces of glass optics and Brewster windows after a short time, even when exposed to a clean environment.)

    Note 1: The purity of medicinal and rubbing alcohol would appear to vary quite a bit. Some cheap brands are apparently only water and isopropyl alcohol while high priced ones may contain ingredients that will cloud your optics. You may have to try a few before finding one that is fairly pure - or just go for the real stuff. :)

    Note 2: The adhesive used to attach the cotton to the Q-tip stick is probably soluble in acetone and perhaps alcohol. Some of it will then go into solution to collect on your optics. Thus, a Q-tip wet with solvent should be used quickly and only once before being discarded.

    For red (632.8 nm) HeNe lasers, the exterior AR coated OC mirror surface should generally be a uniform blue or purple color when clean. However, I have seen at least one that was greenish. The AR coating on lasers of other wavelengths will likely differ in color, but it may not be obvious, especially for IR (or UV) lasers. About the only thing that can be said for sure is that the color of the faint reflection from the AR coated surface shouldn't include much of the lasing color. And, high quality broad-band AR coatings may come very close to being invisible!

    CAUTION: Don't overdo it - optical components may be bonded or mounted using adhesives that are soluble in alcohol or acetone (but probably not water). Too much and the whole thing could become unglued. I still haven't found the itty-bitty collimating lens I lost in this manner. :-( In addition, any plastic optics may be totally ruined by even momentary contact with strong solvents.

    And, about keeping the inner surfaces of those mirrors clean. You say: "I can't even get to them, being sealed inside the tube. What are you talking about?". Well, while the environment inside the HeNe tube should free of contamination, there can always be little particles of unidentified 'stuff' left over from the manufacturing process. So, while there are generally no restrictions on the orientation of these tubes, it is probably not a bad idea for them to be stored and installed horizontally if possible so none of that 'stuff' can fall on the mirrors. This might be excessive caution but it is usually quite easy and painless.

    Why You Shouldn't Touch the Mirrors or Mirror Mounts

    The laser resonator for internal mirror plasma tubes is totally sealed safely inside. However, there are still some very good reasons to avoid touching the output coupler or pressing on the mirrors and their mounts. Some comments:

    If you accidentally touch either mirror, carefully clean it, preferably with lens tissue and alcohol. See the section: Cleaning HeNe Laser Optics

    Damage to Mirror Coatings of Internal Mirror Laser Tubes

    There are two ways the inside surface of the mirrors can be damaged on internal mirror lasers: by having "stuff" land on top of them or by having the original dielectric coatings come off. (Of course, the AR coating on the outer surface of the OC mirror can be damaged by overzealous cleaning or other abuse.)

    Sputtering Overcoat or Debris on Mirrors

    The most likely cause of damage to inner surface of the HR or OC mirror on internal mirror HeNe lasers is due to sputtering as a result of running the tube with reverse polarity for an extended period of time. Where the anode mirror mount is used as the cathode, sputtering of the metal will actually deposit a nice metallic film on the mirror in close proximity to it. In some lasers, after only a few minutes, the power will drop and lasing will eventually cease entirely. After 30 minutes, the coating may be so thick that the mirror will appear opaque. (For an "other color" tube, particularly yellow and green, it may take only a few seconds to ruin the tube due to the low gain.) Note that strictly speaking, the coatings aren't actually damaged, just covered. So, a suitable acid wash or something might restore them to pristine condition. Unfortunately, this is tough to do on a sealed tube but might be possible if you salvage the mirrors for some other laser project. :)

    Apparently, the careful use of reverse polarity may actually be used by some manufacturers to 'tune' the power output of a HeNe tube. This might be needed to reduce the gain of a 'hot' tube that is lasing on an adjacent spectral line in addition to the desired one. However, I can't imagine any hobbyist wanting to ruin a perfectly peculiar tube of this type or to want to reduce output power on any laser! :)

    There are two ways for reverse polarity to occur depending on the style of the HeNe tube. However, they are both due to carelessness or lack of knowledge:

    1. HeNe tubes where the cathode connection is to the cathode-end mirror mount (e.g, the most common types) - In this case, accidentally reversing the power supply leads will do it.

    2. HeNe tubes where all connections are to terminals at the anode-end (Hughes style) rather than to the mirror mounts - It is easy to not realize or forget that the cathode-end mirror mount isn't supposed to be connected at all since there is a separate terminal via a glass feed-through for the cathode connection at the anode-end of the tube. I've done this a couple of times but caught it before detectable damaged occurred.

    As noted elsewhere, the HeNe tube will appear to operate normally - perhaps it will be even easier to start - but degradation will happen in short order and at that point, your options are quite limited - as in there are none.

    Of course, running a tube on AC will do the same thing and an autopsy of one that had died in this manner showed a clear indication of a dark overcoat on the HR mirror, though it wasn't obvious from external examination.

    A drop in power even with correct polarity and current over the course of several hours may also be a result of sputtering but of the actual cathode electrode once it has lost its "pickling". See the section: HeNe Tube Seals and Lifetime. There is nothing that can be done for this either. However, check for other causes like mirror alignment and improper power supply current before giving up.

    A metallic coating on the inside of the glass anywhere in the tube except near the getter may be an indication that sputtering has occurred. For example, Melles Griot HeNe tube cathodes typically have several holes around their perimeter near the end cap/mirror mount. Metallic spots on the glass at these holes are a definitive confirmation of sputtering and likely means end-of-life.

    Running the tube with grossly excessive current (perhaps 2X optimal or more) may also result in sputtering damage though other things will likely die first like the ballast resistor(s) or power supply.

    In rare cases, a bit of debris may find its way to a most inappropriate spot in the center of one of the mirrors. The unfortunate location is probably not a coincidence as a large electric field gradient is present there due to the high intra-cavity flux. Despite clean-room assembly, foreign objects can find their way inside HeNe tubes! This is why I recommend storing and using laser tubes on their side, not vertically!). A speck of dust in exactly the wrong place can result in an interesting, though perhaps useless, multimode beam. :) Sometimes, careful tapping will remedy the situation. I don't know if other more drastic measures (like blasting with a YAG laser) have a reasonable chance of success. I've heard of a Q-switched YAG laser (SSY!) to work in one instance with a HeCd laser B-window, but great care needs to be taken to only blast the debris, not the window. I was unsuccessful when trying this stunt on a 2-B HeNe laser tube and ended up blasting the window. I've also heard of using a Tesla/Oudin coil on all-glass tubes to give the particle and glass the same charge so they repel, but don't know if that's just a laser legend.

    Mirror Coating Vanishes

    This is a problem I guarantee you won't see everyday, at least not on internal mirror laser tubes. No, I'm not referring to sputtered electrode material or other debris on the mirror but something rather strange. :)

    I was sent a HeNe tube with a hole in the Output Coupler (OC) mirror. OK, it isn't quite a hole in the glass, but the dielectric coating on its inside surface is completely obliterated - as though someone had gone in there with abrasive and removed it - wiped it clean (a beautiful job, I might add!) - but only in the central area (slightly larger than the diameter of the actual bore, about equal to the diameter of the inside of the restricted area of the mirror mount - a coincidence?). And, the Anti-Reflection (AR) coating which is apparently placed under the mirror coating is totally intact (at least that's what it appears to be - there is about the same reflection from the inner surface and the AR coated outer surface).

    I have to say that this is the weirdest thing I've ever seen in some time. (Note that damage to external mirrors, even flaking, isn't particularly unusual depending on the storage conditions or prior cleaning attempts but such damage to internal mirrors is unusual. The second weirdest thing would be that HeNe tube where the discharge changes color from anode to cathode. See the section: HeNe Tube Lases but Color of Discharge Changes Along Length of Bore.) I can't imagine that this effect was a result of natural causes and consider any internal cause to be highly unlikely in any case. The discharge looks normal and the operating voltage is normal similar to that of other identical model tubes. The only conceivable explanation from within is that it was run with excessive current for an extended period of time somehow resulting in ion bombardment (inverse sputtering? - see below for some additional info) of the OC mirror which is at the cathode-end of the tube. I don't even know if this is theoretically possible. Since the HR mirror at the anode-end of the tube is in perfect condition, it isn't likely to be an internal optical effect either (too great a light flux in the resonator) since I would think that would do the same thing at both ends. The fact that the diameter of the clear area is significantly larger than the bore also precludes this possibility.

    Total reflection from the inner and outer surfaces of the OC in the area of the hole is about 2 percent which is too bad. I'd love to try to use this tube with an external OC mirror. However, the total single pass gain of a tube of this length is also only around 2 percent so there would probably be insufficient gain to sustain oscillations. At best, it would be marginal. I initially made a half-hearted attempt to get it to lase anyhow but nothing happened. Later, I did a more careful test with some success - see below.

    I've never ever seen a HeNe tube with any internal damage to either mirror before that could not be explained by it having been terribly abused. Thus, I'm inclined to suspect an external cause. Maybe someone was using it to align a high power Nd:YAG resonator and forgot to remove the tube before firing up the big laser. POW! No more mirror. :) This, however, was denied by the former owner. Other possibilities are that the coating was of poor quality and flaked off on its own (though I could find no evidence of any debris) or that this tube was used as part of another high power and/or invisible laser for aiming purposes and the main beam accidentally made its way back to the mirror by reflection from the work-piece.

    I am attempting to find out more about the history of this tube. So far, what I do know is that it was originally part of a Postal scanner of some sort and was operational when removed from service. At some point between then and now, someone or something went in and did a thorough cleaning job. :)

    FLASH - Some new info: I just discovered that for at least the first 5 minutes of operation from a cold start, the negative discharge may decide to originate inside the mirror mount rather than where it belongs at the cathode. And, it may abruptly switch back and forth at random times. Whether this is due to a broken connection between the cathode and mirror mount (unlikely), depletion of the cathode 'pickling', or that the warranty has expired, I do not know. So, the inverse sputtering theory is back in the running even though it would seem more likely that this would more likely result in a metal overcoat than removal of the mirror coating!

    I have now taken some photos of this tube. See Melles Griot 05-LHP-120 HeNe Laser Tube with Missing OC Mirror Coating. The photo on the far left shows a normal 05-LHP-120 with the weird one sitting next to it. The middle shot is of the that one under power with the discharge to the cathode the way it is supposed to be. The photo on the far right shows the discharge taking place to the OC mirror mount instead - probably due to a bad connection between it and the aluminum cathode can.

    As promised, I did some more experiments in getting the tube to lase with an external mirror. It now produces up to about 0.3 mW acting as a two part resonator containing a low reflectance intermediate mirror. With the wiped-clean mirror properly aligned, the weak modes due to the slight reflection from it (in the original tube) and the extended resonator formed with the external mirror compete with one-another. As the tube heats and expands, the output comes and goes periodically. Pressing gently on the external mirror mount to adjust the length of the total cavity ever so slightly results in very distinct power cycles - the classic behavior of an interferometer. A very cool toy if nothing else. :) For more details on these interesting experiments see the section: External Mirror Laser Using HeNe Tube with Missing Mirror Coating.

    I have found a second tube with a similar electrical problem. The resulting sputtering has indeed overcoated the cathode-end mirror to the point that there is no longer any laser output but the coating hasn't fallen off yet. :) Unfortunately, the discharge doesn't remain inside the mirror mount long enough to try the obvious experiment to see if its coating will eventually flake off.

    And I just came across a tube from an HP-5501A laser where there is a (approximately) 1/2 mm hole in the coating at the exact center of the mirror. The HeNe laser power supply went bad and was pulsing the tube continuously, possibly for hours before anyone noticed. Like the other tube with a similar problem, the only evidence is the missing spot on the mirror coating. The tube looks and behaves exactly as a normal tube should, except that there is now no beam.

    Hewlett Packard 5501A HeNe Laser Tube with Missing Coating in Center of Output Mirror shows the unsightly blemish. It's actually fairly sharp edged but the digital camera didn't know how to focus on it. Another 5501A tube had a hole over 2 mm in diameter - the size of the bore in the spacing rod directly against the mirror! In Hewlett Packard 5501A HeNe Laser Tube Showing Bore Discharge it can be seen that the glow terminates well away from the mirror when running normally. I have no idea what happens during repeated starting. But by eye, there was no visible discharge near the mirror.

    HeNe Laser Torture 1

    OK, send the kids to their rooms, put the dog and cat out in the yard, and blindfold the tropical fish. This is going to be ugly. :) My objective is to see if I can blast a hole in the mirror coating of a hapless certifiably dead HeNe laser tube using abuse that may occur due to power supply/tube incompatibility, power supply failure, or increase of tube dropout current due to high mileage.

    My torture machine is a 12 VDC input HeNe laser power supply brick that has lost its regulation ability, so it basically is controlled by the input voltage, and is generally of little practical use except for these "special applications". If it is killed during these experiments, no one will shed any big tears.

    My experimental subject, err, victim #1, is a well used Hewlett Packard 05501-60006, the HeNe laser tube found in the 5501A two-frequency Zeeman metrology laser. This one probably has seen 50,000 hours, if not double or more, of run time, and is now producing very close to 0.00 mW. But it still starts and runs normally. The tube from the 5501A was selected because (1) I already know that the missing mirror coating malady (MMCM) is, if not common, at least not that unusual as can be seen in Hewlett Packard 5501A HeNe Laser Tube with Missing Coating in Center of Output Mirror, as I've seen at least two of these terminally sick laser tubes. And (2) I have a pile of dead ones and this must be a noble use for one or two, before the organs are harvested! The 5501A lasers are typically run 24/7, often unattended and simply idling away their photons not being used. So, it's possible that rapid restarting or sputtering could be going on for days or even weeks without anyone knowing. The tube in the photo above was in such a laser, though I don't know for how long it was actually being pummeled.

    The 136K ohm ballast resistance normally used with the 5501A tube is replaced with a 20K ohm resistor, preceeded by a 3 nF, 15 kV capacitor. The result is a reliable relaxation oscillator at low drive, though the tube does seem to stay on continuously at higher power. This produces pulses which should have a peak current of 100 to 200 mA. If the damage is done by the peak current, then there should be sufficient abuse to produce mirror damage, though it could still take a long time. However, if something like undershoot/reverse polarity or ringing is required, that will necessitate a more complex device.

    I started running this rig at an input voltage sufficient to produce 10 to 20 pulses per second into the laser tube. Thats 10 to 20 starts per second, which we are taught is supposed to be bad for tubes and power supplies. I intend to continue running like this for at least 24 hours, or until the mirror coating shows signs of disappearing. So far, after a few hours, about the only thing that's changed is that the output power at optimal current (run normally) has increased from near 0 uW to about 60 uW. However, I assume this is simply a result of running the tube, not that I've discovered some elixir of life for old lasers! Even though these are supposedly hard-seal tubes, some will respond to extended run time, possibly dramatically.

    My plan was that if after 24 hours, nothing bad happens, more extreme measures would be implemented, like a larger capacitance or lower ballast resistor or both. However, it may take higher peak voltage to make anything happen. If this turns out to be the case, a hard-start tube will be the next test subject. And, should the tube start outputting rated power, I'll quit while I'm ahead. ;-)

    And, not unexpectedly, it's been run at least 24 hours now and the only change is that the output is over 120 uW when run at 3 mA. It's somewhat hard to tell what the exact power output is consistently because even with the Zerodur bore determining the mirror spacing, there is still some change with temperature, and output power is strongly dependent on where on the gain curve it's lasing. But since this tube works well enough to try in a 5501A, torture operations on it have been canceled.

    My next subject, Tube #2 also was brought in outputting exactly 0 mW, but the complexion of its bore discharge seems a bit different, perhaps more white-ish, which could a more terminal condition. Pulsing it for 10 hours with a 100K ohm ballast resistance had no effect on either output power or the mirror coating. When I went to try use a lower ballast of 20K ohms, it promptly blew up one of the 1 nF, 15 kV capacitors making up the 3 nF energy storage capacitor bank. Whether this is due to an extremely high peak voltage due to a sometimes hard-to-start tube, or the rapid dV/dt (seems unlikely), I don't know. So, I built a capacitor from two sheets of 0.02" thick FR4 unplated PCB material and aluminum foil. This worked out to about 9 nF and should be able to handle any voltage the poor abused HeNe laser power supply brick is willing to produce. The bare wood table, chains and shackles setup can be seen in HeNe Laser Torture Machine 1. Without the restraints, the tube would rip out the connections and attempt to excape. ;-) It makes a satisfying snapping sound when it discharges through the tube. This is probably the aluminum foil vibrating - some of which is even visible where the foil isn't in firm contact with the FR4. I also tried eliminating the ballast resistor entirely resulting in discharge flashes that were nearly white, but the constant screaming of the tube from the pain was unbearable. :) The 20K ohm 1 W resistor did eventually get destroyed by the peak current, so I'm now using a 5K ohm 10 W wirewound resistor that should survive, the discharge is still somewhat orange, and there is only occadional moaning from the tube. I will be exercising the modified torture device shortly. However, it will likely not have any more of an effect than the 3 nF setup unless an unfortunate bug wonders by. I need to figure out how to create a reliable undershoot. A nice high value inductor perhaps.....



  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    Problems with Mirror Alignment

    Checking and Correcting Mirror Alignment of Internal Mirror Laser Tubes

    Note: While written specifically with HeNe laser tubes in mind, the techniques described in the following sections apply equally well to HeNe, Ar/Kr ion, and other sealed laser tubes which have internal mirrors on compliant mounts (those attached to the end-caps via narrow sections of ductile metal tubing).

    Some (mostly older) HeNe and other internal mirrors tubes will actually have adjustment screws as part of the tube assembly. I'm not talking about the locking collars found on many Melles Griot and some other tubes to stabilize the mirrors. See Three-Screw Locking Collars on Melles Griot HeNe Laser Tubes. These may be used for adjustment but are not ideal for that purpose. Rather, some tubes have actual three-screw adjusters where the screws run parallel to the tube's axis and press against an adjoining disk. Selected models from Aerotech, Hughes, Melles Griot, Spectra-Physics, and others have been found to have these. Some like those on certain surplus (Xerox) Spectra-Physics laser heads are quite large with fine control of alignment. If your tube is one of these - and its gas fill is still good - the procedures below for mirror adjustment can be considerably simplified. No special tools will be needed and fine control of mirror angle should be easy to achieve with just a tiny (WELL INSULATED!!) hex wrench. This sort of adjuster can often be added to a modern tube as well. See Typical HeNe Tube with Three-Screw Adjusters Added for an example of one approach.

    Precise mirror alignment is critical to proper functioning of HeNe tubes and lasers in general. For a HeNe tube, the mirrors must be aligned (parallel to each other and perpendicular to the tube bore) to a pointing accuracy better than one part in 1/10th of the ratio of bore diameter to resonator length to achieve optimal performance.

    For a typical HeNe tube, this is one part in 2,500. If the alignment is off by one part in 1,000 (1 miiliradian or 1 mR), there will likely be no output at all. You won't fix this by trial and error! Spherical mirrors may have a somewhat wider range where a beam will be produced but still require precise alignment to achieve optimal performance. Alignment (and nearly everything else) is even more critical for HeNe tubes producing non-red (particularly yellow and green) beams as these have much lower gain. And for these, there may be no way to obtain an optimal alignment if the tube is not inside a thermally stabilized enclosure, or possibly at all.

    I now routinely check mirror alignment on any HeNe laser heads or tubes I acquire by gently pressing sideways on the mirror mount at the cathode (grounded) end of the tube. I may also do the basic "walking the mirror" tests as described in the section: Walking the Mirrors in Internal Mirror Laser Tubes which will identify tubes where the alignment of both mirrors was never quite right (most likely when new from the factory). If I can increase power output by more than about 5 percent in either case after a 20 minute warmup, I will adjust alignment as described in subsequent sections, below.

    Where a HeNe tube produces a weak or low quality beam or doesn't lase at all and no other faults have been identified (such as improper operating current, or problems with the gas fill), mirror misalignment is quite possible. However, it does take effort to mess these up as the mirror mount tube(s) must actually be bent. Casual handling won't do it. It would have had to be dropped or used as a hammer! :)

    Other possible causes of less than perfect mirror alignment include the following:

    Note: For really long high power HeNe tubes (e.g., above 15 mW or so), see the comments in the section: How Can I Tell if My Tube is Good?. Your tube may need to warm up for 1/2 hour or more, or it may require external adjusters permanently installed or you may have it mounted incorrectly. DO NOT attempt to remedy the mirror alignment problems by physically bending the mounts if gently rocking the mirrors (see below) doesn't result in any beam. Your likelihood of success is about the same as winning the State Lottery Super Seven. And if there are flashes from rocking the mirrors, adjustments may not be needed in any case as there may be nothing wrong with the tube!

    There are two types of situations:

    The procedures described below are simplified versions of those that can be used for testing and adjusting of mirror alignment on many types of lasers (including HeNe and Ar/Kr ion lasers where one or both mirrors are external to the tube. See the sections: External Mirror Laser Cleaning and Alignment Techniques, Sam's Approach for Aligning an External Mirror Laser with the Mirrors in Place and Daniel's Method for Aligning External Mirror Lasers. The CORD "Laser/Electro-Optics Technology Series" also has a basic alignmnet procedure outlined in the chapter: "1-7 Optical Cavities and Modes of Oscillation".)

    These techniques are also ideal for use with internal mirror argon ion (blue/green) tubes because a readily available red HeNe laser can be used for testing and adjustment (having a different color laser for the alignment procedure simplies it considerably). Here, they have been adapted specifically for use with small internal mirror HeNe tubes.

    Note: It is assumed that your problem HeNe tube has each of its mirror mounts separated from the end-cap/electrode assembly by a restricted area that is not obstructed. If this is NOT the case (at one or both ends), there may already be a mirror adjusting device permanently attached to the tube and it will have to be used (unless it is removed) rather than the tools described below. In its favor, fine adjustment with such a device is more precise (though it will be less convenient for 'rocking the mirror') and alignment problems are less likely in the first place (unless someone was mucking with the screws!). Note that some older HeNe tubes have absolutely no means of adjusting the mirrors - they are bonded directly to the end-cap(s) or glass tube. In that case, best to move on with your life. :)

    Rule #1 of mirror alignment: If it's lasing at all, NEVER EVER allow it to lose that beam entirely without remembering exactly how to get it back! If alignment is lost at both ends of the laser, your job is orders of magnitude more complex than fixing alignment at only one end!

    Minor Problems with Mirror Alignment

    A beam which is much less intense than expected or distorted (not circular with a reasonably smooth gaussian profile for a TEM00 laser) may be due to bad mirror alignment. Checking mirror alignment as long as the laser produces some sort of beam is easy and very low risk. Correcting it may be possible as well.

    If there is no beam at all at the nominal tube current but no evidence of bent mirror mounts or other visible damage, this technique may also be used with care to see if one of the mirrors is SLIGHTLY misaligned. However, if gentle rocking of the mirror mount does not result in a beam (see below), DO NOT attempt to actually bend the mount since there is no way of knowing in which direction the correction (if any) is needed. See the section: Major Problems with Mirror Alignment.

    Despite all the "CAUTIONS" in the following sections about the sky falling if you mess up, don't be too timid about checking and adjusting the mirrors on lasing but weak HeNe laser tubes. If they are not doing the rated power after warmup, the most likely cause is mirror alignment, not age or use. On average, I'd say about 2/3rds of the red HeNe lasers I've gotten surplus (including eBay) could be tweaked up to rated power or above with just alignment of the output-end mirror. As long as you don't lose the beam entirely, it's a fairly low risk effort with potentially high reward. However, before attempting this on a valuable high power tube, practice with junk tubes first. Just keep in mind that the required change in mirror orientation is essentially undetectable to the human eye so always err on the low side. And, some type of laser power monitor is extremely desirable to be able to see small changes in output. A solar cell or photodiode and DMM is perfectly adequate.

    Quick Check of Mirror Alignment

    Here is the instant version of this procedure for quickly checking to see if one mirror mount is slightly misadjusted: (The more complete version follows it.)

    1. Take a sheet of paper and roll it up into a tight tube that just fits over the mirror mount extensions. Wrap adhesive tape around the end you will be using on the tube to prevent it from loosening. This is your mirror rocking 'tool'. (You may need two such tools as some tubes have different diameter mounts at each end.) The nice thing about this tool is that it is virtually impossible to either permanently bend the mount accidentally or to damage it as long as you are reasonably gentle in what you do. Another benefit is that paper is a good insulator so you won't have a shocking experience! For a paper tube of about 6 inches, a couple of pounds of side-ways force can be applied safely on most tubes.

    2. Power up the tube and while holding it steady (careful: high voltage!), use your instant tool to gently rock the mirror mount at each end. If a tube with no output produces a beam for some particular amount of pressure and direction, you know it needs mirror alignment. Similarly, if doing this can increase the output power substantially, one or both mirrors also need alignment. With any luck, only one of the mirrors is very slightly misaligned. However, even if it appears to be the first one you tried, the other may actually be the problem. Therefore, don't jump to conclusions and attempt to correct it just yet.

    If the preceding tests show that alignment is needed, read the following sections for instructions on exactly what to do next.

    Minor Mirror Alignment Procedure

    Here is the more detailed procedure for checking and correcting minor mirror alignment problems:

    1. A (homemade) tool for rocking (and possibly adjusting) the mirror mount is needed. See the section: Means of Adjusting HeNe Tube Mirrors.

    2. Mount the HeNe tube in such a way that the ends are free and clear so the adjuster(s) can be used without interference.

      • For short HeNe tubes (perhaps 15 inches or less), a suitable support is a pair of wooden V-blocks clamped to your workbench. Secure the tube with tape or Velcro straps.

      • For longer HeNe tubes, limit support to exactly two locations at or near the ends. This will permit you to press gently sideways in the center from any direction to very slightly bend the entire tube (GENTLY please!). There will very likely be a noticeable effect on output power especially for tubes over 24 inches in length (even to the point of restoring some sort of beam to a tube that wasn't lasing at all). By determining which direction results in the greatest effect can help to identify how the mirror(s) should be adjusted to restore power permanently.

    3. Start with the cathode-end (arbitrary choice) and power up the HeNe tube at the optimal operating current. If you are using a plate or tube type adjustment tool, take care to assure that it doesn't extend beyond the reduced diameter section of tubing and only applies force to the metal of the mirror mount, not the mirror or its frit seal!

    4. Allow the tube to warm up and stabilize a few minutes before checking or attempting any adjustment of mirror alignment. There is a chance that during this time, the output power will increase to a normal (expected) value. (And then you can see if it can actually be improved.) Where there was no beam at all initially, one may appear (weak or otherwise) at some point during the warmup period making the problem orders of magnitude easier! The longer the tube, the more time is needed for the output to stabilize. A rough guideline is 1 minute for each inch of tube length. I just pulled this 'guideline' out of thin air but it's probably better than nothing! :)

    5. GENTLY rock the tool back and forth as you watch the beam's reflection from a white surface. Do this in X (horizontally) and then in Y (vertically). Go easy! It doesn't take much force to change alignment through the entire range that matters - perhaps a few ounces at most. Even modest finger pressure will do it - but only try it this way at the cathode-end and only if the cathode is grounded! Don't get carried away and actually bend the mount at this time - or break the seal. :-(

      What you should see is the beam power (brightness) pass through a maximum and then diminish on either side of this point. Testing is best done with a laser power meter but one of your eyeballs (or both of them) will work well enough for most purposes.

      • If the maximum is at the relaxed position in both axes (you can try the 45 degree ones as well to be sure), mirror alignment is correct (or at least close enough that the chance of being able to improve it without using more sophisticated adjustment equipment and a laser power meter is vanishingly small.)

      • If the maximum (or any beam at all) is off to one side, you can VERY CAREFULLY try bending the mount permanently to attempt to correct it. First, determine the angle of the tool that results in the least force being needed to correct the alignment.

        CAUTION: The mirror mount is ultimately attached to the glass envelope of the tube. The glass-metal seal may not be that strong. Don't get to carried away! With care this adjustment should be possible - barely. :-)

        Note: Where the maximum intensity results with the mirror very slightly deflected, it is possible that the mirror alignment at the opposite end of the tube is actually to blame and you are simply compensating for its pointing error. Thus, it is better to check the mirrors at both ends of the tube before attempting to adjust either of them. However, the only way to be sure is to measure the maximum beam power AND and also examine the shape of the beam. It should have a circular cross-section, a Gaussian profile, and not have any off-axis arcs or other artifacts) when both mirrors are precisely parallel to each other and perpendicular to the bore of the tube. (Note: Don't confuse a weak spot or spots off to one side due to 'wedge' of the OC mirror with an alignment artifact.)

        • Increase the force gradually until you have a feel of how much it takes to actually deform the mirror mount. Even a significant pointing error will only require a nearly microscopic correction. The change in mirror mount angle that you need to achieve is likely to be a fraction of a mR - less than 1 part in 1,000! Not easy.

        • Approach the desired deflection in small increments and overshoot just enough so that the mirror mount springs back to the optimal position. Avoid repeatedly bending it back-and-forth or you will eventually be using the HeNe tube as a high-tech wall hanging :-(. Depending on the type of adjustment tool used, a bouncing or rocking motion of the tool may result in better precision than simply leaning on it.

        • Once optical output is maximum and this point is *roughly* centered when testing by rocking the mirror mount, pat yourself on the back and consider it as good as it gets. Don't push your luck!

      • In the case where there was no beam at all, if GENTLE rocking doesn't result in a beam at any position, DON'T press your luck at this end! The misalignment may be too great for this approach or the problem may be with the mirror at the other end of the HeNe tube!

      • In the unlikely case where you overshoot and the beam disappears entirely, don't panic (yet). Immediately mark the orientation of the movement so you can try to go in the reverse direction. Unless the change was large, it should be possible to get back to a lasing state relatively quickly. Unfortunately, it sometimes takes much longer.

    6. Gently remove the tool (if relevant) without applying excessive force to the mirror mount.

    7. Repeat this procedure for the anode end of the tube - just be careful not to touch the high voltage!

    Alignment should now be the best that is possible by adjusting the mounts at each end independently. Confirm by rechecking it at both ends and making any very *slight* adjustments that may be needed. This is where the addition of permanently installed adjusters may be desirable. See the section: Home-Built Three-Screw Mirror Adjusters for Internal Mirror Tubes come in handy for tweaking but these may be overkill for inexpensive HeNe tubes.

    However, although the mirrors will be parallel to each other (ignoring the mirror curvature), their central axes may not be aligned with the bore. Thus, power output could still be low - possibly quite low. If only one mirror mount was messed up originally and that is the one you touched, the chance of there still being major problems is small but I've seen many supposedly healthy HeNe tubes where mirror alignment was far from optimal even from the factory!

    If you really want to fully optimize power, you will need to go through the procedure discussed in the section: Walking the Mirrors in Internal Mirror Laser Tubes. The use of the three-screw adjusters is definitely recommended if going beyond this point!

    PERFORM ANY ADJUSTMENTS ONLY AT YOUR OWN RISK! Checking the alignment by gently rocking the mirror(s) is safe and effective. However, actually bending the metal is much more difficult and likely to result in death to your HeNe tube. The required pointing accuracy of much less than 1 mR is not much to fool with! If the brightness change that is bothering you is just barely perceptible or you just *think* that it may not be perfectly centered, LEAVE THE MIRROR ALIGNMENT ALONE! Plexiglas or wood plates (even with any inserts) and plastic tubes are really too soft for precise control beyond the elastic limit (i.e., when actually bending the metal permanently). Your control will be poor and you will be much more likely to bend the mirror mount far off to one side never to work again or break it off completely. The lever type adjusters can be more precise but may result in excessive stress to the mounts if used to make more than very small adjustments since it applies an unbalanced force spreading the mirror mount and end-cap apart.

    Note that with some tubes - generally longer ones that were obtained surplus - there may be no way to achieve truly optimal mirror alignment. See the section: Inconsistent Behavior of HeNe Laser Alignment.

    Means of Adjusting HeNe Tube Mirrors

    Unlike the mirror mounts on high quality lab or industrial lasers, those on inexpensive sealed HeNe tubes are integral to the metal end-caps and not normally considered user adjustable. There is usually a restricted region separating the mirror mount itself from the end-cap and actual deformation or bending is required to alter mirror alignment. As expected, there is no easy way to do this without the assistance of some sort of (homemade) tool - and repeated flexing of the metal tubing beyond its elastic limit WILL result in eventual failure of the seal.

    You cannot just grab the mirror mount in your hand and deform them as though your are Superman (unless you are) since additional leverage and finer control is needed (not to mention the several kV that may be present at one end of the HeNe tube end at least!).

    Here are some suggestions for easily fabricated tools or adapters which will permit fairly precise movement of the mirror mounts. The "plate" and "tube" types are best for 'rocking the mirror' to check alignment without changing it. The "lever" type may be more precise for making initial adjustments since it applies force at the exact place that it is needed. The "three-screw" type is unsurpassed for making fine adjustments in alignment without any risk of permanently ruining the mirror mounts by bending them too much. The "collar" type (Three-Screw Locking Collars on Melles Griot HeNe Laser Tubes) is useful for stabilizing alignment but can be used for final tweaking as well.

    Of course, a nearly infinite number of variations on all of these schemes are possible. However, Vice-Grips(tm) (despite being suggested by a person who should have known better), wrecking bars, and 12 pound hammers are NOT appropriate tools for adjusting the mirrors on HeNe laser tubes (or any other lasers, for that matter)!

    CAUTION: For all of the tools, make sure that, pressure is ONLY applied to the tube of the mirror mount beyond the narrow section - not the part attached to the body of the HeNe tube, or the glass or frit seal of the mirror itself. And, don't go overboard - the amount of force needed isn't that great if applied at the appropriate place in the proper direction. Someone I know ("Dr. Destroyer of Lasers") ruined a possibly salvageable large green HeNe tube from overzealous attempts at alignment by cracking the cathode-end glass-to-metal seal. It is especially important to avoid applying any pressure to the mirror glass (which is quite soft) or the glass frit (glue, glass 'solder') holding the mirror in place which is even softer. On some HeNe tubes, there is just a thin ring of this material and it can be easily fractured. I've done it, hisssss. :-(

    CAUTION: DO NOT use a metal (conductive) material for the tool as the mirror mounts probably connect directly to the high voltage power supply!

    Providing two such tools - for both the cathode and anode ends of the HeNe tube, may simplify some of the alignment procedures. This will also be required if the diameters of the mirror mounts at each end of the tube are not the same.

    Alignment jigs may be used in the factory during tube manufacture but these are made from strong rigid components so that even the smallest adjustment of the thumbscrews actually gets transmitted precisely to the mirror mount. Anything as complex as this is overkill for checking mirror alignment but might be desirable to permit fine tuning while the laser is operating.

    Major Problems with Mirror Alignment

    Where the mirror mounts are obviously bent or damaged, or if the techniques described in the section: Minor Problems with Mirror Alignment don't result in any beam, the HeNe tube starts normally, the power supply is providing approximately the nominal current, and other problems have been ruled out, further testing must be done 'off-line' - not powered. This requires a second (working) laser or special optics. Without one of these, anything you do will be hit or miss (mostly miss) as the mirrors must be nearly perfectly aligned before there will be any output beam at all. What is more likely to happen is that you will end up breaking the seal from repeated bending of the mirror mount.

    If only one mirror is actually misaligned, you can use the procedures from the section: Minor Problems with Mirror Alignment to identify the error (by rocking the mirror and looking for a beam with power on) and then carefully tweaking its alignment. In any case, this should be attempted first (unless you are sure both ends or misaligned).

    Where the mirrors at both ends of the tube are messed up, the chances of ever getting a beam with any testing of this type is quite slim - especially for those high power expensive HeNe tubes. Getting close won't be good enough since rocking either mirror by itself will never result in any beam.

    Unless your baby is a high power and/or expensive HeNe tube, it may not be worth the effort to attempt the procedures described below. While testing and/or correcting major mirror alignment may represent an irresistible challenge, the cost in terms of time, materials, and frustration could prove to be substantial. And, as noted, those longer tubes are exponentially more difficult to align! For anything longer than 8 or 10 inches, your odds of success are probably better in your State's Lottery - and then, when you win, you could just buy a new tube! :)

    As if this isn't enough, if one (or both) of the mirrors on your HeNe tube are not planar (often concave at the high reflector end), or there is an internal Brewster plate or etalon, even more care will be required in equipment setup and subsequent steps may be complicated at that end at least.

    In addition, the output-end (output coupler or OC) mirrors on some lasers have faces which are ground with some wedge and thus their surfaces are NOT quite parallel. This eliminates all ghost beams that are parallel with the main beam (though there will be one or more weak ghost beams off to one side) and also minimizes reflections back into the resonator. Alignment is complicated for a mirror where wedge is present due to non-parallel reflections and slight refraction through the mirror. I don't know how likely wedge is with small internal mirror HeNe tubes but check for it in any case before considering attempting alignment of a non-lasing tube (see the section: Ghost Beams From HeNe Laser Tubes). Wedge is common in large frame HeNe lasers with external mirrors.

    The longer the HeNe tube, the worse it gets!

    I would suggest that if the tube is valuable enough to warrant the expense, see if one of the HeNe laser manufacturers or laser system refurbishers will perform the alignment for you. The ratio of their probability of success compared to your probability of success will approach infinity. OK, perhaps not quite infinity. It probably won't be significantly greater than the ratio of the mass of the Sun to that of a typical electron. :-) I have no idea if this is a viable option or what it might cost.

    Having said that, if you are still determined to proceed, alignment is best done with a working narrow beam laser (i.e., HeNe, argon ion, etc.).

    If you do not have a working laser to use for this purpose, various plans for construction of laser mirror aligners using simple optics and readily available materials are provided in: "Light and Its Uses" [5]). However, some of these are for wide bore tubes and may not work well with the 0.5 to 1.5 mm bores of typical modern HeNe tubes.

    If you have another functioning HeNe laser or tube (you can use the power supply for the one you will be adjusting since it will not be needed until the mirrors are roughly aligned), or possibly even a collimated diode laser or laser pointer) it may be possible to use it as an alignment laser to adjust the mirrors. A low power (i.e., .5 to 1 mW) laser is adequate and preferred since it will be safer as well.

    The general idea is shown in Principle of Mirror Alignment Using Reflected Beam. With the beam of a low power Alignment Laser (A-Laser) and the bore of the Tube Under Test (TUT) are lined up, mirror alignment will be perfect when the beam reflected from the inner (active) surface of the TUT mirror facing the A-Laser is centered in the aperture of the A-Laser (AL-Aperture) and/or the hole in the Bore Sight Card (BSC) next to the TUT. The diagram shows a TUT mirror mount that is bent at an angle much much greater than anything you should EVER encounter!

    Plan on spending a lot of time on this. Therefore, select a location to work where you can spread out and won't be disturbed for hours. The kitchen table is probably not appropriate!

    The following three steps, (3) through (5), may need to be repeated for the High Reflector (HR - fully reflecting mirror) and Output Coupler (OC - beam output) ends of the TUT. If you find a problem at one end and think you fixed it, you can try powering up the tube to see if a miracle occurred before repeating the procedure for the other mirror. :) You can start with either end of the tube if you have no idea of which mirror might be messed up.

    If you are using a non-red laser (e.g., green argon ion) it may be possible to get a clean reflection all the way back in from the far mirror (the HR if the OC is facing the A-Laser). If so, everything should be done without changing TUT position. This is the preferred way of aligning any laser since correct allignment can pretty much be assured by getting the A-Laser beam to bounce up and back inside the tube just as the photons will do when the laser is operating normally. However, with the closer mirror in place, this can be very difficult, confusing, and time consuming unless everything is bolted down rigidly. And, even then, may be virtually impossible due to the many confusing reflections. It is trivial (well, almost trivial!) for lasers with removable mirrors but you don't have that luxury. :)

    So, even if you are using different colored lasers, since you really can't remove - and shouldn't really even move the mirror facing the A-Laser, the reflected strong spots from its surfaces will likely totally obscure the much weaker return from the far end - even if it was aligned perfectly. It might be possible to just deflect it slightly - just enough to move the obscuring spots out of the way. This is easy and safe to to do with those tubes having built-in three-screw adjusters or three-screw locking collars but should probably be avoided where it is necessary to bend the mount unless you can provide a jig (like an adjuster or collar) to just deflect it slightly and temporarily.

    See the section: External Mirror Laser Cleaning and Alignment Techniques for more information - at least to get the general idea. Some changes and simplifications will be required. Also see the section: Daniel's Method for Aligning External Mirror Lasers since this was written specifically for HeNe lasers.

    If what you have is a tube with an internal HR mirror but external adjustable OC, that procedure (also with slight modifications) will be more appropriate.

    However, when using a red laser to align a red HeNe laser (or any time the A-Laser and TUT are similar color lasers) not enough light can pass through the mirrors to get a return spot - it is too small by a factor of 10,000 or so! (In addition, even if you are using different colored lasers, since you really can't remove - and shouldn't really even move the mirror facing the A-Laser, the reflected strong spots from its surfaces will likely totally obscure the much weaker return from the far end - even if it was aligned perfectly.) Assuming this is what you are doing, the procedure will have to be repeated after reversing the TUT end-for-end. This is what is addressed in the remainder of this procedure.

    If mirror alignment was your problem (and for larger tubes, if you believe in minor miracles!), the TUT should hopefully now produce at least some output beam when powered up.

    In either case, see the section: Checking and Correcting Mirror Alignment of Internal Mirror Laser Tubes.

    Walking the Mirrors in Internal Mirror Laser Tubes

    If your laser tube produces any sort of beam and the alignment of both mirrors independently is optimal (either after testing and/or after correcting it as described in the sections starting with: Problems with Mirror Alignment, it's time to either be happy with what you have, or attempt to optimize the output power and beam quality by adjustments to both mirrors.

    Indications for the need of further alignment include:

    See Effects of Walking the Mirrors for an exaggerated (hopefully!) illustration of why this happens. As can be seen, although the mirrors may be perfectly parallel to each other and there is still some output, by not being aligned with the bore/capillary, portions of the beam are cut off, less than the full amount of gain medium participates in the lasing process, and there can be reflections from the walls and other structures in the tube to create artifacts.

    For external mirror lasers with fine adjustment screws on the mirror mounts, the "Walking the Mirrors" procedure isn't really at all difficult: Both mirrors are moved in the same small increments using the micrometer screws (so they remain parallel), first in X until power is maximized, then in Y, and then back and forth optimizing each direction until no further improvement is detected. This aligns the mirrors so they are precisely perpendicular to the bore. Your typical obsessive-compulsive laser physicist type spends his/her life playing with these knobs. :) See the section: Walking the Mirrors in External Mirror Lasers for more info.

    For an internal mirror laser tube without screw adjusters, a modified approach must be used. I will tell you up front that this is a royal pain and is most easily done if you have three hands (or at least a rigid means of mounting the laser tube and the proper tools). But it can be done and for some cases - most commonly where a tube is marginal to begin with due to age or use, or where someone else, (of course)! has played with the alignment - the improvement in performance (power output and beam quality) over adjusting the mirrors independently may be quite dramatic.

    For all measurements of output power, a laser power meter is highly desirable. It doesn't need to be fancy since maximizing power is what's important, not an accurate value. And, an analog meter (one with a needle!) is usually far superior to a fancy digital readout for this purpose since it responds faster and is easier to interpret using the mediocre processing power of the human brain. Anything that will convert photons to a meter reading will be fine including the absolutely trivial ones described in the sections starting with: Sam's Super Cheap and Dirty Laser Power Meter.

    It's just that your basic allotment of eyeballs isn't very good at detecting small changes in intensity! :) Note that mode cycling of your HeNe tube will result in small variations in output power - these can be annoying but need to be mentally discounted in determining the maximum power output readings.

    Tubes that are marginal due to age or use and output a very weak beam seem to benefit the most - a 200 percent or more boost in power is quite possible (though it will still likely be less than their ratings when new). This is probably because the gain is lower and therefore mirror alignment becomes even more critical. The same alignment errors might only result in a 10 or 20 percent reduction in power for a tube in good condition. And the misalignment might have always been present - factory quality control isn't perfect and tubes would be considered good enough as long as their catalog ratings are met or exceeded when new.

    I have improved the performance of several internal mirror HeNe tubes using these techniques. One of these is discussed in the section: Strengthening a Weak Siemens HeNe Tube. Another was a cute little Melles Griot 5" HeNe tube which was only putting out .1 mW. It's output was boosted to about .3 mW by walking the mirrors (still less than the .5 to 1 mW for other similar size tubes). Some additional improvement might be possible with more work.

    Simple Adjustable Optics Platform

    For simple alignment checks, an adjustable platform can be constructed in about 10 minutes from common materials. This isn't quite as precise as an expensive lab jack or the all-metal version shown in: Alignment Laser Three-Screw Platform. However, for many purposes, it may be good enough. Of course, there's no law against making the design below out of metal! The same basic approach can be used for either the alignment laser (as shown in the photo) or the mounting for the Tube Under Test (TUT). From my experience and that of others, the latter arrangement is more intuitive for aligning the TUT's bore to the A-Laser's beam. However, it may be difficult to prevent the platform's position from shifting while checking or adjusting the mirrors compared to a mounting with a fixed base that can be clamped to the workbench. And, easily accommodating multiple size TUTs would be more of a challenge.

    Also see the section: Sam's Eazalign(tm) Internal Mirror Laser Tube Alignment Platform since your internal mirror laser tubes may deserve only the finest in alignment equipment! :)

    You will need:

    There you have it! This will permit a laser or other optical system secured to this platform to be adjusted easily for height, pitch, and yaw. If you need more of a height adjustment range, sit the platform on something else.

    While this isn't quite as precise as one milled out of a solid block of high strength (aircraft quality) aluminum alloy using anti-backlash spring-loaded micrometer adjustment screws, it will suffice for many purposes and costs next to nothing!

    Sam's Eazalign(tm) Internal Mirror Laser Tube Alignment Platform

    Note: It is recommended that the sections starting with: Checking and Correcting Mirror Alignment of Internal Mirror Laser Tubes be read first so that the various terms, alignment techniques, and risks are understood.

    OK folks, this is what you have all been waiting for. :) The ULTIMATE in precision and convenience. Well, sort of, at least if you construct and use it with reasonable care. Depending on the length of the platform, almost any size tube can be checked for alignment or realigned quickly and easily. Does this sound like a sales pitch yet?? :) The Eazalign platform combines the Primary Alignment Laser (PA-Laser), adjustable Tube Under Test (TUT) mount (ATM), optional Bore Sight Mounts (BSMs), and Far Reflector (FR) mirror or Secondary Alignment Laser (SA-Laser) into one handy (but not so compact) package. ;-)

    This approach should also do a decent job with those annoying curved mirrors since the same reference is used at both ends of the TUT without the need to remove and replace it.

    In addition to supporting the various alignment techniques discussed previously, the Eazalign platform adds a way of providing a return beam so that both mirrors can be checked and aligned in place (without turning the tube end-for-end). For small to medium size HeNe tubes (up to 10 inches or so) using a different color A-Laser (e.g., green HeNe to align a red HeNe, this can be accomplished without the need for a second A-Laser by using a flat first surface mirror (the Far Reflector or FR) on an adjustable mount set up to return the A-Laser beam precisely back to its output aperture

    However, for longer tubes or where the PA-Laser is the same color as the TUT and it isn't possible (at least in finite time) to get a clean beam down the bore, the use of SA-Laser will be needed. Where the Bore Sight Method is used to align the TUT to the A-Lasers, the A-Laser colors won't matter.

    The basic setup is depicted in Eazalign Internal Mirror HeNe Laser Tube Mirror Alignment Platform and consists of the following components:

    The PA-Laser and SA-Laser are set up to their beams are precisely aligned with each-other. In other words, the beam from the PA-Laser is centered on the SA-Laser's output aperture and vice versa. If adjustable mounts are not used for both lasers, this can be done with shims. Where the FR mirror is used in place of the SA-Laser, it is adjusted so that the return beam is precisely centered on the PA-Laser's output aperture.

    To use this system:

    That's it! Since only one reference is used (the precisely centered beams of the PA-Laser and FR mirror reflection or SA-Laser) and the TUT is never moved once it is installed on the ATM, this should result in very precise alignment in a short amount of time.

    Initial Alignment and Tweaking of Three-Screw Adjusters

    These are the type found on some HeNe tubes which are way too small to really provide fine control. See Three-Screw Locking Collars on Melles Griot HeNe Laser Tubes. However, they do work well enough with a bit of care and a lot of patience! :) Home-built adjusters can be constructed somewhat larger so the adjustment range will be greater (see the section below).

    Realize that the 3 adjustments are not really independent since each uses the other two as the pivot. However, the net effect is fairly predictable.

    CAUTION: Don't get carried away while turning these screws - it is possible to rip the mirror mount off the tube! The entire adjustment range is less than 1 turn of each screw once they are snug.

    Any of the procedures and setups described above can be used to determine when the mirror is properly aligned. For longer tubes, I recommend the Instalign technique be attempted first. With care (and a bit of luck), this will get you to a lasing state without the requirement for fancy alignment platforms and jigs.

    The following assumes that only one end of the tube is misaligned. Where both ends of the tube are messed up or in an unknown state, your task is just that much more challenging! :)

    Home-Built Three-Screw Mirror Adjusters for Internal Mirror Tubes

    It is possible to construct mirror mount adjustment assemblies operated by thumbscrews or set screws to correct or optimize the alignment of HeNe and other internal mirrors laser tubes with the common mirror mounts consisting of a pair of tubes separated by a compliant narrow section.

    For a small misalignment, this would avoid the risks of actually trying to bend the mounts since the range of motion would still be within the elastic limits of the metal. This type of adjuster is really best for fine tweaking where a beam of some sort is already being produced, not for initial alignment where the mount is bent at a visible angle! And, as noted in the section: Checking and Correcting Mirror Alignment of Internal Mirror Laser Tubes, some (mostly older) HeNe tubes have these built-in for both mirrors or possibly just the OC.

    See Typical HeNe Tube with Three-Screw Adjusters Added for an example of one approach. With even very basic machining skills and a little scrap metal, a set of these should be very easy to fabricate.

    I would recommend 1/8" to 3/16" brass or mild steel for the plates. (In some cases, the inner plate can just be replaced with a protective metal or plastic washer bearing against the metal end cap of the tube.) Aluminum would probably be acceptable as well but might deform or wear too easily if the adjustments get much of a workout. :) (Even acrylic plastic (Plexiglas) or other non-metal material might be adequate for minor corrections and in addition to easier machining, have the benefit of being an insulator!) Drill the holes in the center of the plates (preferably reamed to the correct size) to just fit over the two sections of the mirror mount(s) and use Epoxy or another adhesive to secure them in place. A clamp arrangement would also work (and permit easy removal of the adjusters in the future) as long as it is designed so that there is no tendency to deform the tubes of the mount and not tightened excessively - which could ruin your whole day by cracking the glass-to-metal seal(s). Use fine-thread set-screws with rounded tips.

    The adjusters should be firmly attached (glued with Epoxy or carefully clamped as noted above) to the HeNe tube end-caps or bolted to a rigid baseplate. (However, in the latter case, expansion of the HeNe tube as it warms up will complicate matters.) They could be left permanently in place applying the proper force to the mirror mounts to maintain mirror alignment and always providing the option of making fine adjustments at any time if needed. with these built-in for one or both mirrors.

    When actually adjusting the mount, no single screw should be so excessively tight that there is a chance of liberating the mirror mount from the rest of the tube! The entire useful range is only a fraction of a turn of each screw. If you are headed that way (or the mirror mount is sitting at a 20 degree angle), it will need to be moved into a initial position (using one of the other tools described above) before the three-screw adjuster can be safely used. Adding a tiny drop of penetrating oil to each of the screws and its contact point will minimize the tendency to of the screw to 'stick' thus easing adjustments. Once you have found the best setting, incrementally snug up each of the screws so they are all applying at least a little pressure to the mirror plate. This will maximize long term stability.

    CAUTION: Use a well insulated tool (hex wrench) to avoid a shocking experience!!

    However, the use of such drastic measures may be gross overkill for use with these small inexpensive HeNe tubes unless you have a machinist sitting around with nothing to do. :)

    Sam's Instalign(tm) Procedure for Internal Mirror Tube Mirror Alignment

    The following is a quick way of checking and correcting the alignment of internal mirror HeNe and Ar/Kr ion laser tubes if the bore is precisely centered with respect to the envelope. Centering is usually pretty good on newer tubes but I've seen a some older ones where it was off by a good fraction of a millimeter.

    The idea is to roughly center the beam of an Alignment Laser (A-Laser) in the bore and then rotate the Tube Under Test (TUT) to check for wobble in the reflection back to the A-Laser's aperture. Where the bore is centered, any shift in the position of the reflection will be due to the misalignment of the mirror.

    I've used this approach successfully on HeNe tubes up to about 20" in length (so far). It is particularly effective where only one mirror is misaligned since the result will be close enough to use minimal force in rocking the mirrors to find the precise setting. Add-on adjusters or locking collars will then be sufficient to fine tune it.

    Sam's Approach for Aligning an External Mirror Laser with the Mirrors in Place

    Here is a quick procedure for aligning an external mirror laser where the area between the mirrors and Brewster windows is at least a couple of inches long and accessible.

    As with most of the other techniques, this one requires an Alignment Laser, preferably a well collimated HeNe laser of a different wavelength than the laser being aligned but a same color laser can be used though the transmitted beam will be much weaker.

    Using this approach, I can replace and align either mirror in a low gain HeNe laser with a 15" resonator (see HeNe Laser Tube with Two Brewster Windows Mounted in Home-Built Resonator) in under 5 minutes assuming the A-Laser is still in place and aligned with the LUT's bore.

    Quick Course in Fine Tuning a Large Frame HeNe Laser

    This is an abbreviated version of the procedures discussed in the section: External Mirror Laser Cleaning and Alignment Techniques. It applies to external mirror HeNe, Ar/Kr ion, and other similar gas lasers.

    (From: Steve Roberts (osteven@akrobiz.com).)

    Get an analog power meter in front of the laser before doing this. You must be able to see the changing trends in the power output. This assumes clean optics and a good tube at proper current levels.

    WARNING: This procedure is not for the timid, easily distracted, or faint of heart!

    When tuning a laser, you work with either the verticals or the horizontals, but never both at the same time. Failure to do this makes it easy for you to misalign yourself into non lasing in a fraction of turns on the adjustments.

    Start with the verticals, pick a direction for the front screw to turn, either left or right, detune the laser power by about 30% , then go to the rear VERTICAL screw and peak the power, Leaving the HORIZONTAL screw UNTOUCHED. If the power is greater after peaking then before, keep going in the same direction till it falls off then go back to the peak, and then keep going the other direction, doing the same simple process of slightly detuning, peaking and measuring. If you write down your power meter readings, you will get the idea and will be able to find the vertical sweet spot. You are scanning the cavity lasing path across the bore.

    Then do the horizontals, same procedure, pick a direction, slightly detune, peak with the other mirror, etc.

    Note: this is the short version of this. On most larger lasers, you would move the front and rear mirrors the same direction by about the same amount. However since this is a large frame HeNe, different rules apply, It usually takes a large frame HeNe about 2 to 3 minutes to settle down after a adjustment is changed, keep this in mind and go slow.

    This is a iterative process, you have to repeat the steps many times on both the horizontal and vertical axis till you have the exact peak, if you have one of the many lasers (e.g., argon lasers like the Lexel, Ionics, 60X) that stresses the Brewster stems when adjusting the mirrors, make sure the Brewster covers are off or relaxed.

    Digital power meters take too long to update when tuning a laser, making it easy to scan past the exact peak. and you can't see which way you are going, it is very important to use a wide scale analog meter.

    Daniel's Methods for Internal Mirror HeNe Tube Mirror Alignment

    (From: Daniel Ames (dlames3@msn.com).)

    The following notes are what I observed and used successfully for aligning a 5 to 7 mW red HeNe laser with another red HeNe laser. (In addition to the procedure that follows, there is a simpler one using three HeNe tubes - (1) a laser to produce a beam for alignment, (2) the tube needing alignment, and (3) an identical tube which is used for setting things up. See: Now for the Quick and Easier Shortcut.

    The Basic Procedure

    Last night I was trying to fine tune the HR mirror on a 16 inch Melles Griot HeNe internal mirror laser tube when suddenly, no more beam (ooops). So now I was starting to panic, because I needed to ship this laser tube the next day, plus at the moment, the only working lasers that I have are other red HeNes.

    For the procedure below, I used a 2 mW HeNe for the reference laser (R-Laser). Thanks to the sections starting with: Checking and Correcting Mirror Alignment of Internal Mirror Laser Tubes I found and read up on how to realign a red HeNe with no lasing occurring, with another red HeNe. Well I didn't have the time to construct a bore sight and mounting block to (hopefully) get the R-Laser's beam positioned so that when the TNA (Tube Needing Alignment) was installed into the bore sight's mounting blocks, the reference beam would shine directly on center and parallel to the axis of the TNA's bore.

    So I decided to improvise, using a HeNe tube I had laying around which has Brewster windows for use with external mirrors to substitute for the TNA and mounted it on my previously made alignment platform. This tube is virtually the same diameter as the TNA and could thus be swapped for it with its bore in the same location. With no mirrors, the red beam easily pass through it permit accurate alignment.

    Note: (If you have a dead HeNe tube of the same diameter. you could pull the mirrors off and use it for a bore sight, (but first, read the section: How Can I Tell if My Tube is Good?). And, be advised that the "god of dead lasers" could become very upset with you, plus Sam and myself both agree that this would be sacrilegious) so you will have to make some type of atonement with the laser gods. Maybe, resurrect two other helpless laser tubes for each one used for this purpose. :-)

    The TNA and this special HeNe that I used, are the same diameter (+ or - a few .001"s). This allowed me to align the mirrorless tube (MT) with the reference beam right down the center of its bore (with the two Brewster windows facing upwards), with a very clean red spot exiting the rear Brewster window.

    Next, I carefully removed the MT and without moving the alignment platform, installed the TNA. Of course, the centering of the reference beam was off axis slightly (on the vertical plane) due to the MT's Brewster window's index of refraction. But this didn't matter in this case, even with the the curvature of the internal reflecting surfaces of both lasers' OCs, because the outer reflecting surfaces of both OCs were flat and parallel. So I used the small reflected spot for reflection alignment, not the larger one caused by the curved internal reflecting surfaces of the OCs.

    I chose to use the OC mirror-end on the TNA for reflection alignment back to the R-Laser's OC for two reasons:

    1. When you get the TNA OC aligned so that it's reflected spot is centered in the R-Laser's beam and producing interference patterns dancing around in the spot, you are then ready to power up the TNA to see if it will actually produce a beam. If it does, (good job!) now you can use the TNA's beam to align it with the R-Laser's beam (providing you have two power supplies). (This took me approximately 20 minutes at the most for the initial alignment of the TNA's OC once the R-Laser's beam was aligned down the center of the TNA's bore, and yes, it did actually produce a beam.

      But even if you only have one power supply, I found it a good idea to use the TNA's OC for reflection alignment instead of the HR mirror because if during the fine tuning of the TNA, it happens to get so far off alignment to where there is no longer a beam and you haven't yet moved either the reference beam or the alignment platform, than you can save a lot of time by not having to set everything up all over again!

    2. The OC has a higher (%) of transmission and by aligning the reference laser's beam right down the center of the TNA I found using a white or fluorescent orange sticker behind the rear HR mirror that there was still a small amount of the R-Laser's beam getting through the OC and a weak smaller diameter spot to exit the rear HR mirror of the TNA.

      Although it shouldn't make any difference as to the amount of light from the R-Laser actually getting through the HR and OC mirrors on the TNA tube reguardless of which end of the tube the beam enters, what I discovered is that by shining the R-Laser beam into the OC-end of the TNA, you will see a faint "halo" around the faint dot of the R-Laser's beam. This was very helpful in centering the two axis of the cavities bores. (This probably has to do with the curvature or the OC spreading the beam slightly inside the bore. --- Sam.)

      When shining the R-Laser beam into the HR of the TNA, there was no detectable halo even at only 1" behind the OC mirror. Plus, as Steve Roberts mentioned, by using a bright orange (or red) fluorescent sticker to view the beam (but in this case, the one exiting the TNA), you can see the faint patterns or optical deviations during alignment much more easily. This is very helpful in this procedure as the amount of light from the R-Laser beam actually getting through the TNA's optics is very small. A darkened, but safe place to work, is advisable for this method.

      Note: I used a 13" long Melles Griot tube that was already aligned and lasing well just for testing this procedure. When I got the TNA tube aligned and centered with the reference beam, I could actually see the tiny dot of light that exited the TNA and dancing with its optical interference. You want to get more than just any dot exiting the HR mirror of the TNA, what you want to do, is keep fine tuning the alignment of the reference beam with the TNA, until you see the brightest dot & centered in the halo.

    Now for the Quick and Easier Short Cut

    As long as you have two identical tubes (diameter wise) and one is already lasing well, and providing that the OCs do not have a "wedge mirror" (see section: Ghost Beams From HeNe Laser Tubes, then you actually could save a lot of time by not having to align the reference beam exactly down the TNA's bore.

    What I did: I used the The Lasing Tube (TLT) in the alignment jig as the substitute for the TNA. Then, I just aligned it with the R-Laser's beam until I got the reflected and "round" spot centered and with the reference beam. I was viewing the reflected larger spot on the TNA substitute tube. If the larger reflected spot is oval shaped the you need to re-center the reference beam with the center of the TNA's OC mirror. Once this is done, you could be home free for the initial dual alingment.

    Then, replace the TLT with the TNA and don't move the alignment. With the OCs facing each other, if the reflected spots are not centered, just dial in the reflected spot from the R-Laser to center itself with it's larger and small reflections. There should be a lot of optical interference now - dancing or flickering. If the spots are centered and round, then there is a good chance that this mirror on the TNA does not need adjusting at this point, so turn the TNA around and repeat the above procedure, then power it up. If all went well, it will be lasing somewhat. Then, only fine tuning will be required. See the section: Minor Problems with Mirror Alignment. Now you can fine tune the alignment of whichever mirror needed original adjustments.

    Notes on Viewing the Spot

    One method that I have not seen anywhere on the Web, for fine tuning any laser's beam without the professional and high dollar alignment equipment is the following (I tried it and found it very helpful):

    Just getting a laser to produce a beam, after realigning the optics as you know, is really only step one, but getting the optics (fine tuned) is another step. Sure you could just try to adjust by the brightness, but who wants to stare at a bright laser spot continuously. And, if one doesn't have a power meter, here is what I did.

    I used a clean lens {concave/concave, at a distance greater than the lens's reflected focal length, to eliminate reflected interference patterns on the OC and back through the lens}, in front of the OC with the beam centered in the lens and with a white piece of paper at a close enough distance from the laser for visual clarity, (my setup was 4 feet).

    By using a lens with an appropriate focal length, the beam's spot was spread to about 3" in diameter which made it much easier to see variations in the beam and it's patterns as the optics were adjusted, or even just with a slight amount of pressure on them. Of course, a laser power meter would be a good substitute and helpful, but not everyone has one. The lens's focal length and the distance between the laser's OC & the viewing paper are variable.

    I have fixed (realigned) many HeNe tubes that were not lasing at all to start with. But now, I definitely find it much easier to fine tune them using the lens to widen the spot. Then I watch the outside edges for rings and off center irregularities. Argon ion laser tubes are also very difficult if they are not lasing at all, but at least on many of them, their mirror mounting plates are easier to apply pressure to at the X-X, Y-Y, and Z-Z axis and all combinations, to see if you're not to far off and which way. This would apply, for example, to NEC tubes though Cyonics/Uniphase types use HeNe style mirror mounts.

    Notes on the Alignment Jig

    I found that with my homemade alignment jig, similar to the one discussed in the section: Simple Adjustable Optics Platform but much heavier and with two horizontal adjustment screws, it was actually easier to mount the TNA on it and leave the R-Laser stationary. I don't really know if the leverage of the geometry works out any different, but it appeared to be much easier to make the vertical and horizontal adjustment. But of course this will depend on your alignment jig's design plus, how stationary it is and your preferences.

    Daniel's Method for Aligning External Mirror Lasers

    This was written for HeNe lasers using a dual Brewster window plasma tube and external mirrors. Some slight modification will be needed for single Brewstered tubes or other gas lasers.

    The following assumed that you have a setup consisting of a Reference Laser (R-Laser (a HeNe laser is assumed since for aligning a HeNe laser, that is the toughest as the wavelengths are the same. The laser being aligned consists of the Tube Under Test (TUT) and the resonator with the external adjustable mirrors (front and rear). The pinholes are aids to alignment.

    
       +---------+    |                     |                       |
       | R-Laser |==> :            ||       :     /===========\     :       )|
       +---------+    |                     |          TUT          |
                   R-Laser       Front    Front     (Removed)     Rear     Rear
                   Pinhole       Mirror  Pinhole                 Pinhole  Mirror
    
    Where the R-Laser and TUT are of difference wavelengths, overall difficulty will be somewhat reduced as more light will get through the front mirror. For more on the alignment jig and related topics, see the sections starting with: Daniel's Methods for Internal Mirror HeNe Tube Mirror Alignment since much of the setup is similar.

    Note: One assumption made here for the procedure below to work properly is that the Brewster windows on the TUT are of high quality plane parallel optical glass or quartz and are set at equal and opposite angles (/=====\ NOT \=====\). This should be the case with most commercial tubes. However, if these conditions aren't met, there can be a slight shift in position and/or angle of the beam passing through the tube's bore which would mean that aligning the resonator mirrors with the tube removed would NOT result in exactly the same adjustment settings once the tube is replaced.

    (From: Daniel Ames (dlames3@msn.com).)

    Note: since not all HeNe laser tubes follow the same design, ideally the way you want to orient the resonator's optics is to have the curved mirror for the TUT furthest away from the R-Laser. If this turns out to be the TUT's OC mirror that's all the better, since we will be looking for the reference beam's reflection from this rear mirror to align it back through the pin holes within the cavity. Either mirror as the rear mirror will work for this method, but I prefer to use the curved mirror since it will counteract the (+) divergence of the reference beam.

    First determine which mirror is the [curved] one, simply by observing the reflection of the R-Laser's beam from the two surfaces of each individual mirror. On the curved mirror end of the TUT's resonator, place a white or fluorescent orange card between the Brewster window and it's related optic, then take your strongest Hene and shine it through (from the outside) of this mirror, then dim the lights and see if you can see the beam on the target paper. Chances are, you will, faintly. If so, then you can use a HeNe to align a HeNe. Sounds impossible, but it is NOT as I have actually done it and succeeded with this using only a 5 mW HeNe laser - HeNe to HeNe.

    Note: Some pinholes are helpful. Three (3) pinholes, just barely larger in diameter than the R-Laser's beam would be ideal, unless one has a very symmetrical eye. But at least one pinhole is a must for within the resonator's cavity.

    Remove the TUT from the resonator and align the R-Laser's beam right down the optical center of the mirrors starting at the curved mirror, (hopefully it's the HR).

    Now, we are ready for the actual alignment. This part is even easier if the mirrors are easily removable.

    Flavio's Comments on HeNe Tube Mirror Alignment

    (Portions from: Flavio Spedalieri (fspedalieri@nightlase.com.au).)

    Below is a simple diagram that shows the end configuration of a typical internal mirror laser tube:

           \
    	 \ __   __
    	--|  |_|  |-
              |  |_|  | |====> Laser Beam
    	--|__| |__|-
    	 / ^    ^
           /   |    |
               |    +--- Adjustable part of mirror mount
               +--- Fixed part of mirror mount
    

    The end of the mount is divided in two with a gap between the first and second sections. At the time of manufacture, HeNe laser tubes are aligned for optimum power output.

    On some HeNe tubes (as well as internal mirror argon ion laser tubes), this gap may me covered by a ring with three (3) adjustment 'grub' screws as shown below:

           \     ___
    	 \ _|   |_
    	--| |   | |-
              | |(X)| | |====> Laser Beam
    	--|_|   |_|-
    	 / ^|___|^   (X) denotes adjustment 'grub' screw (1 of 3 shown).
           /   |  ^  |
               |  |  +--- Adjustable part of mirror mount
               |  +--- Ring with three (3) grub screws
               +--- Fixed part of mirror mount
    

    Or see A HREF="3slcmg1.jpg">Three-Screw Locking Collars on Melles Griot HeNe Laser Tubes for photos.

    If the tube has the metal ring with the grub screws, some people have been tempted to re-adjust this - very BIG mistake.. and the reason is this: With the ring in place and the screws tight and sealed from the factory, the whole assembly is very solid. Now, if you try to re-adjust the grub screws, trying to extract more power, more than likely you will throw out the entire tube out of alignment. The screws are so tight, that very slight, and gentle and precise alignment is very difficult to achieve.

    For tubes that DO NOT have this assembly, once the mirrors are out of alignment, it is extremely difficult to re-align the tube. Been there, done that. :(

    Now, the reason that there are troubles with realigning a tube so it is stable are two-fold:

    1. Mirror mounts are very difficult to physically change position (very slight movements) and maintain that position. You are physically bending metal so it is easy to overshoot the desired position. In addition, the mount will not relax to its final position and stay there - there may be some drift or creep over time especially after multiple thermal cycles.

    2. Heat - even if the tube is only on for a few seconds, the very slight temperature differences can be enough to change the mirror alignment. At the factory, the alignment is optimized after a warmup of 30 minutes or more and minimum output power is usually specified after a similar period of time.
    In some older HeNe tubes, I have seen the affects of thermal expansion, the beam will drift in and out of alignment, and this start to occur only after about 5 to 10 sec after power-up.

    WARNING: All the adjustments that you do on the tube, unfortunately have to be done while the tube is powered up - so you have at least one end of the tube (usually the anode) floating at 2 kV or more once the tube is running (and even after power down due to tube and power supply capacitance). If during your adjustments, the tube decides to drop out, and re-start, you will have the 8 to 15 kV starting voltage - so please be very careful!

    As the tube is powered, try and push the mirror mount, and watch the beam, once you get a nice bright output, try and hold that position, and see if it will hold the output as you support that position - Note in which direction / movement you used to achieve this.

    If you have the ring/grub screw assembly, moving one of the screws will not necessarily adjust the mirror in the direction that you want, so you may have to use different adjustment/pressures on all three screws.

    (From: Sam.)

    If the alignment is nearly correct - gentle force or just touching the mirror mount results in full power - I would suggest as an alternative: Instead of actually attempting to bend the mount, add an external 3 screw adjuster to the problem mirror mount. This will operate within the elastic limit of the mounts so the risk of breaking them off from repeated unsuccessful attempts at bending them back and forth is eliminated. Let the tube warm up for at least 30 minutes, then gently adjust the screws to optimize power output.

    Rich's Procedure for External Mirror HeNe Laser Alignment

    As written, this would appear to be apply more to determining if a combination of HeNe tube and mirrors will lase. Modify as appropriate where you are doing this with an existing laser.

    (From: Richard Alexander (pooua@aol.com).)

    1. Use a good optic axis (very important). A good rail is worth the money.

    2. Use a second Hene laser. I pity those who lack this option.

    3. Use good adjustable mounts.

    4. Mount everything except the mirrors onto your optic axis.

    5. Once you have the beam of the 2nd HeNe shining down through the tube of your 1st HeNe (and out the other end), mount the far mirror on the optic axis.

    6. Adjust the mirror mount so that the beam of the 2nd HeNe reflects back through the tube of the 1st HeNe and strikes the 2nd HeNe next to the Output Coupler of the 2nd HeNe.

    7. Put the other mirror on the optic axis.

    8. If your tube is functional, you could apply power to it, and then fiddle with the last mirror until you get a beam.
    With practice, this method can be completed in less than an hour, though 4 or 5 hours is not unusual, either. If you have a 3rd HeNe, or better equipment than we had in tech school, you might get done much faster. Rich's Procedure for External Mirror HeNe Laser Alignment As written, this would appear to be apply more to determining if a combination of HeNe tube and mirrors will lase. Modify as appropriate where you are doing this with an existing laser.

    (From: Richard Alexander (pooua@aol.com).)

    1. Use a good optic axis (very important). A good rail is worth the money.

    2. Use a second Hene laser. I pity those who lack this option.

    3. Use good adjustable mounts.

    4. Mount everything except the mirrors onto your optic axis.

    5. Once you have the beam of the 2nd HeNe shining down through the tube of your 1st HeNe (and out the other end), mount the far mirror on the optic axis.

    6. Adjust the mirror mount so that the beam of the 2nd HeNe reflects back through the tube of the 1st HeNe and strikes the 2nd HeNe next to the Output Coupler of the 2nd HeNe.

    7. Put the other mirror on the optic axis.

    8. If your tube is functional, you could apply power to it, and then fiddle with the last mirror until you get a beam.
    With practice, this method can be completed in less than an hour, though 4 or 5 hours is not unusual either. If you have a 3rd HeNe, or better equipment than we had in tech school, you might get done much faster.

    Dave's General External Mirror Alignment Techniques

    (From: Dave (Ws407c@aol.com).)

    This is my new method of laser alignment. This works well for most narrow bore HeNe and ion lasers. As of today, it is the best yet. :-)

    Ever hold a HeNe or other laser tube in your hand and just hold it up to your eye and sight through the bore and look at something across the room and target it? Quite easy to repeat. I always wished I could shine a laser beam down the same tube with the same accuracy and speed. Especially when trying to align a laser!! I have aligned quite a few lasers over the years via this same tedious method and to be frank, I am sick of it. :-) In the beginning of the hobby I really enjoyed doing this and worked it down to a science but it is still a pain , all that laser light splashing all over the place , fiber optic effect of the light zig-zaging back and forth the bore etc. Well this method works for me and I'm sticking with it :-)

    You will need:

    Here is the procedure:

    1. Remove OC and place light bulb locked in the 3rd hand 8" to 14" in front of the laser where you removed the mirror.

    2. Look through the HR and down the bore, and line up the bulb filament so it is centered, this should be quick and easy. :-) In my case laser is on my right pointing left at the bulb on my left, lined up.

    3. Hold LED flashlight (diffuse source) through HR down the bore.

    4. Place glass slide (beamsplitter) between bulb and laser and orient at 45 degrees so you can sight the LED flashlight, again real quick and easy. :-) Lock in place with 3rd hand, or in this case with 4th hand. :-)

    5. While looking through the slide, twist HR mirror mount adjustments til you see the bulb filament through the slide, center it, done. Again quick and easy. :-)

    6. Remount OC, fire up laser and fish for the beam, your done.

    I like this method because all critical alignment is accurately sighted directly and quickly by eye. This satisfies my natural wanting to look directly down the bore and immediately align the mirror directly by eye. :-)

    I can't believe I haven't tried this before.

    I read the HeNe laser in SciAm and it is pretty much the same setup, but they do not mention to shroud the Brewsters which helps greatly to maximize the contrast. Shrouding the SP-907's Brewsters made it a snap. :-) Yup, tried it a couple of times on the SP-907 and it works awesome. I use 1 steering mirror with the 907 for the tube is too long and this way I sit at the HR-end and tweak while looking down to the mirror.

    Dave's Quickie External Mirror Alignment Technique for the SP-120

    The following works for the Spectra-Physics model 120 and other lasers with spherical OCs where the optics and machining are most excellent. Interchange OC and HR in the procedure below if your laser only has a spherical HR. I doubt it would work reliably depending only on close tolerances for a planar mirror.

    (From: Dave (Ws407c@aol.com).)

    As far as the terrible 3 point mirror mounts on the SP-120, I have developed a way to get the mirrors aligned without any cards or another laser. Just my two hands and a hex wrench. Within 5 minutes I get it every time. :-) I have also been applying this technique to the longer lasers with some good results.

    As you know, if one mirror is aligned correctly, the other is a cinch. I tighten down the OC and then back off each screw 3/4th of a turn. Then I loosen up the HR so it has a lot of play. I put my finger over the HR and wiggle in a repeating all over the place while hunting for a flash out of the OC. When I get a repeatable flash on the OC that's it, no problem to tweak in the HR. Works every time on the SP-120. :-)

    Dave's Preferred External Mirror Alignment Technique for Long Lasers

    For this procedure, both mirrors are left in place. The OC of the Laser Under Alignment (LUA) is facing a Reference Laser (Ref-L). Note that like many of the other techniques, this does require that the Ref-L's beam is sufficiently narrow and collimated that it can get through the LUA's bore with at most minimal wall contact. To what extent this is possible will depend both on the beam characteristics of the Ref-L and the curvature of the LUA's HR (which affects the size/divergence of the reflected beam).

    1. Set up the Ref-L on a scissor lift with XY adjustment and the aperture of Ref-L and LUA approximately 6" apart. For aligning a red HeNe, the Ref-L should be some other wavelength like green which will pass more easily through both of the LUA mirrors.

    2. On opposite end of LUA, a convex lens is held in a "3rd hand" projecting onto a white card or the wall to expand the emerging beam to say an inch or more so you can see what you are doing way over on the other end of the setup while getting the Ref-L perfectly aligned with the bore.

    3. Once the bore is perfectly aligned, insert a glass slide held in another "3rd hand" in between the lasers and reflect the light coming from the laser to be aligned up onto the ceiling or wall or screen.

    4. What you will see reflected if the LUA is totally out of alignment are two spots reflected back from the OC: The mirror surface (bright spot) and the AR-coated surface (dim spot).

    5. Pick an axis and rack this mirror (OC) a few turns to get its reflections out of the way.

    6. Take a walk to the other end of the laser and start twisting the HR mount till you get a 3rd spot projected on the ceiling (preferably over the center of the laser. :) Center this spot for a good clear round spot and if there is some scatter lopsided about this spot, it is OK (bore is slightly out). You want the spot to be round and clear.

    7. The best part: Superimpose the bright OC spot over the HR spot and suddenly there will be an eruption of laser light from the laser that was out of alignment. :-)

    This procedure is nice and easy to perform but even better, REAL EASY to SEE what's going on - no squinting down the bore to look for a light bulb you can't reach. :-)

    Cat's Eye Mirror for Hassle-Free Alignment

    Haven't you always dreamed of just dropping the mirrors into an external mirror laser cavity, power up, and have the laser operating at near optimal performance without ever touching an alignment screw? Well, there is a relatively simple optics configuration that has the potential to make this possible

    While many large bore lasers like the M60 Tank ruby rangefinder laser have used optical roof prisms or even corner cube reflectors for the HR, this isn't practical for narrow bore HeNe and ion lasers. Wny? Well, for one thing, no roof prism or corner cube has perfect edges so there will be some scatter in the region where they are - but for a laser with a 1 mm diameter beam, that's a relatively large percentage of the mode cross-section and effectively kills lasing.

    However, there is a combination of a curved mirror and convex lens called "cat's eye" due to its similarity to the arrangement in, well, a cat's eye. This behaves much like a corner cube reflector but without its problems (at least over a small angle). Rays entering the lens will be reflected directly back in the direction they came, at least close to the optic axis for small angles. The optimal arrangement has an AR coated convex lens placed at one focal length (f) from a mirror with an RoC of f, coated as an HR or OC for the laser wavelength. In principle, any narrow beam laser could benefit from this. The cat's eye reflector has been tested with HeNe lasers but would certainly work as well for other narrow bore lasers - which are those creating the most problems with mirror alignment. Of course, the manufacturing cost would be higher but how much is your time and sanity worth? :)

    Apparently, using such a setup allows the mirror assembly to be held by hand or with a pair of tweezers and get stable lasing. Now, I can do this with my 1-B HeNe laser tubes and a normal HR or OC, the cat's eye makes it even easier. ;-) Of perhaps more importance, since angular sensitivity of the mirror is greatly reduced, it would be possible to say goodbye to the annoying power drift due to alignment changing that often occurs as the laser warms up.

    This was presented in the paper: "Adjustment-free cat's eye cavity He-Ne laser and its outstanding stability", Zhiguang Xu, Shulian Zhang, Yan Li, and Wenhua Du, 2005 Optical Society of America, Optics Express, vol. 13, no. 14, pp. 5565-5573, 11-July-2012.

    Here is the abstract (spelling and grammer NOT corrected):

    "This paper introduces an innovative He-Ne laser which exhibits many advantages to current He-Ne lasers. With cat's eye reflector as the reflecting mirror, the new laser can solve the conventional problems of laser adjustment and power stability. Comprehensive experiments are carried out both in a half-external cavity and a full-external cavity He-Ne laser. Then the results from the cat's eye cavity, plane-concave mirror cavity and concave-concave mirror cavity are compared, which show that in halfe-xternal cavity laser, cat's eye cavity can improve the laser stability up to 10 times better than other cavities and lower the power drift significantly. And in the full-external case the improvements are much greater even up to 60 times and power drift is minished greatly too. The adjustment problem is also considered and solved. A stable and adjustment-free He-Ne laser is finally realized. The examination of a cat's eye reflector is described."

    Now, while it's quite likely that the benefits of the cat's eye configuration were recognized long ago, the added complexity and cost - and especially the losses through the AR coated lens - would likely have prevented it from even becoming widely used. Only with a very long HeNe laser like a Melles Griot 05-LHR-927 or Spectra-Physics 127 would those losses be relatively small compared to the gain. But even so, would still result in a significant reduction in power, not to mention the issues of making sure 2 additional optical surfaces are perfectly clean. One way around the loss problem might be to use an aspherical HR mirror reflecting off-axis to the spherical cavity mirror instead of a convex lens, but then the cost of manufacturing that very special mirror would probably be totally ridiculous.

    My question - which I'm not sure is answered in the paper - is: What happens if cat's eye reflectors are used at both ends of the laser? Is the thing then totally self-aligning? :)

    Mirror Alignment with just an Optical Power Meter

    Here's a way of aligning mirrors very quickly on small to medium length external mirror (one and two-Brewster) lasers. With care, it may work on large frame lasers as well. This approach takes advantage of the fact that the far mirror will be aligned when maximum bore light is returned to the front of the tube. Unlike the laser wavelength, most bore light passes through the mirror and there is ample power to monitor. It works great on one-Brewster HeNe lasers as well as the very difficult to align PMS LSTP-0010 and LSTP-1010 tunable HeNe lasers. I would recommend it only for lasers with screw adjustable mounts, not for HeNe laser tubes with three-screw locking collars or less.

    All that is needed is an optical power meter (laser, photographic, etc.) with enough sensitivity to respond well to the bore light. One with a "suppression range" feature is best but this is not essential. (The suppression range enables the constant light to be cancelled out so that the sensitivity to changes can be increased.)

    To align one mirror, place the sensor of the optical power meter at the other end of the laser, located to pick up the bore light. Set up the meter on a range that allows the maximum deflection of the meter while keeping it on scale, and/or set the suppression range to cancel out most of the constant bore light.

    Now all that's required is to twiddle (technical term!) the far mirror to maximize the power reading. With kinematic or gimbal mounts, this will actually be quite easy. The peak is broad so each axis will have an effect even if the other axis is way off. As the alignment approaches optimal, the reading will increase and with a bit of luck, will then spike as lasing occurs (assuming the other mirror was already aligned).

    For a laser with two adjustable mirrors, just repeat the procedure for the other mirror.

    It takes literally only a couple of minutes to do this for a PMS tunable laser (which uses a 1-B tube with permanently adjusted internal OC), which with its narrow bore is very difficult to align with any of the other techniques.

    Inconsistent Behavior of HeNe Laser Alignment

    With a healthy HeNe laser, the adjustment of mirror alignment will result in behavior that is smooth and repeatable. Except for the mode cycling variations in output power, the result will be fairly stable even from one power cycle to the next. Even the very rudimentary adjustments using the screws on Melles Griot locking collars perform generally in this way. Unless....

    You may come across a laser tube or head where nothing works as expected. After peaking power, the output may drop after a few minutes such that adjustment is again needed. After that, the same thing may happen again. And again. The output power may be extremely sensitive to mirror alignment even to the point were gently clamping the tube in a head cylinder using the nylon screw may cut output power in half or worse. Or, supporting the tube at various points will significantly affect it. And just the weight of a popsicle stick on one of the mirror mounts will change power significantly. What's going on?

    If the laser is from a surplus supplier or eBay, it's quite possible - actually quite likely - that it was a reject, possibly due to a bad design (yes laser designers make mistakes!) or improper manufacturing. In particular, if the mirror specifications were not correct and matched to the bore, the stability of the resonator with respect to the various modes could be so low that each one sees a significantly different gain. So, after optimizing one set of modes, as they drift with respect to the gain curve, there could be very significant power fluctuations. Guess where such tubes end up? :)

    Another possibility is contamination like a hair, fiber, or metal sliver, inside the tube. If it extends into the lasing mode volume (the intracavity beam), then peculiar behavior could result, and could change with time, orientation, vibration, etc.

    The effects of IR (3.39 um) mode competition can appear similar but are not likely to show up with most reasonably modern red (632.8 nm) HeNe lasers, though they can be significant for "other-color" tubes.

    I have several tubes that exhibit this sort of behavior. One is a Melles Griot 05-LHR-990, a 10 mW (rated) tube about 18 inches long. It was obtained in a batch of tubes I bought from one of the well known laser surplus outfits mainly to salvage mirrors. So this tube was even considered a reject by them! (It only cost me $2.) The output power is extremely sensitive to any pressure on the mirror mounts (even with the locking collars tight), pressure applied to the sides of the tube with the nylon screws in the head cylinder, and temperature gradients. By adjusting the locking collars, it's possible to achieve over 14 mW by careful tweaking. However, after a few minutes, the power declines to 12 or 13 mW and realignment of the mirrors at one or both ends is required to get back to the high level. The power is still well above the spec'd value but the variation is annoying. It's a nice tube otherwise. :)

    Another Melles Griot tube I have with a similar but more severe symptoms is also of similar size, though I don't know the exact model number. It's output can vary from 4 to 9 mW with almost no change in mirror alignment, and may switch to a multi-transverse mode (TEM01/10 or something stranger) at the lower powers. I rather suspect that there may be something somewhere in along the length of the bore or inside one of the mirror mounts though I can't find it. There is what looks like hair stuck inside one of the mirror mounts but it doesn't extend anywhere near the beam. But, perhaps, it's buddy is hanging out somewhere else.

    Cleaning and Alignment of the Spectra-Physics 907 HeNe Laser

    This is longest HeNe laser that had been available from Spectra-Physics (now Newport) and the model 127 which uses this tube was listed on their Web site and in the Edmunds catalog until recently. However, many of these are still available surplus either as the SP-127 or just the tube designated 907, 107, or the older 082. The design hasn't changed noticeably in 25 years.

    Unfortunately, one of the deficiencies of this laser is that the probability of it remaining aligned during shipping, even if packed with 10 inches of foam all around - is small. Usually, it's just a slight power loss but I've seen cases where there was no lasing at all and a full realignment was needed.

    Being external mirror lasers with Brewster windows, the optics can get dirty, especially if the rubber sealing boots at each end are cracked, as some tend to be, possibly from overzealous removal by a previous user.

    Sam's Procedure and Comments

    There are 4 sets of alignments in the SP-907 (listed in the order in which they should be dealt with for a total realingment):

    Recommended cleaning

    None. :) Actually, if the rubber boots are in good condition and sealing well and there is no reason to suspect that someone before you has messed with them, it's probably best to leave them in place and not to attempt to clean the Brewster windows or mirrors, at least not until you're determined to eak out the last few photons/second of performance. But here are some procedures that work:

    As noted, put the boots back on as soon as possible as a gradual decline in power is inevitable from from dust collecting on the B-windows and mirrors. For a laser like this, even slight contamination not obvious by eye can result in a power reduction of a few mW.

    Total alignment procedure

    Touch up the mirror alignment after the bore centering.

    Keep in mind that all of these adjustments interact to some extent. So, it's possible to be at a local maxima and lose sight of the global optimization. For example, changing the position of one of the bore centering adjustments may reduce the power initially, but adjusting the other one and realigning the mirrors may get it back and more. These are third order effects though, so doing the procedure above should get you most of the may to a happy laser. :)

    Michael's Procedure

    A complete sequence of steps for cleaning and realigning of the SP-907 from a non-lasing condition can be found at: Dragon's Eye Doings: Laser Alignment Heck. This includes descriptions and many photos. My mirror of the content of this may be found at Sam's Copy of Dragon's Eye Doings: Laser Alignment Heck. This isn't the only way to do it (see my summary in the previous section) but is known to work and is straightforward. Bore straightening isn't covered but can be dealt with as an add-on. :)



  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    Collimation of HeNe Lasers

    Reasons for Poor Collimation

    The output of most HeNe lasers is a very well collimated beam - approaching the theoretical diffraction limited optimum possible for a given bore diameter. It isn't expected to look like that of a flashlight! However, a number of factors can affect this performance - and some are by design:

    Improving the Collimation of a HeNe Laser with a Beam Expander

    The following applies to any laser which outputs a substantially parallel beam but is written specifically for HeNe lasers. Collimation of laser diodes require a slightly different approach - see the section: Beam Characteristics of Laser Diodes.

    Although the divergence of a HeNe laser is already pretty good without any additional optics, the rather narrow beam as it exits from the tube does result in a typical divergence between 1 to 2.5 mR (half of total angle of beam). 1 mR is equivalent to an increase in beam diameter of 2 mm per meter.

    As noted in the section: HeNe Laser Tubes and Laser Heads, beam divergence is inversely proportional to the beam diameter. Thus, it can be reduced even further by passing the beam through beam expander consisting of a pair of positive lenses - one to focus the beam to a point and the second to collimate the resulting diverging beam. Though the beam will start out wider, it will diverge at a proportionally reduced rate.

    A small telescope can be used in reverse to implement a beam expander to collimate a laser beam and will be much easier to deal with than individual lenses. (This is how laser beams are bounced off the moon but the telescopes aren't so small.) Using a telescope is by far the easiest approach in terms of mounting - you only need to worry about position and alignment of two components - the laser tube and telescope. The ratio of original to expanded beam will be equal to the magnifying power of the telescope. Even a cheap 6X spotting scope will reduce divergence six-fold.

    If you want to use discrete optics:

    For optimal results, the ratio of collimating lens diameter to focal length (D2/F2) should greater than or equal the ratio of HeNe beam diameter to focusing lens focal length (D1/F1). This will ensure that all the light is captured by the collimating lens.

    The beam will be wider initially but will retain its diameter over much longer distances. For the example, above, the exit beam diameter will be about 10 mm resulting in nearly a 10 fold reduction in divergence.

    Adjust the lens spacing to obtain best collimation. A resulting divergence of less than 1 mm per 10 meters or more should be possible with decent quality lenses - not old Coke bottle bottoms or plastic eyeglasses that have been used for skate boards. :-)

    Note that some HeNe tubes have wide divergence by design using an external negative lens glued to the OC. For these, removing this lens with a suitable solvent may be all that is needed to produce the divergence you want. See the section: HeNe Laser Beam Characteristics.



  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    Beam Polarization of HeNe Lasers

    Typical Polarization Characteristics and Problems

    Sealed HeNe tubes with internal mirrors which are linearly polarized usually incorporate an internal Brewster plate to suppress the unwanted polarization orientation - and these HeNe tubes are more expensive. Where mirrors are external, the Brewster windows on the plasma tube accomplish this function and the output of these is always linearly polarized.

    Common inexpensive internal mirror HeNe tubes produce a beam that is either randomly polarized or slowly changing in polarization (as the tube heats) - possibly with a combinations of polarization states present simultaneously. Placing a polarizing filter in the beam of one of these lasers results in a variation in brightness, usually taking place over a few seconds possibly with sudden shifts as various modes compete for attention inside the resonator. The presence of any of these characteristics makes such a laser unsuitable for many experiments and applications. These tubes are normally designated as 'random polarized' (with an 'R' somewhere in the model number) which translates as: "The manufacturer has no idea of what the polarization characteristics will be at any given time". :)

    If the polarization were truly random, meaning all polarization states are present simultaneously (or on a short enough time scale that it doesn't matter), a simple polarizing filter in the beam path will produce a linearly polarized beam at the expense of at least one half the output power (that which is blocked because its polarization orientation is wrong and because of losses in the filter). However, where the polarization orientation of the laser is slowly changing, this approach will result in unacceptable varying output intensity from the polarizing filter. Additional optics including polarizing beamsplitters, mirrors, and combiners can in principle, at least, produce a stable polarized beam but these are complex and expensive.

    Determining if a Laser Tube is Linearly Polarized

    There are several ways to determine if a HeNe (or other internal mirror) laser tube or head has random or linear polarization (all external mirror laser with Brewster angle windows are linearly polarized): Note: Random polarized lasers are generally not truly random but will tend to (longitudinal) mode cycle and some modes will tend to favor a particular polarization orientation. This will result in the polarization changing gradually or suddenly (especially as the tube is heating up) so at any given time, they may emit a polarized or partially polarized beam but its orientation will not be predictable. Thus, to confirm that your tube is polarized requires that the polarization remains constant - not just for an instant (30 seconds to 1 minute should be enough time to wait to know for sure). If the beam is reflected off of a non-metallic reflective surface (which acts somewhat as a polarizer), you may see a large variation in brightness due to the polarization changing especially if it is a low gain or short tube where fewer modes are active simultaneously.

    Even a polarized tube may show a small amount of variability of the low intensity beam passed by a polarizing filter or reflected from a Brewster angle plate - this is normal and one reason why the specifications only say 500:1 or 1000:1 and not infinity:1. The reason is that the tube's linear polarization results from the cavity gain being maximized by the internal Brewster plate at the polarization angle. However, gain function with respect to angle is not a singularity - there is still enough gain for a few degrees on either side to maintain oscillation. And, some samples are better than others. Also see the section: Typical Polarization Characteristics and Problems.

    Unrandomizing the Polarization of a Randomly Polarized HeNe Tube

    The best option where a polarized beam is required is to start with a HeNe laser that produces a polarized beam! However, for the experimenter, there is at least one alternative - magnets to the rescue!

    I have found that placing powerful magnets alongside a random polarized tube will result in a highly linearly polarized beam. While this may be common knowledge at the Afternoon Teas attended by laser physicists (assuming they drink tea), it certainly isn't something found in popular books on lasers.

    A type of magnet that works quite well has a strength of several thousand gauss. The ones I used came from the voice coil positioner of a moderate size hard disk drive. They are rare earth magnets with dimensions of about 1.25" x 2.5" x .375" with the broad faces being the N and S poles. The amount of polarization is most pronounced by placing one of the broad faces of the magnet against the tube near its mid-point. Some adjustment may be needed to optimize the effect. I do not know how much magnetic field strength is needed but even moving this magnet 1/4" away from the tube surface greatly reduced the ratio of light intensity in the two orthogonal polarization axes.

    CAUTION: These types of magnets are very powerful. In addition to erasing your credit cards and other magnetic media, they will tend to crush, smash, or shatter anything (including flesh or your HeNe tube) between them and/or between them and a ferrous metal. Some portions of a HeNe tube or laser head may contain parts made from iron or steel. These rare earth magnets also tend to be quite brittle. In addition, the violent uncontrolled movement may place you and a HV terminal in the same space at the same time as well! Take care.

    With the magnet's N or S pole placed on the side of the tube, the result was a vertically polarized beam. By rotating a polarizing filter in the beam path, beam intensity could be varied from nearly totally blocked to nearly totally transmitted and the polarization orientation followed the magnet as it was rotated around the tube.

    The control wasn't perfect - a small amount of light with a slowly varying polarization did sneak through. However, it was significantly less than 1 percent of total beam power for these particular tube and magnet combinations (I have tried this with 2 different tubes with similar results). The constant portion of the residual beam may have just been a result of the imperfect nature of the polarizing filter.

    By using two similar magnets - one on either side of the tube with N and S poles facing each other (mounted on an aluminum U-channel for support and so they would not crush the tube), the variation in residual beam intensity was virtually eliminated. I do not know if this effect was due to the increased magnetic field or its more homogeneous and symmetric nature. This was also used successfully with an enclosed HeNe laser head:

                            __S__
                           |_____| Rare earth magnet
          ____________________N_______________________
         |                                            |
         |             HeNe laser head                |=====> Polarized HeNe beam
         |____________________________________________|
                            __S__
                           |_____| Rare earth magnet
                              N
    

    Use of Magnets to Generate Polarized HeNe Laser Beam shows acceptable locations for one pair of magnets along side a typical 1 mW HeNe tube. This placement was found to be effective but possibly not totally optimal - experimentation may be required. Under some conditions, a single magnet slighlty separated from the tube seemed more effective, possibly because the field was spread over a longer stretch of bore.

    As far as I could tell, with this dual magnet configuration, the output beam characteristics were similar to those of a polarized HeNe tube. However, additional and/or more powerful magnets might be necessary with other tubes.

    Output power did not appear to be affected significantly. A measurement done later on a Melles Griot 05-LHR-911 HeNe laser head showed that when the polarization effect was most complete, the output power decreased by about 5 percent. A polarizing filter would nearly totally block the beam at one orientation and have minimal effect 90 degrees away from this.

    I do not know about the stability or reliability of this scheme but the only other effects seem to be to increase the required input starting/operating voltage and/or magnitude of the negative resistance of the tube slightly (current dropped by about 10 percent with the magnets using an unregulated power supply) and possibly to shift to point of maximum beam power to a higher tube current (5 mA instead of 4 mA for one tube - but this could have just been my imagination as well). With that 05-LHR-911, the operating voltage at 5 mA increased from 1,500 V to 1,550 v with one set of magnets and to 1,600 V with two sets. And, the laser would not stay on in a stable manner with the magnets very near the anode (cable) end of the laser head, but I didn't think to try and adjust the current setting to see if that would help.

    As a side note, output power *increased* by about 5 percent with a magnet some distance from the laser head, possibly due to the Zeeman splitting suppressing IR losses, but of course there was no effect on polarization.

    Where the capillary of the plasma tube is exposed as with many older lasers, and the magnets can be placed in close proximity to the bore, their strength can be much lower. Some commercial lasers (like the Spectra-Physics model 132) offered a polarization option (-01) which adds an assembly consisting of several ferrite magnets glued between steel plates that screws in place alongside the tube with the pole pieces (the steel plates extending beyond the magnets) above and below the tube bore. I performed some tests using a near-mint condition SP-132 (from around 1973) and the original magnet option, the extinction ratio and power stability are not as good with the magnets compared to the more common approach using a Brewster plate or Brewster window tube though. The Spectra-Physics specifications only claim an extrinction ratio of about 30:1 compared to 500:1 or better for a laser using a Brewster plate or window. I doubt that magnets are used for polarization in any modern HeNe lasers.

    Since it is possible to control the polarization orientation with permanent magnets, the next step would be do this with electromagnets. This would permit polarization to be dynamically controlled. Adding a fixed polarizer would provide intensity modulation without any connection to the power supply or expensive electro-optic devices. Hopefully, by using multiple sets of coils distributed along the side of the HeNe tube, a lower field strength would be adequate. Liquid helium cooled superconducting electromagnets would definitely add to the cost of the project. :-) Perhaps, someday, I will try this out.

    Magnets and Mode Plots

    Several years after doing this initial experiment, I recorded the mode behavior of a common barcode scanner HeNe tube with different magnetic field configurations. The laser itself is the same one used to construct the system described in the section: Sam's Very Simple Stabilized HeNe Laser, mainly because it already had photodetectors mounted for the two polarization orientations. It was also known that the particular tube was nicely behaved with respect to mode sweep, with no tendency to flipping. Two sets of magnets were used. The plots were obtained for each magnet set over about 8 minutes as follows:

    The following two sets of magnets were used for these tests:

    1. Three pairs of medium strength ferrite magnets positioned on each side of the tube resulting in a fairly uniform transverse field over about half the length of the tube roughly centered between the mirrors. These magnets are 2" high by 5/8" wide by 1/4" thick, placed 5/8" apart side-by-side. Plot of Spectra-Physics 088 Mode Behavior in a Moderate Transverse Magnetic Field shows the results.

      An aluminum frame with the 3 sets of magnets was placed over the tube in each of the 4 possible orthogonal orientations for about 1 minute. The strength and configuration of these magnets results in the beam being somewhat polarized, but with significant double frequency ripples in intensity of both modes. Surprisingly, the preferred polarization orientation is orthogonal to the orientation of the magnetic field! When the magnets are rotated 90 degrees, the effects essentially swap polarizations.

      Note how the total power is significantly higher than with no field for the first two magnet orientations. There is also a the difference in shape of the modes depending on orientation of the magnets, ranging from pulse to sinewave.

      Orientations of the moderate strength magnetic field other than horizontal or vertical produce intermediate effects, but with low stability and even some flipping behavior.

    2. A single pair of very strong rare earth magnets similar to those in the setup described above, but somewhat longer along the direction of the tube (about 2-1/2 inches). Each magnet was 1-3/8" high by 1-3/4" wide by 1/4" thick located toward the anode-end of the tube. Plot of Spectra-Physics 088 Mode Behavior in a Strong Transverse Magnetic Field shows the results.

      The super strength pair of magnets was placed over the tube, again for about 60 seconds in each orientation. In all cases, the polarization preference was very strong and lined up with the magnetic field. The suppression of the mode orthogonal to the direction of the magnetic field was nearly perfect. But, the total output has declined for all but the second orientation (Ver-N) and quite dramatically for the last one (Ver-S). That reduction is at least 20 percent and quite obvious simply observing the brightness of the spot on the photodiode.

      It may be possible to find a preferred orientation of the high strength magnets where the total output power is maximized with good stability. Ver-N seemed particularly good with total power at least equal to that without magnets.

      A closeup of the first case is shown in Plot of Spectra-Physics 088 Mode Behavior in a Strong Transverse Magnetic Field (Hor-N). However, note from the plot that although the S-Mode (blue) is quite close to 0, the P-Mode (red) and total power are rather lumpy. Of course, this is probably not the ideal magnetic field configuration to force linear polarization being only a single pair of magnets. And, the polarization ratio is probably only about 50:1, not the 500:1 or 1000:1 of a normal linearly polarized HeNe laser.

      Orientations of the strong magnets at other angles aligns the polarization with them, but, but with various amount of a reduction in total output power.

    The individual polarized modes and total power were captured for the plots. The two orthogonal polarization orientations (shown in red and blue) were detected using the waste beam and photodiodes that were already present in this laser. A trans-impedance op-amp buffer converted their uA-level outputs to the +/-10 V range of the data acquisition system. The total power (shown in dark green) used the main beam with a photodiode and resistor load. The scale factors of the three signals are fairly close but not perfectly matched. However, even accepting errors in the scale factors, there are additional unexplained discrepancies between the sum of the modes and total power, which should be equal. This is especially evident for the high strength magnets comparing the dominant mode to total power (since there is almost no power in the other mode). I'm not positive of the cause but suspect some interference effects in the detector optics for the waste beam. While not actually destabilizing the laser, multiple reflections could explain the variation in mode amplitude, though why it seems worse for one of the modes is a mystery. But the ripple remains with the optics channel when swapping the photodiode electrical connectors. And, rotating the pickup assembly so it is at a slight angle definitely makes a difference. I have seen some high frequency noise of the laser output power possibly due to plasma oscillation in the tube with the high magnetic field. This may be resulting in some differences in response through the electronics. It looks like the total power is fairly well behaved, but the waste beam power has the irregularities. So, perhaps the preamp needs some attention. But I did try increasing the time constant of the op-amp buffer by a factor of 10 with no noticeable change in the response. Hmmm. With the magnets centered between the mirrors, the discrepancy between total power and the sum of the mode power was even worse. When pushed toward the anode-end of the tube, it seems to quiet down, though that might have just been a coincidence. I'm going with the optics interference explanation for now. :)

    Magnet Tests with a Polarized HeNe Laser Tube

    In an effort to determine how strong the effect of a magnetic field actually is on the polarization, I later tried the same scheme as described above but with a small polarized tube - the type with an internal Brewster plate. Not surprisingly, the effect of the Brewster plate is much more pronounced than that of even very powerful magnets. Here is the output power after warmup with different magnetic fields:

    1. No magnetic field: 0.75 mW.
    2. Strong transverse magnetic field (aligned with polarization): 0.70 mW.
    3. Strong transverse magnetic field (orthogonal to polarization): 0.50 mW.
    4. Weak magnetic field (almost any orientation): 0.80 mW.

    In all cases, the polarization was unchanged and output power was at least as stable as without any magnetic field. Thus, even the strong magnetic field was insufficient to overcome the losses of the Brewster plate at the (wrong) orthogonal polarization orientation but did reduce the gain at the (correct) aligned polarization orientation enough to cut output power by 33%. (For this short tube, lasing would probably have been killed entirely if forced to have its polarization orthogonal to the correct orientation.) These results are not unexpected - except perhaps for (4) - I do not know if the increase in power was simply a result of the usual Zeeman splitting effect suppressing the IR wavelengths or something else. A noticeable increase in output power due to Zeeman splitting is usually associated with long high power HeNe tubes, not the 0.5 mW tube used for these tests.



  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    Repairing Leaky or Broken HeNe Tubes

    Gas Fill Problems with HeNe Tubes

    HeNe tubes can fail due to slow leakage through soft (Epoxy) seals (not found on modern tubes), or actual damage or defective manufacturing leading to a crack or break. Helium loss through diffusion is a special case and may be remedied easily without major effort or investment. See the section: Rejuvenating HeNe Tubes. Other gas fill problems will require a very non-trivial amount of work and access to sophisticated. Minor structural damage may be repaired but then a refill will be needed. See the section: Repairing and Refilling a HeNe Tube at Home? However, if your tube is broken in half, you should probably just salvage the mirrors (for possible future use) and move on to other things!

    The major HeNe laser manufacturers and laser repair companies may offer regassing services for larger more expensive HeNe tubes (high power internal mirror tubes or those with Brewster windows designed to operate within an external resonator). Figure on $1000 or more to regas an HeNe tube, and more still if there is physical damage (assuming they will bother with it at all).

    Whether the cost of such an operation can be justified is another matter. For a high quality research laser it probably makes sense as the tube alone may cost several thousand dollars or more - if a replacement can be obtained at all. Even a basic HeNe tube with Brewster windows may cost over $1000 (being much less common and thus much more expensive than the internal mirror variety). And replacements are becoming even less available and more expensive by the day! However, for small sealed internal mirror HeNe tubes, low cost replacements are readily available at perhaps 1/10th to 1/4th the cost of a regassing service (even cheaper if you are willing to use a surplus tube).

    However, where the tube has high mileage and died from use and age, it may not simply be a matter of regassing. The following is from the Melles Griot FAQ Page:

    "While regassing can provide some extension of the output performance in some gas lasers like the CO2, argon and the higher powered side arm HeNes (which have external optics), it is not recommended or provided for smaller internal mirror coaxial tubes. Typical end-of-life failure for a HeNe tube is cathode sputtering. This occurs when the protective oxide layer on the cathode is expended through continuous bombardment by the laser discharge. There is no cost effective way of regenerating this layer. When the oxide layer is expended, the discharge itself vaporizes the "raw" aluminum and deposits this material, in its vapor state, on other surfaces such as the optics and the bore."

    So, while refilling may help some, the sputtered aluminum coating will remain on critical surfaces, and without cathode replacement, the sputtering will continue and the tube will quickly die - again. A careful visual inspection of the bore and mirrors may reveal whether a suspect tube is worth saving - a black or metallic film in the vicinity of the cathode could (that's not a getter flash) indicates that serious sputtering has taken place. On Melles Griot tubes, it will show up as a metallic coating on the glass facing the small holes near the end of the cathode. A white-ish color discharge on a tube that's hard to start and hard to keep running is often an indication of an end-of-life cathode-depleted tube. (However, discoloration in the bore, while also an indication of a high-mileage tube, may have no noticeable effect on performance since this is due to an independent phenomenon.)

    Rejuvenating HeNe Tubes

    These techniques may be used on HeNe tubes that are marginal due to loss of helium though diffusion or a slight amount of contamination from air leakage due to age. Where the tube lases weakly, a helium soak or reactivation of the getter may help. If there is no lasing at all, many other causes are possible.

    The best way to determine if loss of helium or slight contamination is your problem is to check the spectrum of the discharge. See the section: Instant Spectroscope for Viewing Lines in HeNe Discharge.

    Helium Soaking

    Where just the helium (remember how slippery those He atoms can be!) has leaked out, there may be an alternative to the dumpster or a major refurb. HeNe tubes which do not lase well or at all due to loss of helium can sometimes be rejuvenated by soaking them in helium at normal atmospheric pressure for a few days or weeks. You don't need a pressure chamber or any other fancy equipment - a few helium filled party balloons and a garbage bag will do just fine.

    However, there could be other causes like misaligned mirrors or excessive tube current (due to a defective power supply). Check for these possibilities first and confirm loss of helium with a spectrometer capable of actually measuring the relative intensity of the spectral lines if possible. From my experience, just viewing the discharge with a diffraction grating will not reveal a low helium condition unless it is extremely severe - as in almost none remaining. (I've yet to actually see this. If anyone has a HeNe tube with certifiably low helium, please send me mail via the Sci.Electronics.Repair FAQ Email Links Page. I'd be interested in testing it.)

    The point to realize is that it is the partial pressure of each gas inside and out that matters. Neon is a relatively large atom and does not diffuse through the tube at any rate that matters. However, helium is able to excape even when the pressure difference is small. For a typical HeNe tube at only 2 Torr (1/380th of normal atmospheric pressure), the partial pressure of helium in the tube is still much much greater than its partial pressure in the normal atmosphere. So, helium leaks out even though the total pressure outside is several hundred times greater. Conversely, soaking a HeNe tube in helium at 1 atmosphere will allow helium to diffuse into the tube at several hundred times the rate at which it had been leaking out. Thus, only a few days of this treatment may be needed if the problem is low helium pressure. Assuming that the desired partial pressure of He is 2 Torr, the ratio of age:soak-time will be about 380:1 or pretty close to 1 day of soak per year of the tube's age.

    Helium loss is most likely with soft-seal tubes - those with an Epoxy-type adhesive holding the mirrors or Brewster windows in place. However, it is also possible for hard-seal tubes using frit seals or optical contacting to lose helium though probably at a slower rate and rejuvenation will also take proportionally more time. Checking the intensity of the He lines with a spectroscope is really the only way to know for sure if He loss is the problem and to also monitor the soaking process.

    Almost any sort of helium supply will work for atmospheric pressure diffusion including welding supply grade and even the stuff sold for filling party balloons. (Note, however, that these sources are mostly the common isotope of helium, He4, not the light isotope, He3 which may be what was originally in your tube - see the additional comments below.) A party tank of helium may be as little as $15 or $20 or just buy a few prefilled balloons and empty their contents into an air-tight plastic bag containing the HeNe tube. However, make sure what you are getting is really helium and NOT hydrogen!! In addition to the flammability issues, any significant H2 that makes its way into your HeNe tube will make the situation worse - probably terminal. Also note that as much as 50 percent of what is in those party tanks may actually be air, nitrogen, and/or some other unidentified gas, so the process may take somewhat longer (approximately by 100 divided by the percent of actual helium) though most of these contaminants won't hurt the tube.

    The required amount of effort hardly seems worthwhile for a $15 1 mW HeNe tube but it is something to keep in mind for other more substantial and expensive types.

    Note that there are a few types of tubes that won't benefit from helium soaking even if they have certifiably leaky seals. Those are tubes where the seal is between the interior and another sealed chamber as with some older Aerotech HeNe lasers. In these, the leaky seal is on a Brewster window but the laser mirrors are attached with frit or Epoxy to an external sealed chamber which is filled with air. The only thing helium soaking will do is slightly increase the partial pressure of He in that external chamber which will essentially no effect on the internal gas fill. It might be possible to drill a hole in the metal end-plate or melt a hole in the glass of the external chamber though, hopefully without contaminating the Brewster window. Or, use a diamond saw to cut one end off entirely and install the mirror on an adjustable mount.

    (From: Mark W. Lund (lundm@xray.byu.edu).)

    I have rejuvenated HeNe laser tubes with low helium pressure. Since the partial pressure of 1 atmosphere helium is much higher than inside the tube you don't really need to use high pressure, or even increased temperature. I just put them in a garbage bag and blasted some helium into it from time to time. The length of time necessary in my case was a few days, but depending on the glass type, thickness, and sealing method this may vary. It would be good to test the power every couple of days so you don't overshoot too much.

    One warning: Helium has a lower dielectric strength than air, so don't try to operate the laser in helium as it may arc over.

    (From: Philip Ciddor (pec@dap.csiro.au).)

    My information is very old, but may be helpful. Early 2 mW red tubes had about 2 torr of He, so soaking in 760 torr (1 atmosphere) of He for 1 day per year of life roughly restored the initial He pressure, since diffusion rate is proportional to pressure difference. I have no data on the gas mix in current green or IR tubes, but if you can find it, similar scaling may be feasible.

    (From: Sam.)

    Gas fill probably isn't all that different for non-red HeNe tubes so the same general recommendations should apply. However, since their gain is lower, nearly everything about near-IR (1,523.1 nm and 1,152.3 nm), orange (611.9 nm), yellow (594.1 nm), and particularly green (543.5 nm) HeNe tubes is more critical including power supply current and mirror alignment. So, it is important to eliminate other possible explanations for low or no output or other problems before blaming loss of helium.

    I cannot overemphasize the importance of carefully monitoring the amount of helium that has diffused back into the HeNe tube (by removing it from the bag of He and testing with a spectroscope periodically and for a laser beam) - once helium pressure goes too high, the only (non-invasive) way of lowering it is to wait a few years or decades. :-) If power is just low and you are trying this, put the tube in the helium soak for a couple of days and then check power output again. If it has increased, repeat this procedure a couple days at a time until power levels off or starts to decrease. If power decreases after the first soak, helium loss isn't your problem!

    If it's possible to wrap the tube such that only the seals are inside the helium and not the electrode connections (the glass envelope shouldn't leak at any rate that matters), monitoring of power can be done without having to remove the tube from the helium container or whatever.

    CAUTION: Apparently, most modern HeNe tubes are actually filled with the light isotope of helium, He3, rather than He4 which for all intents and purposes, is the one found in nature (99.9998%). He3 has a higher energy state which may be better for exciting certain transitions. Thus, helium soaking with common He4 could result in problems including reduced maximum power, greater frequency spread, reduced stability, or something else. As noted above, once the HeNe tube has been helium soaked, the effects are irreversible without waiting many years. The only practical way to determine what isotope(s) of helium your tube used is probably to ask the manufacturer - even a high resolution spectrometer won't help if the helium has escaped. For a common red HeNe tube, there is little to lose by using common He4 though results may not be optimal. However, if the tube is from a specialized research laser, it would probably be best to have a professional laser refurb company or the original manufacturer deal with it. You could make matters worse.

    WARNING: In addition to not attempting to operate the HeNe tube itself in a helium atmosphere due the lower breakdown voltage, there may even be problems with He diffusing into power supply components or ballast resistors and lingering there. So, if possible, remove the HeNe tube from its laser head or system enclosure for the helium soak. Or else, wait awhile (your guess is as good as mine) after dumping the helium before applying power.

    Reactivating the Getter

    Where some air has entered due to age (not an actual leak, that you can forget about unless you want to try refilling the tube at home - see below), it may be possible to reactivate the getter and absorb/combine/react with the unwanted molecules. However, don't expect miracles.

    Note that not all HeNe tubes have getters. For some that do, the getter may never have been activated in the first place (if the gas fill was already deemed pure enough after pinch-off). See the section: Gas Fill and Getter for info on the getter in a HeNe laser tube. And, if the getter was activated, the source of the active material (in the getter electrode) may have been totally depleted during manufacture so there may be no more remaining.

    This only has a chance of working if the gas pressure is nearly correct - not if it has changed by a factor of 100. The closest example I have of the effect of the getter on tube vacuum is for a typical TV or monitor CRT:

    (From an engineer at Philips)

    "A regular CRT-type getter can reduce gas pressure from about 10-6 Torr to its final value of 10-9 Torr IFF the gases can be gettered at all. H2, O2, N2, CO, and CO2 can be gettered. CH4 (Methane) can not be gettered but by heating, it can fall apart into C (a solid) and H2 that can be gettered. Noble gases can not be gettered either, so their gas pressure will determine the final gas pressure in a picture tube."

    Of course, for a HeNe or Ar/Kr ion laser, those inert gas molecules ARE the desired result! :) Unfortunately, since the typical gas laser operates at a pressure 1,000,000 times higher than a CRT (a few Torr), any effect of the getter on detectable contamination is likely to be minimal. How to tell? If the color of the discharge is more towards white or pink than it should be and there is still at least some evidence of lasing, the getter has a good chance of returning it to normal assuming all its active material isn't already used up. If the color is too orange, then the helium loss may be indicated and a helium soak may be all the tube needs. See the section: Helium Soaking.

    However, there is probably nothing to lose if the tube is unusable and you won't be going the entire route of refilling it. Heating the getter can be achieved in a variety of ways including (depending on design and what you have available): DC current, glow discharge, Sunlight and Fresnel lens, RF, and induction heating, even a microwave oven. See the sections starting with: Methods to Activate Neon Sign Electrodes and Getters. The Solar heater approach is low tech and known to work where there is no visible 'white cloud of death' (heating the white stuff (which is probably unavoidable with the Sun's rays) seems to release previously trapped stuff making the situation much worse). See the section: Simple Solar Heater.

    I've also tried using a 1 watt fiber-coupled laser diodes with a focusing lens to heat the getter but although an incandescent spot could be seen on the getter, there was no significant change in performance. Perhaps I didn't let it cook long enough. A 10 or 20 watt diode or YAG laser might work better. :) But a CO2 laser will not work since 10.6 um can't get through the glass.

    The idea is to drive off some of the material remaining in the getter electrode onto the walls of the tube. If nothing appears or it turns milky immediately, the getter probably isn't capable of helping much - though even in this case, try out the tube again - it may have helped just enough. Lack of results could also mean that the getter electrode hasn't been made hot enough or the material it contained had already been fully used up.

    Note: If you expect to try your hand at actually refilling a leaky tube, DON'T attempt to reactivate the getter - you may need it later!

    The same approach can be used with ion laser tubes if they are made of glass and you can locate the getter. Those that are of all ceramic construction may still have a getter, but it may need to be heated by a precisely controlled current flow between the cathode end-bell and filament or something equally obscure like that - not easily guessed! Also, since these tubes are generally much more expensive than HeNe tubes, it may pay to have it professionally refurbed.

    Once the tube has been revived (or perhaps even before you make the attempt), adding an additional layer of Epoxy/TorrSeal at the tip of the exhaust tube, mirror(s), and any other possible areas of leakage would be a good idea. This is particularly relevant for modern hard-seal tubes since they shouldn't really leak at all (at least on time-scales that humans can understand). Thus, any contamination generally means an actual defect at the frit seals or exhaust tube (tip-off). Soft-seal tubes leak by design :) but adding an additional layer of sealant at the mirrors, end-caps, tip-off, and other suspect locations can reduce this somewhat. At least it won't hurt - unless you accidentally glop it on the OC mirror! :(

    I've successfully revived a couple of Melles Griot HeNe laser tubes which had getter electrodes but no visible getter spots (which means the material is transparent). One was a hard-seal tube that must have been contaminated in some way since after treatment, it has worked essentially unchanged for over a year. The other was a green HeNe laser tube that had an Epoxy seal at one end. However, all attempts to revive Spectra-Physics HeNe lasers have failed miserably and generally made matters worse. Heating the "white cloud of death" material (including what's no doubt inside the getter ring) must release whatever it previously trapped.

    Repairing and Refilling a HeNe Tube at Home?

    If you are really really ambitious, have lots of time on your hands, have access to lab supplies, laser grade He:Ne gas mixture, and a high vacuum system in good condition, you, too, can refill an HeNe tube - at least as an experiment. Whether it could be sealed off and then expect to have a long life is another matter.

    First, any physical damage would have to be repaired. For example, if an overzealous attempt at mirror alignment resulted in a mirror breaking off at the frit seal, it would have to be reattached - in as precisely the same position as possible using new glass frit or Epoxy (though that will leak over time). If someone yanked on the anode wire on a large HeNe tube broke the metal-to-glass seal, that would have to be repaired - again with Epoxy or by actually heating the glass to fuse it together. However, the latter risks shattering the entire tube if you aren't experienced in glass working. If you don't know where the leak is, then you need to find it first. :)

    Once the HeNe tube is known to be gas-tight, the seal is cracked at the exhaust tube, it is put on a high vacuum system to pump it down and backfilled with pure He:Ne gas mix several times while baking out impurities.are very finicky about gas purity.

    For more information on this sort of endeavor, see the chapter: Amateur Laser Construction, the section: Home-Built Helium-Neon (HeNe) Laser, and the introductory chapter: Home-Built Laser Types, Information, and Links for relevant information. Good luck! :-)

    Regrinding or Otherwise Compensating for a Chipped Mirror

    So that wonderful HeNe tube you had fallen in love with (we all have our peculiarities!) got knocked loose from its mounting and fell face down on the granite countertop. Exactly what were you doing with your laser? :) Now, there is a chip in the OC mirror which extends into the area of the beam and the beam is but a shadow (well not quite but you get the picture!) of its former self. OK, it ouptus a unique pattern but not quite what you had in mind! (Note that for the most part, a similar accident with the HR mirror at the other end of the tube wouldn't affect anything but its external appearance.) Is there any hope?

    Well, assuming the chip isn't too deep, it is possible to grind it out and then polish the resulting surface to optical quality. To do this properly will require a means of holding the tube just slightly off of perpendicular (to add some wedge - see below) to a rotating platform on which various grades of wet grinding compound can be introduced starting with something coarse like 400 grit and going up in stages to 1,200 grit or more, following by lapping with optical rouge for the final polish. That should get you a reasonably decent result after considerable effort and cost. But don't expect it to be to 1/10th λ!

    One thing you won't be able to reproduce is the anti-reflective (AR) coating present on most HeNe OCs. (Well, not unless you have access to some vacuum coating equipment!) That is the reason I suggest grinding it on a slight angle - the resulting wedge will divert the reflected beam away from the axis of the cavity and minimize instability and interference.

    I was given a cute little HeNe tube with such a chip in the OC mirror. Now, this certainly wasn't worth spending much of anything to repair (it was only a .8 mW, barcode type HeNe laser after all!). So, I decided to experiment using the minimalist approach: emery paper. I started with 400 grit to remove the chip and then 1,200 grit. I also deliberately attempted to grind the surface parallel to the actual mirror rather than with wedge to see what would happen. All this just by hand so the result is also somewhat convex rather than perfectly flat. I need to find some rouge to attempt the final polishing if I ever bother.

    Even without fine polishing, the beam was much much cleaner than it used to be (formerly being spread out off to one side in random directions!). Just for grins and giggles, I went back to 600 grit to see what effect an even more random ground surface would have on the beam. The interference patterns are really quite interesting - sort of like a stellar globular cluster - so I may just leave it the way it this way. :)

    Another alternative where the area of the beam just touches the chip might be to push the mirror mount side-ways beyond the restricted area. With care, it may be possible to shift it by as much as .5 mm which could be enough.

    Or, use some optical cement to glue a flat piece of glass to the mirror filling the voids. With the proper material that closely matches the index of refraction of the mirror glass, such an approach may result in a beam that isn't too terrible. :)

    Using a Microwave Oven to Evaluate and Revive HeNe Laser Tubes

    WARNING: These are dangerous procedures, at least for your laser tube! Attempt at your own risk. There is a good chance that the tube will be ruined totally as a result of the glass or glass-to-metal seals cracking. There is also a chance that the procedures will make the situation worse or that any apparent improvement will be temporary.

    Note that using a microwave oven is safe for just checking to see if the tube is gas-intact and has approximately the correct discharge color. In this case, only a second or two is needed so heating is minimal. See the section: How Can I Tell if My Tube is Good?.

    The whacko procedures below may be used to provide an idea of what is wrong with a HeNe tube as well as to at least partially revive them in some cases. The difference between evaluation and revival is basically in cooking time and how many times the procedure is repeated.

    Alternative sources of RF energy can be used in place of the kitchen microwave but may not be quite as convenient or as readily available. :)

    I have had some modest success in at least partially reviving some old soft-seal HeNe laser tubes with the power output from 4 of 6 weak tubes being improved significantly, though not to anywhere near the rated specifications. However, one tube was destroyed due to the glass cracking (the first one I tried, not having a feel for the safe cook time), and on another, the power went down slightly. To what extent these results are due to getter reactivation or other phenomena is not currently known. The effects of the microwaves (whether it be from the discharge or just due to heating) would also appear to be useful as a diagnostic tool for evaluating HeNe tube condition.

    Since the entire tube or whatever has to be inside the oven (don't even think about drilling holes in the side or door!), this stunt probably only applies to smaller helium-neon laser tubes and maybe the getters in receiving tubes if you remember what they are. :) Here goes:

    However, what you may find is that the power increases immediately after treatment but then decays back to its original value or below over the span of a day or so, or faster if the tube is powered. But, this may be repeatable, so if you just need a temporary boost, go for it! :)

    I would appear that the microwave treatment may do any or all of the following: